K961935 · Sharplan Lasers, Inc. · GEX · Nov 1, 1996 · General, Plastic Surgery
Device Facts
Record ID
K961935
Device Name
SHARPLAN LASERS, INC. SILKTOUCH CO2 FLASH SCANNER
Applicant
Sharplan Lasers, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 1, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
It is indicated for soft tissue incision of recipient sites for laser assisted hair transplantation.
Device Story
Motorized laser beam deflection system; attaches to CO2 laser articulating arm. Redirects focused laser beam via moving mirrors within specified pattern scan area. Limits tissue exposure time while maintaining power density for vaporization at low average power. Includes controls for scan area and exposure time. Used with CO2 handpiece. When off, laser functions normally for manual incision/vaporization. Operated by surgeons in clinical settings. Benefits: precise tissue vaporization for hair transplant recipient sites.
Clinical Evidence
Histological information submitted to support safety and efficacy for soft tissue incision.
Technological Characteristics
Motorized laser beam deflection system; moving mirror assembly; attaches to CO2 laser articulating arm; manual controls for scan area and exposure time.
Indications for Use
Indicated for soft tissue incision of recipient sites for laser assisted hair transplantation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Norcross Scientific SurgiScan CO2 Scanner
A-Z Surgical Super Punch
Robbins Acrotorque Hand Engine
Related Devices
K960820 — SHARPLAN LASERS, INC. SILKTOUCH XJ SCANNER · Sharplan Lasers, Inc. · May 10, 1996
K960521 — SHARPLAN LASERS, INC. SILKTOUCH FLASH SCANNER · Sharplan Lasers, Inc. · Apr 26, 1996
{0}
NOV 1 1996
K 96 1935
# 510(k) Summary of Safety and Effectiveness
## Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory
In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
The safety and effectiveness of the Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner is based on a determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices which include the Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, and the Robbins Hand Engine.
1. Sharplan Lasers, Inc.
1 Pearl Court
Allendale, NJ 07401
George J. Hattub, Director of Regulatory Affairs
May 6, 1996
1-800-394-2000, ext. 531
2. Model: Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory
3. Predicate Devices: Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, the Robbins Acrotorque Hand Engine.
4. Description: The Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory is a motorized, laser beam deflection system that is attached to a CO2 laser articulating arm. The device redirects a focused laser beam through moving mirrors, within a specified pattern scan area. Similar to the common surgeon’s technique of manually scanning a focused beam over a treatment area, the constantly moving beam of the SilkTouch Flash Scanner limits the time of laser exposure on tissue while providing sufficient power density for tissue vaporization at low average power levels. The SilkTouch Flash Scanner has controls for varying the scanned area setting the exposure time. This device is used with a CO2 handpiece. When the unit is turned off, the connected laser operates to its specifications and allows incision or manual scanning for vaporization purposes.
5. The Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory is substantially equivalent to the following devices: the Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, and the Robbins Acrotorque Hand Engine. It is indicated for soft tissue incision of recipient sites for laser assisted hair transplantation.
{1}
# 510(k) Summary of Safety and Effectiveness (continued)
Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory
6. From a clinical performance perspective, the aforementioned devices while are not identical, perform the same function for the same indication. Regarding the technological similarities; while two of the predicate devices are of the same photonic technology with minor differences, the non-laser devices relies upon a mechanical technology to achieve its clinical effect. Based upon the histological information submitted, Sharplan Lasers, Inc. believes that the expanded indications for this device raises no safety or efficacy concerns.
**Advisory:** This information was prepared for the sole purpose of compliance with the Safe Medical Devices Act of 1990. It does not imply that the procedures described herein can be performed with the equipment described without substantial risk of personal injury or death to patients due to operator error or in procedures requiring a high degree of skill.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.