← Product Code [MKO](/productcode/MKO) · K962587

# CONCEPTUS FALLOPIAN TUBE CATHETEIZATION WITH MICROENDOSCOPY SYSTEM (K962587)

_Conceptus, Inc. · MKO · Jan 31, 1997 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K962587

## Device Facts

- **Applicant:** Conceptus, Inc.
- **Product Code:** [MKO](/productcode/MKO.md)
- **Decision Date:** Jan 31, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1690
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Conceptus FTCM System is intended for use in selective catheterization/cannulation of the fallopian tubes in order to evaluate proximal tubal occlusion or patency under direct visualization.

## Device Story

The FTCM System is a set of catheters, guidewires, and a microendoscope used for fallopian tube catheterization and visualization. The system includes a Soft Seal Cervical Catheter for access, a Soft Torque Uterine Catheter for guiding instruments to the utero-tubal ostium, and a Variable Softness Catheter for accessing the proximal fallopian tube. Guidewires provide column strength for catheter positioning. A microendoscope and articulating hysteroscope provide direct visualization of the cervical canal, uterine cavity, and tubal ostia. The device is operated by a physician in a clinical setting. The optical system conveys images to an eyepiece or video camera/monitor. By enabling direct visualization and selective cannulation, the device assists in diagnosing tubal patency or occlusion, potentially guiding further clinical management of infertility or tubal disease.

## Clinical Evidence

Clinical data provided to substantiate equivalence; specific study design, sample size, and performance metrics not detailed in the provided summary text.

## Technological Characteristics

System components include catheters (Soft Seal, Soft Torque, Variable Softness), guidewires, articulating hysteroscope, microendoscope, and eyepiece. Materials include latex (balloon). Sensing/actuation is mechanical/manual. Visualization via fiber-optic microendoscope and telescopic system. Connectivity is via standard ACMI-type light connectors and optical image transmission. No software or electronic processing described.

## Regulatory Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

## Predicate Devices

- Conceptus FTC System ([K962587](/device/K962587.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

CONFIDENTIAL
FEB - 3 1997
K962587
P193

# XVI. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## Contact Person

Nancy Naddaff
Clinical Research and Regulatory Affairs Specialist
Conceptus, Inc.
1021 Howard Avenue
San Carlos, CA 94070

## Device Description

The FTCM System combines well-established procedures of HSG, co-axial catheterization and microendoscopy. The FTCM System is composed of the following components: Soft Seal Cervical Catheter, Soft Torque Uterine Catheter and Accessory Valve, and/or Articulating Tip Hysteroscope, Variable Softness Catheter, guidewires: the Robust Guidewire, the Traveler Guidewire, the Supple Guidewire and the Pirouette Guidewire, and a Microendoscope with Eyepiece. Each component is described below;

### a. Soft Seal Cervical Catheter

The Soft Seal Cervical Catheter is a dual lumen device with a latex balloon located at the distal end. The Cervical Catheter is packaged with a malleable Obturator which can be shaped with a gentle curve and will increase column strength of the catheter when inserted completely into the Cervical Catheter. The proximal end of the Cervical Catheter has two ports, one to inflate the balloon and one to access the through-lumen. A two-way stopcock on the balloon port maintains inflation during use. The through-lumen allows injection of dye, contrast medium, or sperm into the uterus and/or serves as a conduit for the Obturator and/or Conceptus Uterine Catheter.

### b. Soft Torque Uterine Catheter, Mandrel and Accessory Valve

The Soft Torque Catheter serves as a conduit to the proximal portion of the fallopian tube for the Conceptus Variable Softness Catheter, Guidewire and Microendoscope. The Soft Torque Catheter is a single lumen device with a curved distal tip. The Catheter is packaged with a Mandrel which will straighten the tip curve when inserted completely into the catheter. The hub located at the proximal end of the Catheter has two ports. The side port is used for the attachment of a syringe for injection and is in the same plane and direction as the tip curve. The end port is used for introduction of the Mandrel and / or Variable Softness Catheter and has a valve to prevent reflux during injection. The Soft Torque Catheter can access the utero-tubal ostium through a rigid hysteroscope with direct visualization.

### c. Articulating Tip Hysteroscope

The Articulating Tip Hysteroscope is used to permit direct viewing of the cervical canal, the uterine cavity and utero-tubal ostium by a telescopic system introduced into the uterus through the cervix. The scope has a rigid shaft with a flexible articulating tip. The device is

Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM) 510(k)
232

{1}

CONFIDENTIAL
233

equipped with a separate eyepiece plug and light guide plug. The optical system conveys an image to the user's eye or to a video camera.

d. Variable Softness Catheter

The Variable Softness Catheter is an over-the-wire catheter which is intended to access the proximal portion of the fallopian tube. The catheter is used through the Conceptus Soft Torque Uterine Catheter or Articulating Tip Hysteroscope. It is a single lumen 3 French (1 mm) catheter with graded shaft flexibility. The hub located at the proximal end of the catheter has two ports. The side port is used for the attachment of an irrigation tube or a syringe for injection. The endport is used for introduction of the guidewire and microendoscope and has a valve to prevent reflux during injection of dye and/or irrigation fluid. The catheter has a proximal marker which indicates when the Conceptus Variable Softness Catheter is about to exit the Conceptus Soft Torque Catheter.

e. Guidewire

The Conceptus Guidewires are intended for use to increase column strength of the Variable Softness Catheter during positioning within the proximal portion of the fallopian tube. The guidewires have a tapered core wire and a distal coil section for flexibility, and a proximal handle to lock the guidewire into position with the Variable Softness Catheter. The two models differ only with respect to the wire diameter profile and the mid-to-distal coil length. This difference in taper and coil lengths imparts varying degrees of flexibility to the guidewire, and guidewire selection will be based on physician preference.

f. Microendoscope

The Conceptus Microendoscope is an ultra-thin, flexible fiberscope designed for use with the Conceptus Variable Softness Catheter. It is to be used with a Conceptus Eyepiece which is detachable and has adjustable focus. The microendoscope is equipped with a separate image plug and light plug. The light plug has an ACMI-type male connector. The length of the cable provides easy handling when imaging on a television monitor.

g. Eyepiece

The Conceptus detachable Eyepiece is designed to fit onto the image plugs of the Conceptus Microendoscope. The Eyepiece has an adjustable focus and an alignment marker on its tip. The cleaning plug should always be inserted for cleaning and storage.

**Indications for Use**

The Conceptus FTCM System is intended for use in selective catheterization/cannulation of the fallopian tubes in order to evaluate proximal tubal occlusion or patency under direct visualization.

Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM) 510(k)

{2}

CONFIDENTIAL
K962587
P393

# Substantial Equivalence

The Conceptus FTCM System is substantially equivalent to the Conceptus FTC System. Comparative performance testing demonstrated that the Conceptus FTCM System is substantially equivalent to the Conceptus FTC System with respect to force required to navigate the fallopian tube model. Additionally, performance testing and clinical data were provided for the Conceptus FTCM System which further substantiated equivalence.

Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM) 510(k)
234

---

**Source:** [https://fda.innolitics.com/device/K962587](https://fda.innolitics.com/device/K962587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
