K961025 · Medicinelodge, Inc. · GAT · Apr 15, 1996 · General, Plastic Surgery
Device Facts
Record ID
K961025
Device Name
EXPANDED BODY SUTURE
Applicant
Medicinelodge, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Apr 15, 1996
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EXPANDED BODY suture is indicated for use in soft-tissue approximation and/or ligation in rotator cuff repair procedures.
Device Story
EXPANDED BODY suture is a braided, nonabsorbable, sterile, poly(ethylene terephthalate) surgical suture. Device features an expanded midsection (15-25 mm length; 2-4x standard USP diameter) designed to firmly engage drill/needle holes in bone tissue during rotator cuff repair. Increased midsection surface area distributes force more evenly to surrounding bone, potentially increasing pull-out strength compared to standard sutures. Used by surgeons in orthopedic procedures. Suture is provided in USP sizes #0-#5, with or without attached 420C stainless steel needles. Device is sterilized via gamma radiation.
Clinical Evidence
Bench testing only. Comparative testing performed using USP #5 EXPANDED BODY suture versus USP #5 TI•CRON predicate in Last-a-Foam polyurethane foam (simulated bone). Testing verified conformance to USP #871 (Needle Attachment) and USP #881 (Knot Pull) standards. Results confirmed the device meets or exceeds USP requirements and performs equivalently to the predicate in simulated pull-out applications.
Technological Characteristics
Materials: Poly(ethylene terephthalate) suture; 420C Stainless Steel (ASTM F-899) needles. Colorants: D&C Blue #6, D&C Green #6. Design: Braided, nonabsorbable, with expanded midsection (15-25 mm length, 2-4x standard diameter). Sterilization: Gamma radiation. Sizes: USP #0-#5. Connectivity: None.
Indications for Use
Indicated for soft-tissue approximation and/or ligation in rotator cuff repair procedures. No known contraindications.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}
04/11/96 15:41 9166465432 HAYES MEDICAL --- CDRH ODE 006
K961025
APR 15 1996
# Premarket Notification
510(k) Summary of Safety and Effectiveness Information
## DATE OF SUMMARY PREPARATION
March 5, 1996
## MANUFACTURER
MedicineLodge, Incorporated
152 South 600 West
Logan, Utah 84321
Telephone: (801) 753-7675
Fax: (801) 753-7698
## MANUFACTURER CONTACT PERSON
Alan Chervitz, President
MedicineLodge, Inc.
152 South 600 West
Logan, Utah 84321
Telephone: (801) 753-7675
Fax: (801) 753-7698
## REGULATORY CORRESPONDENCE
Correspondence related to this submission may be addressed to:
Brian T. Cleary
Director of RA/QA
Hayes Medical, Inc.
819 Striker Avenue, Suite 10
Sacramento, CA 95834-5432
Telephone: (916) 646-5441
Fax: (916) 646-5432
## PROPRIETARY DEVICE NAME
EXPANDED BODY Suture
## COMMON NAME
Braided Nonabsorbable Polyester Surgical Suture
27
{1}
04/11/85 0:40 9166465432
HAYES MEDICAL --- CDRH ODE
007
## CLASSIFICATION NAME
Nonabsorbable Poly(ethylene terephthalate) Surgical Suture
## CLASSIFICATION REFERENCE
21 CFR § 878.5000
## DEVICE PRODUCT CODE
79 GAT
## REGULATORY CLASS
In accordance with FDA classification of nonabsorbable synthetic poly(ethylene terephthalate) sutures as Class II medical devices, this device is proposed for placement in Class II.
## SPECIAL CONTROLS
At this time, Food and Drug Administration generated Performance Standards applicable to the EXPANDED BODY suture are not in force. In accordance with 21 CFR § 878.5000, testing has been performed confirming that the EXPANDED BODY suture meets applicable United States Pharmacopoea (USP) requirements as described in the USP Monograph for Nonabsorbable Surgical Suture, with the exception of suture diameter.
## USP COMPLIANCE TESTING
As required by 21 CFR § 878.5000, device testing has been conducted to verify that the EXPANDED BODY suture meets all applicable USP standards for nonabsorbable surgical sutures, with the exception of suture diameter. MedicineLodge, Inc. has conducted the following tests on a representative USP #5 EXPANDED BODY suture to verify conformance with the USP standards listed below.
