← Product Code [LJT](/productcode/LJT) · K962695

# PORT-A-CATH II VENOUS IMPLANTABLE ACCESS SYSTEMS PRECONNECTED WITH POLYURETHANE CATHETER (K962695)

_Sims Deltec, Inc. · LJT · Oct 1, 1996 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K962695

## Device Facts

- **Applicant:** Sims Deltec, Inc.
- **Product Code:** [LJT](/productcode/LJT.md)
- **Decision Date:** Oct 1, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5965
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

PORT-A-CATH® II Venous Systems Preconnected with Polyurethane Catheter are indicated when a patient requires repeated venous access for injection or infusion therapy and/or venous blood sampling.

## Device Story

Implantable venous access system consisting of titanium/polysulfone portal and radiopaque polyurethane catheter. Device provides subcutaneous access to venous system for repeated infusion, injection, or blood sampling. Portal/catheter connection is pre-assembled during manufacturing using polyurethane strain relief. System includes access needle and optional introducer sets. Used by clinicians in hospital or clinical settings to facilitate long-term vascular access; reduces need for repeated venipuncture; improves patient comfort and safety during therapy administration.

## Clinical Evidence

No clinical data. Bench testing only. In-vitro mechanical testing performed on the catheter-to-portal connection per FDA guidance. Biocompatibility testing conducted on all system components.

## Technological Characteristics

Portal housing: titanium/polysulfone. Septum: silicone. Catheter: radiopaque polyurethane. Dimensions: 1.9 mm or 2.6 mm O.D. catheters. Pre-connected portal/catheter assembly with polyurethane strain relief. Mechanical device; no software or energy source.

## Regulatory Identification

A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.

## Predicate Devices

- PORT-A-CATH® II Dual Lumen Venous Implantable Access Systems Preconnected with Silicone Catheter ([K955407](/device/K955407.md))
- PORT-A-CATH® II Venous Implantable Access Systems with Polyurethane Catheter ([K932840](/device/K932840.md))
- PORT-A-CATH® II Venous Implantable Access Systems with Polyurethane Catheter ([K942024](/device/K942024.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K962695

OCT 1 1996

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

PORT-A-CATH® II Implantable Venous Access Systems
Preconnected with Polyurethane Catheter

July 8, 1996

## I. GENERAL INFORMATION

Applicant's Name and Address: SIMS Deltec, Inc.
1265 Grey Fox Road
St. Paul, MN 55112

Contact Person: Lisa Stone
Manager, Regulatory Affairs
Tel. (612) 628-7224

Common/Usual Name: Subcutaneously Implanted Intravascular Infusion Port and Catheter

Proprietary Name: PORT-A-CATH® II Venous Implantable Access Systems Preconnected with Polyurethane Catheter

Equivalence Device Comparison: PORT-A-CATH® II Venous Implantable Access Systems with Polyurethane Catheter (manufactured by SIMS Deltec, Inc.)

PORT-A-CATH® II Dual Lumen Venous Implantable Access Systems Preconnected with Silicone Catheter (manufactured by SIMS Deltec, Inc.)

## II. DEVICE DESCRIPTION

PORT-A-CATH® II Venous Systems Preconnected with Polyurethane Catheter are similar in design and function to the current commercially available PORT-A-CATH® II Venous Systems with Polyurethane Catheter.

These systems differ from current commercially available systems in that the portal/catheter connection is made during the manufacturing process, by placing a polyurethane strain relief over the catheter and outlet tubes of the portal.

These systems consist of a standard or low profile portal (titanium/polysulfone) preconnected to a radiopaque catheter (polyurethane) and an access needle. Two catheter options will be available: a 2.6 mm O.D. catheter and a 1.9 mm O.D. catheter. Additionally, systems will be made available with introducer sets.

{1}

510(k) Summary of Safety and Effectiveness
Page 2 of 3

## III. INTENDED USE OF DEVICE

PORT-A-CATH® II Venous Systems Preconnected with Polyurethane Catheter are indicated when a patient requires repeated venous access for injection or infusion therapy and/or venous blood sampling.

## IV. DEVICE COMPARISON

|   | PORT-A-CATH® II Systems Preconnected with Polyurethane Catheter | PORT-A-CATH® II Dual Lumen Systems Preconnected with Silicone Catheter K955407 | PORT-A-CATH® II Systems with Polyurethane Catheter K932840 K942024  |
| --- | --- | --- | --- |
|  MANUFACTURER | SIMS Deltec, Inc. | SIMS Deltec, Inc. | SIMS Deltec, Inc.  |
|  INDICATIONS FOR USE | A system is indicated when a patient requires repeated venous access for injection (bolus) or infusion therapy and/or venous blood sampling. | A system is indicated when a patient requires repeated venous access for injection (bolus) or infusion therapy and/or venous blood sampling. | A system is indicated when a patient requires repeated venous access for injection (bolus) or infusion therapy and/or venous blood sampling.  |
|  PORTAL DIMENSIONS |  |  |   |
|  Height | STANDARD LOW-PROFILE 14.7 mm 11.5 mm | DUAL-LUMEN 16.0 mm | STANDARD LOW-PROFILE 14.7 mm 11.5 mm  |
|  Base | 30.5 mm 25.0 mm | 50.0 (L) mm x 30.0 (W) mm | 30.5 mm 25.0 mm  |
|  Septum Diameter | 11.4 mm 9.5 mm | 11.4 mm | 11.4 mm 9.5 mm  |
|  MATERIALS |  |  |   |
|  Portal Housing Septum Catheter | Titanium/Polysulfone Silicone Polyurethane | Titanium/Polysulfone Silicone Silicone | Titanium/Polysulfone Silicone Polyurethane  |
|  CATHETER DIMENSIONS |  |  |   |
|  I.D. | 1.0 mm 1.6 mm | 1.1 mm/1.1 mm | 1.0 mm 1.6 mm  |
|  O.D. | 1.9 mm 2.6 mm | 3.4 mm | 1.9 mm 2.6 mm  |
|  Length | 76 cm 76 cm | 76 cm | 76 cm 76 cm  |
|  PORTAL/CATHETER PRECONNECTED | YES | YES | NO  |

{2}

510(k) Summary of Safety and Effectiveness
Page 3 of 3

# V. SUMMARY OF STUDIES

## A. Functional Testing

*In-vitro* mechanical testing was conducted according to the "Guidance on 510(k) Submissions for Implanted Infusion Ports," October 1990. Only catheter to portal connection testing was performed, since these preconnected systems are the same as current commercially available systems, except for the connection scheme.

Biocompatibility testing was conducted on all system components.

## B. Clinical Studies

Clinical studies were not deemed necessary regarding PORT-A-CATH® II Venous Systems Preconnected with Polyurethane Catheter due to their similarity in materials, design and function to current SIMS Deltec commercially available systems.

## C. Conclusions Drawn from the Studies

The results of the testing indicated that PORT-A-CATH® II Venous Systems Preconnected with Polyurethane Catheter function according to specification and the materials used in the system are biocompatible. Therefore, these systems are considered acceptable for human use.

---

**Source:** [https://fda.innolitics.com/device/K962695](https://fda.innolitics.com/device/K962695)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
