HEARTPORT CABLE BULLDOG CLAMP

K962366 · Heartport, Inc. · DXC · Nov 7, 1996 · Cardiovascular

Device Facts

Record IDK962366
Device NameHEARTPORT CABLE BULLDOG CLAMP
ApplicantHeartport, Inc.
Product CodeDXC · Cardiovascular
Decision DateNov 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Heartport Cable Bulldog Clamp is intended for the temporary occlusion of a vessel, including the Internal Mammary Artery or vein grafts, and other clamping procedures. The Cable Bulldog Clamp may be introduced to surgical sites through a thoracic port or incision for use during endoscopic cardiac surgical procedures.

Device Story

Device is a vascular clamp for temporary vessel occlusion; used during endoscopic cardiac surgery. Introduced to surgical site via thoracic port or incision. Provides clamping force to occlude vessels like Internal Mammary Artery or vein grafts. Operated by surgeons to facilitate cardiac procedures. Benefits include enabling minimally invasive access to vessels.

Clinical Evidence

Bench testing only. In vitro evaluation of clamping force demonstrated performance equivalent to predicate devices and meeting clinical requirements. Biocompatibility of materials established.

Technological Characteristics

Vascular clamp; mechanical actuation; materials similar to predicate devices with proven biocompatibility. Designed for endoscopic delivery via thoracic port.

Indications for Use

Indicated for temporary occlusion of vessels, including Internal Mammary Artery or vein grafts, during endoscopic cardiac surgical procedures.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Heartport™ Cable Bulldog Clamp Appendices # APPENDIX A. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. NOV - 7 1996 The assigned 510(k) number is: K962366 ## Applicant Information: Date Prepared: June 17, 1996 Name: Heartport, Inc. Address: 200 Chesapeake Drive Redwood City, CA 94063 Contact Person: Ms. Isabella R. Abati Director, Regulatory Affairs Phone Number: (415) 306-7900 Fax Number: (415) 306-7905 ## Device Information: Classification: Class II Trade Name: Heartport™ Cable Bulldog Clamp Common Name: Cardiovascular Devices - Vascular Clamp, 21 CFR 870.4450 Product Code: 74DXC ## Equivalent Devices: The Heartport Cable Bulldog Clamp are substantially equivalent in intended use and/or method of operation to a combination of the following predicate devices: 1. Baxter Healthcare - Fogarty® Spring Clips (Softjaw®, Safejaw®, and Double Safejaw® Spring Clips) and Spring Clip Applicators 2. Baxter Healthcare - EDWARDS Parallel Jaw Spring Clip and Applicator Forceps 3. Baxter Healthcare - EDWARDS Single Softjaw and Double Softjaw Handleless Clamps Heartport, Inc. CONFIDENTIAL Page 24 {1} Heartport™ Cable Bulldog Clamp Appendices # 510(k) Summary of Safety and Effectiveness (continued) ## Intended Use: The Heartport Cable Bulldog Clamp is intended for the temporary occlusion of a vessel, including the Internal Mammary Artery or vein grafts, and other clamping procedures. The Cable Bulldog Clamp may be introduced to surgical sites through a thoracic port or incision for use during endoscopic cardiac surgical procedures. ## Comparison To Predicate Devices: The Heartport Cable Bulldog Clamps are equivalent in intended use and operational characteristics to the following devices: (1) Baxter Healthcare - Fogarty® Spring Clips (Softjaw®, Safejaw®, and Double Safejaw® Spring Clips) and Spring Clip Applicators; (2) Baxter Healthcare - EDWARDS Parallel Jaw Spring Clip and Applicator Forceps; and (3) Baxter Healthcare - EDWARDS Single Softjaw and Double Softjaw Handleless Clamps The Heartport Cable Bulldog Clamp combines the same functions as the individual predicate devices into a single device. ## Non-Clinical Test Results: ### Performance The Heartport Cable Bulldog Clamp was evaluated *in vitro* for clamping force. The performance testing demonstrated that the Heartport Cable Bulldog Clamp equivalent to the predicate device tested and meets the anticipated clinical requirements for its intended use. ### Biocompatibility The materials used to fabricate the Heartport Cable Bulldog are similar to the predicate devices. All materials used in the Heartport Cable Bulldog Clamp have proven biocompatibility. ## Summary: Based on the intended use, product information, performance and biocompatibility data provided in this Premarket Notification, Heartport Cable Bulldog Clamps have been shown to be substantially equivalent to currently marketed predicate devices. Heartport, Inc. CONFIDENTIAL Page 25
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%