← Product Code [DSP](/productcode/DSP) · K962177

# BARD TAPERSEAL INTRA-AORTIC BALLOON CATHETER & INTRODUCER KIT (K962177)

_C.R. Bard, Inc. · DSP · Sep 4, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K962177

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [DSP](/productcode/DSP.md)
- **Decision Date:** Sep 4, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3535
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Refractory left ventricular failure, Cardiogenic or septic shock, Unstable refractory angina, Impending infarction, Ischemia related ventricular arrhythmias, Weaning from cardiopulmonary bypass, Support and stabilization during coronary angioplasty, Intraoperative pulsatile flow generation, Associated mechanical complications of acute myocardial infarction, Support and stabilization of high-risk patients undergoing general surgical procedures.

## Device Story

Intra-aortic balloon (IAB) catheter; percutaneous insertion; double-lumen design; pre-furled balloon; integrated sliding hemostasis device for puncture site bleeding control. Used in clinical settings (OR/ICU) by physicians for hemodynamic support. Device functions as mechanical circulatory assist; inflates/deflates to augment cardiac output; reduces myocardial oxygen demand. Kit includes insertion accessories (introducer, guidewires, needle, dilator) and console adaptors. Output is mechanical pulsatile flow; assists clinical decision-making by stabilizing hemodynamics in acute cardiac failure or high-risk surgical patients.

## Clinical Evidence

Bench testing only. Testing verified the catheter's ability to withstand insertion forces associated with sheathless insertion. No clinical data provided.

## Technological Characteristics

Double-lumen catheter; materials include polyurethane, silicone, stainless steel, polycarbonate, polyolefin, and thermoplastic elastomer. Features pre-furled balloon and integrated sliding hemostasis device. Dimensions: 30cc and 40cc volumes. Mechanical device; no energy source or software.

## Regulatory Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Sheathless 40cc Pre-Wrapped Double-Lumen 9.0 F. Intra-Aortic Balloon Catheter (Model IAB-04240S)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K962177

Bard Vascular Systems Division

C.R. Bard, Inc.

25 Computer Drive

Haverhill, MA 01832

508-373-1000

SEP - 4 1996

BARD

## VI. 510(k) SUMMARY FOR THE BARD® TAPERSEAL® IAB CATHETERS

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

### A. Submitter Information

Submitter’s Name: Bard Vascular Systems, C.R. Bard Inc.

Address: 25 Computer Drive, Haverhill, MA 01832

Phone #: (508) 373-1000

Fax #: (508) 374-6200

Contact Person: Sandra L. Perreand

Date of Preparation: May 29, 1996

### B. Device Name

Trade Name: Bard® TaperSeal® IAB Catheters

Common/Usual Name: Intra-Aortic Balloon Catheter

Classification Name: Intra-Aortic System, Balloon, Intra-Aortic and Control

### C. Predicate Device

Trade Name: Sheathless 40cc Pre-Wrapped Double-Lumen 9.0 F. Intra-Aortic Balloon Catheter. (Model IAB-04240S)

### D. Device Description

Each double lumen percutaneous TaperSeal® IAB catheter is pre-furled and has a hemostasis device attached to the catheter which may be slid down the catheter and into the insertion site as necessary in the event of excessive bleeding from the puncture site. Each TaperSeal® IAB catheter is packaged into a balloon kit that contains airway tubing, a syringe, syringe adaptor, Kontron and/or Datascope adaptor, and pressure tubing with a stopcock. The TaperSeal® IAB catheters also come packaged with an insertion kit which contains a percutaneous tear away introducer, two guidewire, an angiography needle, and a vessel dilator.

### E. Intended Use

- Refractory left ventricular failure,

- Cardiogenic or septic shock,

- Unstable refractory angina,

- Impending infarction,

{1}

-Ischemia related ventricular arrhythmias,
-Weaning from cardiopulmonary bypass,
-Support and stabilization during coronary angioplasty,
-Intraoperative pulsatile flow generation,
-Associated mechanical complications of acute myocardial infarction,
-Support and stabilization of high-risk patients undergoing general surgical procedures.

## F. Technological Characteristics Summary

Both the TaperSeal and Arrow IAB catheters are very similar, with only a few minor differences with respect to design and materials.

|  Characteristic | Bard* TaperSeal* 30cc & 40cc IAB Catheters | Arrow Sheathless 40cc Pre-Wrapped Double-Lumen 9.0 F. IAB Catheter  |
| --- | --- | --- |
|  Balloon Membrane | Polyurethane/Silicone | Polyurethane/Silicone (Cardiothana™)  |
|  Outer Lumen | Polyurethane | Polyurethane  |
|  Inner Lumen | Stainless Steel coiled around a Polyurethane tube | Stainless Steel Tube  |
|  Catheter Tip | Polyurethane | Stainless Steel  |
|  Bifurcation | Polycarbonate | Polyolefin  |
|  Tracer Ring | Stainless Steel | None  |
|  Tie down | Thermo Plastic Elastomer | Thermo Plastic Elastomer  |
|  Hemostasis Device | Polyolefin | Polyolefin  |
|  Hemostasis seal | Silicone | None  |
|  Hemostasis Cap | Nylon | None  |

## G. Performance Data

Bench testing was performed to verify that the TaperSeal Catheter can withstand insertion forces associated with sheathless insertion. All testing was completed successfully.

![img-0.jpeg](img-0.jpeg)

Sandra L. Perreand
Regulatory Affairs Program Manager

9/3/96

Date

---

**Source:** [https://fda.innolitics.com/device/K962177](https://fda.innolitics.com/device/K962177)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).