The CEDIA Core TDM Multi-Cals are used to calibrate the CEDIA assays for carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid in human serum and plasma.
Device Story
CEDIA Core TDM Multi-Cals are liquid calibrators used to calibrate specific therapeutic drug monitoring (TDM) assays. Input consists of calibrator matrix containing bovine serum albumin, stabilizers, preservatives, and specific drug analytes (carbamazepine, phenobarbital, phenytoin, theophylline, valproic acid). Device functions as a reference standard to establish assay calibration curves. Used in clinical laboratory settings by laboratory technicians. Output is a set of known concentration values used to adjust assay performance. Benefits include ensuring accurate quantification of therapeutic drugs in patient serum/plasma, facilitating clinical decision-making regarding drug dosing.
Clinical Evidence
Bench testing only. Performance established via method comparison demonstrating equivalent correlation between subject calibrators and predicate kit calibrators.
Technological Characteristics
Calibrator matrix composed of bovine serum albumin, stabilizers, and preservatives spiked with carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid. Liquid form factor. Quality controlled against in-house reference calibrators value-assigned by comparison to predicate devices.
Indications for Use
Indicated for the calibration of CEDIA assays for carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid in human serum and plasma samples.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K031856 — PRECISET TDM I AND PRECISET TDM II CALIBRATORS · Roche Diagnostics Corp. · Aug 6, 2003
K033879 — VALIDATE TDM CALIBRATION VERIFICATION TEST SET, MODEL 126 · Maine Standards Co. · Feb 27, 2004
K032574 — DIMENSION DRUG CALIBRATOR II, MODEL DC49C · Dade Behring, Inc. · Oct 6, 2003
Submission Summary (Full Text)
{0}
K96/659
Diagnostics
# 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
| --- | --- |
| 1. Submitter name, address, contact | Boehringer Mannheim Corporation
2400 Bisso Lane
Concord, CA 94524-4117
(510) 674-0690 extension 8415
Contact Person: Mary Koning
Date Prepared: April 26, 1996 |
| 2. Device name | Proprietary name: CEDIA Core TDM Multi-Cals
Common name: Therapeutic drug monitoring calibrators for use in the calibration of the CEDIA assays for carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid.
Classification name: Calibrators, drug mixture |
| 3. Predicate device | The Boehringer Mannheim CEDIA Core TDM Multi-Cals are substantially equivalent to the kit calibrators included in CEDIA Carbamazepine (K914857), Phenobarbital (K904413), Phenytoin (K905689), Theophylline (K961462), and Valproic Acid (930734/2) Assays.. |
| 4. Device Description | The CEDIA Core TDM Calibrators are manufactured using bovine serum albumin, carbamazepine, phenobarbital, phenytoin, theophylline, valproic acid, stabilizers, and preservatives. The drugs are appropriately spiked into the calibrator matrix to the correct calibrator concentration levels. The calibrators are in process checked and quality controlled against in-house reference calibrators (prepared using a similar procedure) which have been value assigned by comparison to the predicate device to ensure correct assay calibration. |
Continued on next page
page 21
40
{1}
Diagnostics
# 510(k) Summary, Continued
## 5. Intended use
The CEDIA Core TDM Multi-Cals are used to calibrate the CEDIA assays for carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid in human serum and plasma.
## 6. Comparison to predicate device
The Boehringer Mannheim CEDIA Core TDM Multi-Cals are substantially equivalent to the CEDIA Carbamazepine (K914857), Phenobarbital (K904413), Phenytoin (K905689), Theophylline (K961462), Valproic Acid (930734/2) Assays.
The following table compares the CEDIA Core TDM Multi-Cals with the predicate devices, CEDIA Carbamazepine, Phenobarbital, Phenytoin, Theophylline, and Valproic Acid assay’s kit calibrators. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate device is provided in attachment 6.
## Similarities:
- Similar formulation.
- Similar concentrations of carbamazepine, phenobarbital, phenytoin, theophylline, and valproic acid.
- For the calibration of the CEDIA Carbamazepine, Phenobarbital, Phenytoin, Theophylline, and Valproic Acid assays.
## Differences:
| Feature | Core TDM Multi-Cals | Kit Calibrators |
| --- | --- | --- |
| Configuration | sold separately from reagents | sold with reagents in kit |
Continued on next page
page 22
{2}
Diagnostics
# 510(k) Summary, Continued
## 6. Comparison to predicate device, (cont.)
### Performance Characteristics:
| Feature | Core TDM Multi-Cals | Kit Calibrators |
| --- | --- | --- |
| Open Vial Stability | Stable for 30 days | Stable for 30 days |
- Method Comparison: equivalent correlation between kit calibrators and CEDIA Core TDM Multi-Cals.
page 23
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.