SMART-TENS, MODEL 456

K961376 · Bio-Medical Research, Ltd. · GZJ · Sep 12, 1996 · Neurology

Device Facts

Record IDK961376
Device NameSMART-TENS, MODEL 456
ApplicantBio-Medical Research, Ltd.
Product CodeGZJ · Neurology
Decision DateSep 12, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Smart-TENS is used for transcutaneous electrical nerve stimulation for the purposes of relief of chronic intractable pain.

Device Story

Smart-TENS is a compact, battery-powered transcutaneous electrical nerve stimulator (TENS) for chronic intractable pain relief. Device features two independent channels delivering electrical pulses via skin surface electrodes. Clinician selects prescribed programs; patient controls include amplitude intensity, burst mode initiation, sequence repetition, sequence advancement, and treatment pausing. Device includes internal compliance recording for clinician review. Operation involves impedance-based output transformer to achieve near net zero charge delivery. Used in clinical or home settings; operated by patients under clinician-prescribed parameters.

Clinical Evidence

No clinical testing was performed. Safety and effectiveness supported by bench testing of stimulation output parameters and adherence to EN60601-1 and EN60601-1-2 standards.

Technological Characteristics

Portable, battery-powered TENS device; two independent channels; output transformer-based impedance control for near net zero charge delivery; conforms to EN60601-1 (IEC 601-1) and EN60601-1-2 (IEC 601-1-2) standards.

Indications for Use

Indicated for the relief of chronic intractable pain in patients requiring transcutaneous electrical nerve stimulation therapy.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K961376 # 510(k) Notification for Smart-TENS: Appendix IX SEP 12 1996 # 510(k) Summary of Safety and Effectiveness This summary is submitted in compliance with the FDA interim rule 21 CFR 807.92. (a) (1) Submitted by: Bio-Medical Research Ltd., BMR House, 16 Merchants Road, Galway. Republic of Ireland Telephone: +353 91 566911 Fax:: +353 91 566907 Contact Person: Isaac Finnegan Position/ Title: Design Engineer - Regulatory Affairs Date of preparation: 4 April, 1996 (2) Trade name of device: Smart-TENS, Model 456 Common name: TENS Classification name: Transcutaneous Electrical Nerve Stimulator for pain relief; §882.5890 (3) Identification of predicate or legally marketed device: Bio Tens Model: ST-601 (Skylark Device Co. Ltd., 510(k) # K912178) Page 1 of 3 {1} # 510(k) Notification for Smart-TENS: Appendix IX ## (4) Description of device: The Smart-TENS is a compact, battery powered transcutaneous electrical nerve stimulator. It is used for the relief of chronic intractable pain by providing transcutaneous electrical pulses to areas of the body that require therapy for the indicated medical conditions. The parameters of the programs in the Smart-TENS cannot be set by the user. Two channels are available on the Smart-TENS. Each channel operates independently and if desired they may be used simultaneously. The clinician may set the prescribed program from a list of defined programs. The patient may then use this program or another available program called MultiTens which provides a pattern of sequences which allows variation and modulation on the stimulation pulse. The patient may also initiate burst mode on certain programs, cause one of the pattern of sequences in MultiTens Mode to repeat continuously, advance to the next sequence or pause the treatment session, increase or decrease the amplitude intensity of the stimulation or lock the intensity controls. Use of the device is recorded internally, allowing the clinician to determine compliance with the prescribed treatment regimen. ## (5) Intended uses: The Smart-TENS is used for transcutaneous electrical nerve stimulation for the purposes of relief of chronic intractable pain. These uses are similar to the predicate marketed device identified in section (3) of this summary. ## (6) Technological comparison The Smart-TENS is similar to the Bio Tens Model: ST-601 in that both are portable, compact battery powered TENS devices. Output for either device is suitable for the relief of chronic pain using standard skin surface electrodes. Both devices are similar in basic operational design and use the impedance of the output transformer to achieve a near net zero charge into the skin. The Smart-TENS is effectively a more user controllable device which indicates patient compliance. {2} # 510(k) Notification for Smart-TENS: Appendix IX (b) (1) Non-clinical tests: Comparisons of stimulation outputs for the Smart-TENS and the predicate Bio Tens Model: ST-601 show similar results that are suitable for TENS. To minimize potential electrical and mechanical hazards, Bio-Medical Research adheres to recognised and established industry practice and all devices are subject to final performance testing. The Smart-TENS is designed to conform to EN60601-1 (IEC 601-1) and conforms to EN 60601-1-2 (IEC 601-1-2). (2) Clinical tests: No clinical testing was performed. (3) Test conclusions: Testing of the stimulation output parameters of the Smart-TENS indicate that the device is safe, that it provides appropriate stimulation output for effective relief of chronic pain and that it performs as well as or better than the legally marketed predicate device identified in section (3) of this Summary. Page 3 of 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...