K962600 · Rexam Medical Packaging, Inc. · KCT · Aug 9, 1996 · General Hospital
Device Facts
Record ID
K962600
Device Name
REXAM SELF-SEAL TYVEK POUCH
Applicant
Rexam Medical Packaging, Inc.
Product Code
KCT · General Hospital
Decision Date
Aug 9, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Intended Use
The Rexam Self-Seal Tyvek® Pouch is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider using a sterilization process where a non-cellulose pouch is indicated. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The chemical process indicator on the Rexam Self-Seal Tyvek® Pouch is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Device Story
Rexam Self-Seal Tyvek® Pouch serves as a sterile barrier system for medical devices. Constructed from medical-grade Tyvek® and polyethylene/polyester film; features pressure-sensitive self-seal tape closure for manual sealing by healthcare providers. Optional chemical sterilization process indicator provides visual confirmation of sterilization parameters. Used in clinical settings to enclose instruments prior to sterilization; maintains sterility until point of use. Device facilitates manual sealing as an alternative to heat sealing.
Clinical Evidence
Bench testing only. Seal strength and seal integrity tests were performed and compared to the predicate device. Seal strength results were statistically analyzed at a 95% confidence level, confirming performance significantly above minimum limits. No clinical data provided.
Technological Characteristics
Materials: Medical grade Tyvek® sealed to polyethylene/polyester film. Closure: One-inch wide pressure-sensitive self-seal tape. Optional chemical sterilization process indicator. Non-sterile device intended for use in sterilization processes.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices requiring sterilization in processes where non-cellulose pouches are appropriate; maintains sterility post-sterilization. Optional chemical indicator monitors sterilization process parameters.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Medi-Oxide™ Pouch
Related Devices
K070765 — VIS-U-ALL II SELF SEAL POUCH · STERIS Corporation · Oct 12, 2007
K141412 — STERILUCENT SELF-SEAL STERILIZATION POUCH · Gs Medical Packaging, Inc. · Dec 3, 2014
K973827 — KIMBERLY-CLARK SELF-SEAL POUCH FOR LOW TEMPERATURE STERILIZATION · Kimberly-Clark Corp. · Feb 11, 1998
K051242 — WINNER SELF SEAL STERILIZATION POUCH · Winner Medical (Usa), Inc. · Dec 13, 2005
K152669 — PMSSteripack Self-Seal Sterilization Pouch with Chemical Indicator (KP) · Pms Tip Teknolojileri Sanayi VE Ticaret Limited Std · Jan 15, 2016
Submission Summary (Full Text)
{0}
AUG-07-1996 09:38 FROM
TO 913014803002 P.06
# REXAM
*X962600*
AUG - 9 1996
## 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K962600
### Applicant Information:
- **Date Prepared:** August 5, 1996
- **Name:** Rexam Medical Packaging
- **Address:** 1919 S. Butterfield Road, Mundelein, IL 60060-9735
- **Contact Person:** Quality and Regulatory Manager
- **Phone Number:** (847) 362-9000
- **Fax Number:** (847) 918-4660
### Device Information:
- **Trade Name:** Rexam Self-Seal Tyvek¹ Pouch
- **Common Name:** Self-Seal Tyvek®/Film Sterilization Pouch with or without Chemical Sterilization Process Indicator
- **Classification Name:** Sterilization Wrap, Chemical Sterilization Process Indicator
### Predicate Device:
- **Trade Name:** Medi-Oxide™ Pouch
- **Manufacturer:** Cur*Med, Curwood, Inc., a Bemis Company
### Device Description:
The Rexam Self-Seal Tyvek® Pouch is constructed of medical grade Tyvek® sealed to polyethylene/polyester film. A chemical sterilization process indicator may be incorporated to monitor one or more of the sterilization process parameters. The final seal of the pouch is applied by the health care provider using the pressure sensitive self-seal tape closure. The self-seal tape closure facilitates manual sealing as an alternate method to heat sealing.
¹Tyvek is a registered trademark of E.I. duPont de Nemours & Company.
Page 2
{1}
AUG-07-1996 09:39 FROM
TO 913014803002 P.07
# REXAM
## Intended Use:
The Rexam Self-Seal Tyvek® Pouch is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider using a sterilization process where a non-cellulose pouch is indicated. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The chemical process indicator on the Rexam Self-Seal Tyvek® Pouch is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
## Comparison of Technological Characteristics:
The Rexam Self-Seal Tyvek® Pouch and the Medi-Oxide™ Pouch are both constructed of medical grade Tyvek® sealed to polyethylene/polyester film. The pressure sensitive tape closure on the Rexam Self-Seal Tyvek® Pouch and the Medi-Oxide™ Pouch are both one inch in width, and the seal is applied by the customer in the same manner for both products. The Medi-Oxide™ Pouch is manufactured without a chemical sterilization process indicator, and the Rexam Self-Seal Tyvek® Pouch may be manufactured with or without a chemical sterilization process indicator.
The performance of the self-seal tape closure on the Rexam Self-Seal Tyvek® Pouch was measured by seal strength and seal integrity tests, and compared to the performance of the self-seal tape on the Medi-Oxide™ Pouch. The seal strength test results were statistically analyzed at a 95% confidence level, and the test results indicate that both products were significantly above the minimum seal strength limit. The Rexam Self-Seal Tyvek® Pouch performed as well as the Medi-Oxide™ Pouch on the seal integrity tests.
The information submitted includes data that demonstrates that the Rexam Self-Seal Tyvek® Pouch is as safe and effective as the legally marketed Medi-Oxide™ Pouch, and that there are no new issues raised of safety and efficacy.
Page 3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.