AUTOMIX 3+3/AS COMPOUNDER SYSTEM

K961008 · Baxter Healthcare Corp · LHI · Oct 10, 1997 · General Hospital

Device Facts

Record IDK961008
Device NameAUTOMIX 3+3/AS COMPOUNDER SYSTEM
ApplicantBaxter Healthcare Corp
Product CodeLHI · General Hospital
Decision DateOct 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

This device is an electromechanical positive displacement fluid compounding device that provides for the compounding of fluids via gravimetric weighing.

Device Story

Electromechanical positive displacement fluid compounding system; utilizes gravimetric weighing for accurate fluid transfer. Modification includes 'Positive Identification Solution Family Monitor' to verify source solution identity and placement. Used in clinical settings for compounding fluids; operated by trained personnel. System provides automated verification of source solutions to ensure correct formulation. Benefits include increased safety, speed, and accuracy in fluid compounding processes.

Clinical Evidence

No clinical data. Functional acceptability of hardware and software modifications demonstrated via bench testing, verification, and validation procedures.

Technological Characteristics

Electromechanical positive displacement compounding system; gravimetric weighing mechanism; integrated computer software for solution identification; Class II device (21 CFR 880.5440).

Indications for Use

Indicated for the compounding of fluids via gravimetric weighing using an electromechanical positive displacement system. Prescription use only.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1. 1008 510(k) Premarketing Notification Clintec Automix® 3 + 3 / AS Compounder System March 11, 1996 OCT 10 1997 Attachment 9. # 510(k) Summary (as required by 21 CFR 807.92) Submitted by: Dennis A. Ocwieja Vice President Regulatory Affairs and Quality Assurance Clintec Nutrition Company Three Parkway North, Suite 200 Deerfield, IL 60015 phone: (847) 317-3058 fax: (847) 317-3139 Date Prepared: 03/08/96 Trade/Proprietary Name: Clintec Automix® 3 + 3 / AS Compounder System Common/Usual Name: Electromechanical Positive Displacement Fluid Compounding System Classification Name: Compounding System with Integrated Computer Software Classification: Class II in 21 CFR § 880.5440, Set, I.V., Fluid Transfer Predicate Device: Clintec Automix® 3 + 3 Compounder System The Automix® 3 + 3 / AS Compounder System is an electromechanical positive displacement fluid compounding device that provides for the safe, fast, and accurate compounding of a wide variety of fluids via gravimetric weighing. It has the same operational and functional intended use and remains the same in all respects to the currently marketed system, except for the proposed modifications to incorporate a system to aid in the verification of source solution identity and placement. The Positive Identification Solution Family Monitor, an equipment and software upgrade designed to aid in the verification of the identity of the source solutions being compounded, is a modification and upgrade to the compounding system currently marketed which is found substantially equivalent, for purposes of the Federal Food, Drug and Cosmetic Act only, to the Automix® 3+3 Compounder System under K894827, dated October 6, 1989. Performance testing for the Clintec Automix® 3 + 3 / AS Compounder System described in the submission demonstrates the functional acceptability of the proposed hardware and software modifications and enhancements. The design changes are documented and their verification and validation procedures are described in the application. 000315 MAR 11 1996 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Ms. Marcia Marconi Vice President Regulatory Affairs Baxter Healthcare Corporation Route 120 & Wilson Road Roound Lake, Illinois 60073 OCT 10 1997 Re: K961008 Trade Name: Automix® 3+3/AS Compounder System Regulatory Class: II Product Code: LHI Dated: September 26, 1997 Received: September 30, 1997 Dear Ms. Marconi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 {2} Page 2 - Ms. Marconi through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} OCT 9 '97 14:49 FROM REGULATORY AFFAIRS 10214803002-344771 PAGE.002/002 510(k) Number (if known): K961008 Device Name: Automix® 3+3/AS Compounder System Indications For Use: This device is an electromechanical positive displacement fluid compounding device that provides for the compounding of fluids via gravimetric weighing. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number K961008 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) 10/09/97 **TOTAL PAGE.002**
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...