AMO ARTHOSCOPES
K962330 · American Microptiks, Inc. · HRX · Sep 11, 1996 · Orthopedic
Device Facts
| Record ID | K962330 |
| Device Name | AMO ARTHOSCOPES |
| Applicant | American Microptiks, Inc. |
| Product Code | HRX · Orthopedic |
| Decision Date | Sep 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.1100 |
| Device Class | Class 2 |
Intended Use
Like other legally marketed arthroscopes, the AMO Arthroscope is indicated for use in minimally invasive procedures of the knee, shoulder and ankle joints, for joint examination, biopsy, arthroscopy, diagnosis of joint disease. The device provides a clear view of the joint structures such as femoral condyles, medial meniscus, cruciate ligament, and other anatomical structures.
Device Story
AMO Arthroscope; rigid endoscope for visualization of joint structures (femoral condyles, meniscus, ligaments). Input: optical light transmission through fiber optics; output: direct visualization of joint anatomy via eyepiece. Used in clinical arthroscopic procedures by surgeons. Device construction: eyepiece, body, light guide, shaft (outer/inner tubes with fiber optics). Non-sterile delivery; requires user-performed sterilization (ethylene oxide gas) prior to use. Provides clear view for diagnosis and biopsy; facilitates minimally invasive joint surgery.
Clinical Evidence
No clinical data. Substantial equivalence based on design, material, and functional comparison to predicate device.
Technological Characteristics
Rigid arthroscope; construction includes eyepiece, body, light guide, and shaft. Shaft contains outer tube, inner tube, and fiber optic light-carrying fibers. Optical system housed in inner tube. Sterilization: ethylene oxide gas (user-performed).
Indications for Use
Indicated for minimally invasive procedures of knee, shoulder, and ankle joints, including joint examination, biopsy, arthroscopy, and diagnosis of joint disease.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
- Storz (Karl Storz) 7200BW Arthroscope
Related Devices
- K031972 — ASAP ARTHROSCOPE MODELS; 10-0001-00, 10-0002-00, 10-0003-00, 10-0004-00, 10-0005-00, 10-0006-00, 10-0007-00, 10-0008-00, · Asap Endoscopic Products GmbH · Sep 15, 2003
- K133018 — OSI Q SYSTEM ARTHROSCOPES · Orthopedic Sciences, Inc. · Feb 26, 2014
- K112548 — ALTOMEC ARTHROSCOPE · Altomec Endoscopy, Inc. · Aug 15, 2012
- K080560 — HENKE SASS WOLF OF AMERICA ARTHROSCOPE · Henke Sass Wolf of America, Inc. · Mar 18, 2008
- K202659 — Stryker Arthroscope · Stryker · Dec 1, 2020
Submission Summary (Full Text)
{0}
SEP 11 1996
# SUMMARY OF SAFETY AND EFFECTIVENESS
American Microptiks, Inc. "AMO Arthroscope"
501K Number K962330
Updated as of September 6th, 1996
1. Submitter information:
| Name and Address | American Microptiks, Inc.
420 East Main St.
Building 3, Unit 18
Branford CT 06437 |
| --- | --- |
| Phone | (203) 483-6428 |
| Fax | (203) 483-9829 |
| Contact Person | Dr. George Hallack |
2. Device Name:
Classification Name: Arthroscope
Proprietary Name: AMO Arthroscope
3. Predicate Device:
An equivalent device to the AMO Arthroscope is the Storz (Karl Storz) 7200BW Arthroscope, manufactured in Germany at:
Mittelstrasse 8
D-4200 Tuttlingers
Postfach 230
This device is distributed in this country by Karl Storz Endoscopy - America, Inc, Charleston MA.
The above identified equivalent device uses the same basic design principles and technological features as the proposed AMO Arthroscope.
{1}
Updated 9/6/96
4. Description of Device:
The AMO Arthroscope is a device that will be provided non-sterile for arthroscopic procedures. The device has a basic design similar to those already sold legally in the U.S.A. Its construction consists of an eyepiece, body, light guide and a shaft. The shaft contains an outer tube, an inner tube, with light-carrying fibers sandwiched between these tubes. The inner tube contains the optical system.
5. Indications for Use:
Like other legally marketed arthroscopes, the AMO Arthroscope is indicated for use in minimally invasive procedures of the knee, shoulder and ankle joints, for joint examination, biopsy, arthroscopy, diagnosis of joint disease.
The device provides a clear view of the joint structures such as femoral condyles, medial meniscus, cruciate ligament, and other anatomical structures.
6. Description of Safety and Substantial Equivalence:
The biological safety of the AMO arthroscope has been defined through the selection of materials that demonstrated appropriate levels of biocompatibility. The proposed AMO Arthroscope was designed with similar base materials. These materials are similar to those used in other legally marketed brands of arthroscopes, including the predicate device which is marketed for the same indications.
In summary, biocompatibility, function and device design have been developed to ensure the device is safe, effective, and substantially equivalent in materials, function and intended use to commercially available arthroscopes.
The device will be sold non-sterile, to be sterilized prior to each procedure by the user. Our recommended sterilization method, to be validated, is ethylene oxide gas.