CARDIOSERV P
Device Facts
| Record ID | K962547 |
|---|---|
| Device Name | CARDIOSERV P |
| Applicant | Marquette Electronics, Inc. |
| Product Code | DRO · Cardiovascular |
| Decision Date | Mar 6, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5550 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
CardioServ is intended to be used for the emergency resuscitation of cardiac arrest victims and clinical cardiac dysrhythmia. CardioServ is intended to be used by trained operators CardioServ is designed for external and internal defibrillation (including cardioversion) CardioServ is capable of monitoring the heart rate with adjustable alarm limits. CardioServ is designed for external pacing with adjustable current and frequency.
Device Story
Portable defibrillator with integrated ECG monitor, recorder, and external transcutaneous pacemaker; used for emergency resuscitation and dysrhythmia management. Device inputs include cardiac electrical signals via electrodes/paddles; processes signals for heart rate monitoring and alarm triggering; delivers therapeutic electrical shocks or pacing pulses. Operated by trained clinicians in clinical settings. Output includes real-time ECG visualization, heart rate data, and therapeutic energy delivery. Assists clinicians in rapid cardiac rhythm assessment and intervention; facilitates life-saving resuscitation.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI DF2-1989, ANSI/AAMI ES1 1993, IEC 601-1, IEC 601-1-2, IEC 601-2-4, and EN 60601-2-31. Quality assurance included requirements specification reviews, code inspections, software/hardware testing, safety/environmental testing, and independent final validation.
Technological Characteristics
Portable defibrillator/monitor/pacemaker. Sensing: ECG monitoring. Actuation: Defibrillation and external pacing (adjustable current/frequency). Standards: ANSI/AAMI DF2-1989, ANSI/AAMI ES1 1993, IEC 601-1, IEC 601-1-2, IEC 601-2-4, EN 60601-2-31. Includes built-in recorder.
Indications for Use
Indicated for cardiac arrest victims and patients with clinical cardiac dysrhythmia requiring emergency resuscitation, defibrillation, cardioversion, or external pacing. Intended for use by trained operators.
Regulatory Classification
Identification
An external transcutaneous cardiac pacemaker (noninvasive) is a device used to supply a periodic electrical pulse intended to pace the heart. The pulse from the device is usually applied to the surface of the chest through electrodes such as defibrillator paddles.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “American National Standards Institute/American Association for Medical Instrumentation's DF-21 ‘Cardiac Defibrillator Devices’ ” 2d ed., 1996, and (2) “The maximum pulse amplitude should not exceed 200 milliamperes. The maximum pulse duration should not exceed 50 milliseconds.”
Predicate Devices
- HELLIGE CardioServ SCP 910
- Marquette Series 1500 Responder
Related Devices
- K962551 — CARDIOSERV S · Marquette Electronics, Inc. · Nov 19, 1996
- K963120 — RESPONDER 2500 DEFIBRILLATOR AND CARDIAC CARE SYSTEM · Marquette Electronics, Inc. · Feb 28, 1997
- K040637 — POWERHEART AED G3 PRO · Cardiac Science, Inc. · Aug 6, 2004
- K972241 — ZOLL REVIVER · Zoll Medical Corp · Mar 16, 1998
- K200849 — Tempus LS - Manual · Remote Diagnostic Technologies, Ltd. A Philips Company · Jul 24, 2020