ACUFEX TIBIAL ANCHOR SCREW & SPIKED WASHER

K961649 · Smith & Nephew Endoscopy, Inc. · HWC · Jul 16, 1996 · Orthopedic

Device Facts

Record IDK961649
Device NameACUFEX TIBIAL ANCHOR SCREW & SPIKED WASHER
ApplicantSmith & Nephew Endoscopy, Inc.
Product CodeHWC · Orthopedic
Decision DateJul 16, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Acufex® Tibial Anchor Screw & Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures.

Device Story

Device consists of two-part tibial anchor screw and spiked washer; used for soft tissue-to-bone fixation in orthopedic procedures. Outer screw (9mm diameter, 20-30mm length) features cancellous threads for bone contact; inner screw (7mm diameter, 20-25mm length) secures spiked washer (14-20mm diameter) to outer screw. Spikes at washer periphery provide tissue grip. Device implanted by orthopedic surgeons in clinical/OR settings to secure tendons/ligaments. Provides mechanical fixation to facilitate healing. Materials include titanium and reinforced polyacetal polymer.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials: Titanium and reinforced polyacetal polymer. Design: Two-part screw system (outer 9mm, inner 7mm) with spiked washer (14-20mm). Mechanical fixation principle. Sterile, single-use.

Indications for Use

Indicated for fixation of soft tissue, including tendons and ligaments, to bone during orthopedic surgical procedures.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0} JUL 16 1996 K 96/649 # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is : This summary was prepared on April 26, 1996 ## A. Submitter Smith & Nephew Endoscopy 130 Forbes Boulevard Mansfield, MA 02048 ## B. Company Contact Tedd Gosian Clinical Regulatory Manager ## C. Device Name Trade Name: Acufex® Tibial Anchor Screw & Spiked Washer Common Name: Fixation Screw & Washer Classification Name: Single/Multiple component metallic bone fixation appliances and accessories ## D. Predicate/Legally Marketed Devices Concept® Soft Tissue Anchoring Implants (K871037) Synthes® Spiked Washer ## E. Device Description Acufex® Tibial Anchor Screw & Spiked Washer is a device used in soft tissue to bone fixation. The Tibial Anchor Screw is a two part screw. The outer larger part has cancellous threads, which come in direct contact with bone, and is 9 mm in diameter with a length of 20 - 30 mm. The inner part, 7 mm diameter and 20 - 25 mm length, is used to attach the Spiked Washer directly to the outer part. The spiked washer is 14 - 20 mm in diameter with spikes set at the periphery. ## F. Intended Use The Acufex® Tibial Anchor Screw & Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures. {1} 104 # G. Substantial Equivalence The AcufeX® Tibial Anchor Screw & Spiked Washer is intended for use for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures. The AcufeX® Tibial Anchor Screw & Spiked Washer is similar in design, function, materials and intended use to other devices currently marketed and in commercial distribution, namely Concept® Soft Tissue Anchoring Implants (K871037) and Synthes® Spiked Washer. Risks to health have been addressed through the specified materials, processing controls, quality assurance, and compliance to the Medical Device Good Manufacturing Practices regulations. A summary comparison of the characteristics of the AcufeX® Tibial Anchor Screw & Spiked Washer and the substantially equivalent device is presented in the table below. | | Current Product | Substantially Equivalent Product | | | --- | --- | --- | --- | | Attribute I | AcufeX® Tibial Anchor Screw & Spiked Washer | Concept® Soft Tissue Anchoring Implants | Synthes® Spiked Washer | | Indication | Soft Tissue Fixation | Soft Tissue Fixation | Ligament Fixation | | Length | Outer: 20 - 30 mm Inner: 18 - 25 | Screw: 25 - 55 mm | | | Diameter | Outer: 9 mm. Inner: 7 mm Washer: 14 - 20 mm O.D. | Screw: 6.5 mm - Washer: 14 to 20 mm O.D. | Washer: 13.5 mm O.D. | | Materials | Screw: Titanium Washer: Titanium Reinforced Polyacetal Polymer | Screw: Titanium Washer: Titanium | Washer: Stainless Steel reinforced polyacetal | | Labeling | Sterile, Single Use Only | Sterile, Single Use Only | Sterile, Single Use | Applicant _Vicki R. Brown_ Date _4/26/96_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...