MAMMARY SIZER

K961356 · General Surgical Innovations · MRD · Sep 23, 1996 · SU

Device Facts

Record IDK961356
Device NameMAMMARY SIZER
ApplicantGeneral Surgical Innovations
Product CodeMRD · SU
Decision DateSep 23, 1996
DecisionSN
Submission TypeTraditional
Device ClassClass U

Indications for Use

The GSI, Inc. Mammary Sizer is intended to assist in evaluation of the appropriate mammary prosthesis volume for each patient prior to implantation.

Device Story

Device: Mammary Sizer; used during plastic/reconstructive surgery. Input: saline/fluid via filler tube. Operation: balloon inflated to evaluate mammary prosthesis volume/symmetry in sub-glandular or sub-muscular tissue planes. Output: physical volume expansion for visual/tactile assessment by surgeon. Benefit: assists in selecting appropriate implant size/symmetry prior to permanent implantation.

Clinical Evidence

Bench testing only; includes biocompatibility, material strength, and volume rating testing.

Technological Characteristics

Material: smooth semi-elastomeric polyurethane. Components: balloon, balloon tubing, fluid leakage prevention attachment, female luer lock. Form factor: inflatable balloon sizer.

Indications for Use

Indicated for patients undergoing plastic and reconstructive surgery requiring determination of mammary implant volume and symmetry prior to permanent implantation.

Predicate Devices

Submission Summary (Full Text)

{0} SEP 23 1996 K961356 General Surgical Innovations, Inc. 3172A Porter Dr. Palo Alto, CA. 94304 SMDA Summary Attachment B 510(K) SUMMARY Document Mail Center (HFZ-401) Center for Devices and Radiological Health Food and Drug Administration 9200 Corporate Boulevard Rockville, Maryland 20850 USA Summary of Safety and Effectiveness **Contact Person** Theresa M. Nazaroff General Surgical Innovations, Inc. 3172A Porter Dr. Palo Alto, CA 94304 **Device Name** Mammary Sizer **Substantial Equivalence and Conclusions** The Mammary Sizer which is proposed for use in the sub-glandular of sub-muscular tissue plane is substantially equivalent to Mentor's Volume Sizer for Mammary Implants that are marketed for use in the area of plastic and reconstructive surgery. **Intended Use** The GSI, Inc. Mammary Sizer is intended to assist in evaluation of the appropriate mammary prosthesis volume for each patient prior to implantation. {1} k961306 Page 2 of 2 ## Device Description The GSI Mammary Sizer is designed to be used during plastic and reconstructive surgery when it is desirable to determine the correct volume and/or symmetry of mammary implants prior to implantation. The device is constructed of smooth semi-elastomeric polyurethane. The device consists of a balloon and balloon tubing used for filling the sizer. There is an attachment to prevent fluid leakage and a female luer lock at the proximal end of the filler tube. ## Testing Results and Conclusions Testing performed to support the appropriateness of the Mammary Sizer for its intended use include biocompatibility, material strength, and volume rating. Theresa M. Nazaroff General Surgical Innovations, Inc. March 30, 1996
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...