The Ciba Corning 400 System is intended for the point-of-care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.
Device Story
Ciba Corning 400 System is a point-of-care and laboratory analyzer for whole blood samples. It utilizes electrochemical sensors to measure pH, blood gases (pCO2, pO2), electrolytes (Na+, K+, Ca++, Cl-), glucose, and hematocrit. The device employs thick-film hybrid, solid-state planar sensors in a disposable cartridge, replacing traditional internal-fill electrodes. Input is 75 uL of whole blood; sensors convert electrochemical potential (potentiometry, amperometry, conductimetry) into electrical signals. The system processes these signals to output analyte concentrations. Used by clinicians in laboratory or point-of-care settings to provide rapid diagnostic data for clinical decision-making.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Electrochemical sensing system using thick-film hybrid, solid-state planar sensors. Measurement methods: potentiometry (pH, Na+, K+, Ca++, Cl-, pCO2), amperometry (pO2, glucose via glucose oxidase), and conductimetry (hematocrit). Reference electrode uses silver/silver chloride in potassium chloride. Designed for use with compact disposable measurement cartridges. Requires 75 uL sample volume. Maintenance-free sensor design.
Indications for Use
Indicated for point-of-care and laboratory testing of arterial, venous, and capillary whole blood samples to measure pH, pCO2, pO2, sodium, potassium, ionized calcium, chloride, glucose, and hematocrit.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
K962021 — CIBA CORNING 348 SYSTEM · Ciba Corning Diagnostics Corp. · Jun 24, 1996
K080370 — ABL 80 CO-OX SYSTEM · Sendx Medical, Inc. · Jul 31, 2008
Submission Summary (Full Text)
{0}
Company Confidential
K961657
# Summary of Safety and Effectiveness
JUN 25 1996
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
## 1. Submitters Information
Contact person: William J. Pignato
Director of Regulatory Affairs
Address: Ciba Corning Diagnostics Corp.
63 North Street
Medfield, MA 02052
Phone: 508 359-3825
Date Summary Prepared: April 29, 1996
## 2. Device Information
| Proprietary Name: | Ciba Corning 400 System |
| --- | --- |
| Common Name: | Analyzer for Blood Gas, electrolytes and metabolites |
| Classification Name: | Electrode/Sensor measurement of blood gases, blood pH and blood electrolytes |
| Classification Number: | Calcium - 21 CFR 862.1145, Class II |
| | Chloride - 21 CFR 862.1170, Class II |
| | Glucose - 21 CFR 862.1345, Class II |
| | Hematocrit - 21 CFR 864.6400, Class II |
| | Potassium - 21 CFR 862.1600, Class II |
| | Sodium - 21 CFR 862.1665, Class II |
| | pCO2 - 21 CFR 862.1120, Class II |
| | pO2 - 21 CFR 862.1120, Class II |
| | pH - 21 CFR 862.1120, Class II |
## 1. Predicate Device Information
Name: Model 860 Analyzer
Manufacturer: Ciba Corning Diagnostics Corp.
510(k) Number: D.C. # K934907
## 2. Device Description
The 400 Series system analyzer is a point of care and laboratory testing analyzer used to for the direct measure of whole blood samples for the determination of the following parameters:
Ciba Corning 400 System
4/18/96
Page 2
{1}
Company Confidential
- partial pressures of carbon dioxide; $p\mathrm{CO}_{2}$
- partial pressure of oxygen $pO_{2}$
- pH
- sodium; $\mathrm{Na}^{+}$
- potassium; $\mathrm{K}^{+}$
- ionized calcium; $\mathrm{Ca}^{++}$
- chloride; $\mathrm{Cl}^{-}$
- glucose
- hematocrit; Hct
## 3. Statement of Intended Use
The Ciba Corning 400 System is intended for the point-of-care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.
## 4. Summary of Technological Characteristics
The 400 Series System uses measurement technology that is based on electrochemical phenomena. The device use potentiometry, amperometry and conductimetric methods to convert the potential generated by the sensor to an electrical signal which the system then converts to a value that represents that concentration of a specific analyte in the whole blood sample.
The 400 Series sensors (i.e., electrodes) provide direct measurement of the specific analytes or substances in the sample. Each sensor in the 400 system is highly selective for one substance over others.
The sensors employ the use of thick-film hybrid technology and a solid-state design in place of the traditional electrodes that use internal fill solutions that are used in the traditional blood gas/electrolyte systems.
Features the planar-format sensors include:
- Suitability for use with the compact disposable measurement cartridge used with the 400 Series
- Sensors require a small volume of sample (75 uL)
- Sensors are maintenance free
Ciba Corning 400 System 4/18/96 Page 3
{2}
Company Confidential
The sensors use the following measurement technology:
| Sensor | Measurement Technology |
| --- | --- |
| | |
| pH, Na+, K+, Ca++, Cl- | potentiometric method using ion-selective electrode technology |
| reference | silver electrode in potassium chloride and silver chloride |
| pCO2 | potentiometric method |
| pO2 | amperometric method |
| glucose | amperometric method using an enzyme electrode that uses glucose oxidase |
| hematocrit | conductimetric method |
Ciba Coming 400 System
4/18/96
Page 4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.