1) Suture Needle Attachment Testing (USP # 871)
2) Suture Tensile (Knot Pull) Test (USP # 881)
Results from these tests confirm that the USP #5 EXPANDED BODY suture exceeds applicable requirements for USP tests #871 and #881 for nonabsorbable surgical sutures. MedicineLodge, Inc. believes that the results of this testing are representative of the performance of all USP sizes of this device and demonstrate that all USP size EXPANDED BODY sutures will exceed these requirements.
## MATERIALS
The following materials are used in the manufacture of the EXPANDED BODY suture.
28
{2}
04/11/96 15:41 9166465432
HAYES MEDICAL --- CDRH ODE
008
Poly(ethylene terephthalate): in accordance with 21 CFR § 878.5000, the EXPANDED BODY suture is constructed of poly(ethylene terephthalate). This material is similarly used in the manufacture of the all predicate devices and its clinical use is therefore well established. The expanded midsection of the device is similarly comprised of this material.
Color Additives: the EXPANDED BODY suture will be colored using the following coloring additives:
- D&C Blue #6
- D&C Green #6
- White (Standard/Undyed)
D&C Blue #6 and D&C Green #6 are similarly used to dye all predicate devices. FDA has indicated that these colors are approved for the dying of polyester sutures.
Stainless Steel: the needle portion of the EXPANDED BODY suture is manufactured from 420C Stainless Steel (ASTM F-899).
## INDICATIONS FOR USE
The EXPANDED BODY suture is indicated for use in soft-tissue approximation and/or ligation in rotator cuff repair procedures.
## DEVICE DESCRIPTION
The EXPANDED BODY suture is a braided, nonabsorbable, sterile, surgical suture composed of Poly(ethylene terephthalate) intended for use in rotator cuff soft-tissue repair procedures. This device is prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component.
The EXPANDED BODY suture is identical to currently marketed polyester sutures within this device class with the exception of the following:
1) a new labeled indication for use stating specific use in rotator cuff soft tissue repair procedures, and
2) a new design which incorporates an expanded midsection equidistant from both ends of the suture possessing a larger outer diameter (two to four times) than the standard USP size.
The outer diameter of the expanded midsection thus has a cross-sectional area equal to or greater than the cross-sectional area of the needle or drill used to create the hole in the tissue. This ensures that the expanded midsection firmly engages the hole created by the drill or needle in the tissue. The rationale behind this new design is discussed in detail on the following page.
29
{3}
04/11/96 15:42 9166465432
HAYES MEDICAL --- CDRH ODE
009
MedicineLodge, Inc. plans to provide the EXPANDED BODY suture in USP sizes #0 through #5 (metric sizes 3.5 through 7). All sutures will be 30 inches in length, available in 5 different colors (undyed white, dyed green, dyed blue, and white/blue and white/green combinations), and provided with and without attached needles.
For all EXPANDED BODY sutures, the length of the expanded midsection of the device will range from 15 mm to 25 mm and the diameter of the expanded midsection will range from two to four times the diameter of the corresponding USP size of the suture. All device labeling will specify exact expanded midsection dimensions for each USP size.
## SUBSTANTIALLY EQUIVALENT PREDICATE DEVICES
MedicineLodge, Inc. believes that several commercially available devices are substantially equivalent to the EXPANDED BODY suture. Within the proposed class, the following devices were used as predicate devices for comparison in establishing the substantial equivalence of the EXPANDED BODY suture:
1) TI•CRON - Nonabsorbable Polyester Surgical Suture
Davis + Geck, Manati, PR; Danbury, CT
2) POLYVIOLENE - Nonabsorbable Polyester Surgical Suture
Look, Inc., Norwell, MA
## SUBSTANTIAL EQUIVALENCE COMPARISON
To begin, the materials, method of manufacture, method of sterilization, and device packaging are identical to predicate devices. Additionally, as with all predicate devices, the EXPANDED BODY suture meets all USP requirements applicable to nonabsorbable sutures, with the exception of suture diameter.
As indicated previously, the only differences that distinguish the EXPANDED BODY suture from the selected predicate devices are:
1) a new labeled indication for use stating specific use in rotator cuff soft tissue repair procedures, and
2) a new design which incorporates an expanded midsection possessing a larger outer diameter (two to four times) than the standard USP size.
Devices that have previously received 510(k) clearance within 21 CFR § 878.5000 are specifically indicated for use in all types of soft tissue repair, including specific stated use in cardiovascular, ophthalmic and neurological procedures. Unlike the proposed labeling for the EXPANDED BODY suture, the predicate devices do not specify a labeled indication for use in orthopedic soft-tissue repair.
