PHILIPS MULTIDIAGNOST 4

K961374 · Philips Medical Systems, Inc. · OWB · Aug 19, 1996 · Radiology

Device Facts

Record IDK961374
Device NamePHILIPS MULTIDIAGNOST 4
ApplicantPhilips Medical Systems, Inc.
Product CodeOWB · Radiology
Decision DateAug 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2

Indications for Use

The MultiDIAGNOST 4 is indicated for use in Radiographic/Fluoroscopic, Angiographic, and Interventional diagnostic imaging examinations.

Device Story

MultiDIAGNOST 4 is a universal tilt C-arm system for diagnostic imaging. It captures X-ray images for radiographic, fluoroscopic, angiographic, and interventional procedures. Operated by clinicians in clinical settings, the system provides visualization to assist in diagnostic decision-making. It functions as an image-intensified fluoroscopic X-ray system.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, safety standards, and historical performance of similar predicate devices.

Technological Characteristics

Universal tilt C-arm X-ray system; image-intensified fluoroscopy; complies with 21 CFR Part 1000 and UL 187 safety standards.

Indications for Use

Indicated for patients requiring radiographic, fluoroscopic, angiographic, or interventional diagnostic imaging examinations.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Submission Summary (Full Text)

{0} PHILIPS K96/374 Philips Medical Systems P.O. Box 10000, 5680 DA Best, The Netherlands # PHILIPS Department of Health and Human Services Center for Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section Qual. Ass. Dpt. XSB/XCB XB030-960317/RR/gd 1996.03.01 # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION for # PHILIPS MULTI DIAGNOST 4, UNIVERSAL TILT C-ARM SYSTEM This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. The undersigned certifies that the 510(k) Pre-Market notification for the above referenced product contains adequate information and data to enable CDRH to determine substantial equivalence. This information and data is summarized as follows: 1. The Multi Diagnost 4 system subject to Federal Performance Standards, defined in 21CFR - part 1000; 2. The Multi Diagnost 4 system will be manufactured in accordance with voluntary safety standards, such as UL 187; 3. The information for Users contains comprehensive information to insure safe and effective use; 4. Past experience with substantially equivalent predicate devices has shown our device to be safe and effective when used as directed in the Information for Users. Ing. R.W. Rijntjes Approbation officer Quality Assurance dept. XSB / XCB Philips Medical Systems Nederland BV Best, The Netherlands. ![img-0.jpeg](img-0.jpeg) Philips Medical Systems Nederland B.V. Vaanpluis 6, Best, The Netherlands Tel: +31 40 27 911 11 Fax: +31 40 27 622 05 Commercial Register Eindhoven {1} HUMAN SERVICES, INC. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Mr. Peter Altman Director of Regulatory Affairs Philips Medical Systems, Inc. North America Company 710 Bridgeport Avenue SHELTON CT 06484 FEB 19 2013 Re: K961374 Trade/Device Name: PHILIPS MultiDIAGNOST 4 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: May 22, 1996 Received: May 23, 1996 Dear Mr. Altman: This letter corrects our substantially equivalent letter of August 19, 1996. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {2} Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of *In Vitro* Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, ![img-1.jpeg](img-1.jpeg) Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number (if known): Unknown Device Name: Philips MultiDIAGNOST 4 Indications For Use: The MultiDIAGNOST 4 is indicated for use in Radiographic/Fluoroscopic, Angiographic, and Interventional diagnostic imaging examinations. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David L. Segura (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K961374 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-4) 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...