← Product Code [DQA](/productcode/DQA) · K962156

# 3304 PULSE OXIMETER (K962156)

_Bci Intl., Inc. · DQA · Aug 28, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K962156

## Device Facts

- **Applicant:** Bci Intl., Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Aug 28, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The 3304 provides fast, reliable SpO2, pulse rate and pulse strength measurements. It may be used in the hospital or clinical environment, during emergency air or land transport and in the home. The oximeter will operate accurately over an ambient temperature range of 32 to 104°F (0 to 40°C). The oximeter works with all BCI oximetry probes providing SpO2 and pulse rate on all patients from neonate to adult. The oximeter permits continuous patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.

## Device Story

The BCI 3304 is a portable, tabletop pulse oximeter for continuous patient monitoring. It interfaces with BCI oximetry probes to acquire SpO2 and pulse rate signals. The device processes these inputs to display real-time SpO2, pulse rate, and pulse strength via an LED interface. It features adjustable alarm limits with audible and visual alerts for system status (e.g., probe disconnect, low battery, artifact). Data communication is supported via a printer/PC port, with optional analog output and remote digital alarm adapters. Used by clinicians or patients in hospitals, clinics, homes, or during emergency transport, the device aids in clinical decision-making by providing reliable oxygen saturation data. It is an updated version of existing BCI oximeters, maintaining similar functionality in a lightweight, transportable form factor.

## Clinical Evidence

Clinical evidence consists of controlled desaturation studies comparing the 3304 to a co-oximeter (OSM-3). Results showed a standard deviation of 1.92 for the 70-100% SpO2 range and 2.31 for the 50-69% range. The R-squared value was 0.978 over the entire range, supporting accuracy claims.

## Technological Characteristics

Tabletop pulse oximeter with LED display and keypad interface. Features SpO2 probe interface, printer/PC port, optional analog output, and optional digital alarm adapter. Operates in 0-40°C range. Utilizes standard pulse oximetry sensing technology. Compliant with respiratory device reviewer guide standards for EMC, electrical, and mechanical safety.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- BCI 3301
- BCI 3302
- BCI 3100 Oximeters

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

AUG 28 1996
K962156

# Summary of Safety and Effectiveness

|  Submitter: | BCI International, Inc.
W238 N1650 Rockwood Drive
Waukesha, WI 53188  |
| --- | --- |
|  Address: |   |
|  Telephone: | (414) 542-3100  |
|  Contact: | VP Regulatory Affairs  |
|  Prepared: | May 31, 1996  |
|  Proprietary Name: | 3304 Pulse Oximeter  |
|  Common/Classification Name: | Pulse Oximeter  |
|  Predicate Devices: | BCI 3301, 3302 & 3100 Oximeters  |

## New Device Description:

The new BCI 3304 pulse oximeter is an updated version of existing pulse oximeters legally marketed by BCI International. This updated device is designed to provide full featured monitoring capabilities in a light weight, transportable design. The system consists of a small table top oximeter with a wall mount charger. The system features an $\mathrm{SpO}_2$ probe interface, display of patient data via an LED display ($\mathrm{SpO}_2$, Pulse Rate, Pulse Strength), system status LEDs (Probe, Lo Batt, Silence, Alarm, Artifact, Search), and the function keypad area consisting of eight keys (O/I, ID/Clear, Up and Down Arrows, Alarm Select, Alarm Silence, Alarm Volume, Pulse Volume). The oximeter has a printer / pc port that is used for data communication, an optional analog output and an optional digital alarm adapter for a remote location.

## Intended Use:

The 3304 provides fast, reliable $\mathrm{SpO}_2$, pulse rate and pulse strength measurements. It may be used in the hospital or clinical environment, during emergency air or land transport and in the home. The oximeter will operate accurately over an ambient temperature range of 32 to $104^{\circ}\mathrm{F}$ (0 to $40^{\circ}\mathrm{C}$). The oximeter works with all BCI oximetry probes providing SpO2 and pulse rate on all patients from neonate to adult. The oximeter permits continuous patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.

{1}

Performance Data:

The design of this device utilizes currently available technology found in many legally marketed devices. Testing was done to ensure that it would perform within the environment(s) for which it is to be marketed. The testing was performed in accordance with the guidelines and standards found in the reviewers guide for respiratory devices. This testing included EMC, electrical, mechanical durability, safety (operator and patient), and temperature/humidity. The results of the testing demonstrated that the device was in compliance with the guidelines and standards referenced in the reviewers guide and that it performed within its specifications and functional requirements.

Additionally, clinically controlled desaturation studies were done to demonstrate that the 3304 accurately displays the patient’s blood oxygen level within its accuracy limits as compared to a co-oximeter (OSM-3). Statistical analysis on the data collected from the studies were compared to those from a co-oximeter during the controlled subject desaturation runs. The results from the clinical studies support the accuracy claims of the device, with a standard deviation of 1.92 for the measurement range of 70-100% and 2.31 for the range of 50-69%. The R squared value was 0.978 over the entire range. (R squared - measure of how true the regression line is. R squared = 1 is a perfect fit.) On the basis of these results and the above-referenced testing it is our determination that the device is safe, effective, and performs as well as or better than the legally marketed predicate device(s).

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Respectfully,

![img-0.jpeg](img-0.jpeg)

Donald Alexander
VP Regulatory Affairs

---

**Source:** [https://fda.innolitics.com/device/K962156](https://fda.innolitics.com/device/K962156)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