However, it should be mentioned that all predicate device labeling states no known contraindications. It is our understanding that suture devices are used regularly in orthopedic soft tissue approximation and/or ligation.
30
{4}
04/11/96 15:42 9166465432
HAYES MEDICAL --- CDRH ODE
010
Medical literature documents the widespread use of nonabsorbable polyester suture devices for soft-tissue repair of the shoulder, including rotator cuff repair. Hence, although the labeled indication for use of the EXPANDED BODY suture is new, the practice of using nonabsorbable, polyester suture devices for the same indication is well documented in medical literature.
## Design Difference - Expanded Midsection
The rationale behind the expanded midsection design is to increase the outer diameter of the suture to more firmly engage and fit the hole created by the needle or drill in the bone tissue during suture insertion. This increased outer diameter also results in an increased suture surface area contained in the formed hole.
This larger surface area potentially reduces the chances of suture or bone failure by more evenly distributing any applied force to the surrounding bone tissue. The expanded midsection feature more evenly distributes the applied force over the bone surface, the EXPANDED BODY suture potentially increases the maximum pull-out strength of the suture relative to standard, similarly sized polyester sutures.
A representative USP #5 EXPANDED BODY design configuration was used for all comparative testing. This design combines dimensions for the expanded midsection that fall at the lower end of the design limits for a USP #5 EXPANDED BODY suture - lower-end design limit length (16.5 mm) and less than minimum outer diameter (1.27 mm < 2x = 1.499 mm). The predicate device used for comparison was a USP #5 TI•CRON (Davis + Geck) nonabsorbable, polyester suture.
Comparative testing was performed using Last-a-Foam Polyurethane Foam as a simulated bone material. Use of this material for comparative device testing assures that the EXPANDED BODY and TI•CRON suture samples are tested in a uniform material so that variation of material properties is controlled. In this way, the results of the test measure only the difference in performance of the two suture types.
The test results confirm that the EXPANDED BODY suture performs equivalent to the TI•CRON suture in this simulated application.
## Conclusion
Based on the design concept, use of standard materials, feature comparisons to the selected predicate devices, and device testing, MedicineLodge, Inc. believes that sufficient evidence exists to conclude that the EXPANDED BODY suture is substantially equivalent to existing legally marketed nonabsorbable polyester suture devices.
Additionally, the information presented herein should support our request for labeled "indications for use" in soft-tissue approximation and/or ligation in rotator cuff repair procedures.
Table 1 on the following page summarizes the relevant feature comparisons between the EXPANDED BODY suture and the selected predicate devices.
31
{5}
04/11/96 15:43 9166465432 HAYES MEDICAL --- CDRH ODE 011
# TABLE 1
Feature comparisons between the EXPANDED BODY suture and Ti•Cron (Davis + Geck) and Polyviolene (Look, Inc.) predicate devices (differences are indicated in **bold** print)
| | EXPANDED BODY
Polyester Nonabsorbable
Surgical Suture, USP | TI•CRON
Polyester Nonabsorbable
Surgical Suture, USP | POLYVIOLENE
Polyester Nonabsorbable
Surgical Suture, USP |
| --- | --- | --- | --- |
| *Indications* | Soft-tissue approximation
and/or ligation in rotator
cuff repair procedures. | General soft tissue
approximation and/or
ligation, including use in
cardiovascular, ophthalmic
and neurological procedures. | General soft tissue
approximation and/or
ligation, including use in
cardiovascular, ophthalmic
and neurological procedures. |
| *Contraindications* | None known | None known | None known |
| *Materials* | Poly(ethylene terephthalate) | Poly(ethylene terephthalate) | Poly(ethylene terephthalate) |
| *Expanded
Midsection* | Length: 15 mm to 25 mm
Outer Diameter: 2 to 4
times the corresponding
USP size | No | No |
| *Sterilization* | Gamma | Gamma | Gamma |
| *Pledgets* | No | Yes (PTFE) | Yes (PTFE) |
| *Needles* | With and without | With and without | With and without |
| *Sizes* | USP 0 through 5 | USP #7-0 through #5 | USP #11-0 through #5 |
| *Colors* | Blue, Green, White;
Blue/White and Green/White
combinations | Blue and White | Green and White |
| *Manufacturer* | MedicineLodge, Inc. | Davis + Geck | Look, Inc. |
| *510(k) Approved* | N/A | Yes | Yes |
32
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.