MONOJECT BLUNT I.V. ACCESS CANNULA/MONOJECT HYPODERMIC SYRINGE WITH BLUNT I.V. ACCESS CANNULA
Device Facts
| Record ID | K960982 |
|---|---|
| Device Name | MONOJECT BLUNT I.V. ACCESS CANNULA/MONOJECT HYPODERMIC SYRINGE WITH BLUNT I.V. ACCESS CANNULA |
| Applicant | Sherwood Medical Co. |
| Product Code | FMI · General Hospital |
| Decision Date | Dec 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5570 |
| Device Class | Class 2 |
Intended Use
The Monoject™ Blunt I.V. Access Cannula is a sterile, lubricated plastic cannula with a female luer hub. It is used to provide access to pre-pierced IV line injection sites and pre-pierced medication vials.
Device Story
Sterile, lubricated plastic cannula with female luer hub; used for accessing pre-pierced IV injection sites and medication vials; available as individual cannula or pre-attached to 3ml, 6ml, or 12ml piston syringes; operated by clinicians; facilitates needle-free access to IV systems; reduces risk of needlestick injuries.
Clinical Evidence
Bench testing only. Comparative performance testing included insertion force, bend test, and break resistance. Biocompatibility testing included cytotoxicity, intracutaneous irritation, acute systemic toxicity, sensitization, hemolysis, subchronic toxicity, and drug compatibility.
Technological Characteristics
Plastic cannula; female luer hub; sterile; lubricated; available as standalone or pre-attached to piston syringes (3ml, 6ml, 12ml).
Indications for Use
Indicated for accessing pre-pierced IV line injection sites and pre-pierced medication vials in clinical settings.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
- B-D Interlink Syringe Cannula
- Monoject Lifeshield Blunt Cannula
- Monoject Lifeshield Blood Collection Device
Related Devices
- K974363 — BECTON DICKINSON BLUNT PLASTIC CANNULA · Bd Becton Dickinson Vacutainer Systems Preanalytic · Jun 3, 1998
- K031923 — ULTRASITE VALVE · B.Braun Medical, Inc. · Aug 11, 2003
- K013949 — UNIVERSAL VIAL ACCESS SPIKE WITH NEEDLELESS CONNECTOR · Churchill Medical Systems, Inc. · Mar 27, 2002
- K013087 — SINGLE DOSE DISPENSING PIN, MODEL 2201 · Alaris Medical Systems, Inc. · Nov 13, 2001
- K960661 — MAXCESS NEEDLE FREE CONNECTOR (8200) · Solopak Medical Products, Inc. · Mar 21, 1996