DISPOSABLE GAS SAMPLING INTERFACE KITS

K960972 · Marquest Medical Products, Inc. · CCK · Oct 18, 1996 · Anesthesiology

Device Facts

Record IDK960972
Device NameDISPOSABLE GAS SAMPLING INTERFACE KITS
ApplicantMarquest Medical Products, Inc.
Product CodeCCK · Anesthesiology
Decision DateOct 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1400
Device ClassClass 2

Intended Use

Indicated use - When a sample of a patient's expired gas is to be analyzed. Targeted population - Any patient requiring monitoring of expired gases. Environment of use - Hospital, anesthesia, ICU, PACU, respiratory therapy

Device Story

Small-bore tubing with luer fittings; facilitates transport of patient expired gas samples to external gas analyzers. Used in hospital environments (anesthesia, ICU, PACU, respiratory therapy) by clinicians. Device features various lengths, diameters, and optional in-line filters. Single-use disposable design.

Clinical Evidence

Bench testing only; performance testing of luer fittings confirms equivalence to predicate.

Technological Characteristics

Small-bore tubing with luer fittings; materials include PVC, PC, ABS, Acrylic, and PE. Optional in-line filter. Single-use, disposable.

Indications for Use

Indicated for patients requiring expired gas monitoring in hospital settings including anesthesia, ICU, PACU, and respiratory therapy.

Regulatory Classification

Identification

A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} The New Marq of Excellence K900972 OCT 18 1996 Marquest Medical Products, Inc. 11039 East Lansing Circle Englewood, Colorado 80112 U.S.A. 303 790 4835 phone 303 799 0210 fax Non-Confidential Summary of Safety and Effectiveness March 1, 1996 page 1 of 2 Marquest Medical Products, Inc. 11039 East Lansing Circle Englewood, CO 80112 Tel - (303) 790-4835 Fax - (303) 799-0210 Official contact: Tom Dielmann, Director, Regulatory Affairs / Quality Assurance Proprietary or Trade Name: Gas Sampling Interface Kits Common/Usual Name: Disposable Gas Sampling Lines Classification Name: Analyzer, Gas, Carbon Dioxide, Gaseous Phase Intended device: Gas Sampling Interface Kits Predicate devices: Gibeck, Inc. - disposable gas sampling lines - K841722 Device description: The intended product is a small bore tube with luer fittings which permit a sample of patient's expired gas to be delivered to an analyzer. The device may be of various lengths, tubing diameters and have an in-line filter. Intended use: Indicated use - When a sample of a patient's expired gas is to be analyzed. Targeted population - Any patient requiring monitoring of expired gases. Environment of use - Hospital, anesthesia, ICU, PACU, respiratory therapy Comparison to predicate devices: | Attribute | Marquest | Dryden (Gibeck) | | --- | --- | --- | | Use | | | | Indicated for use with expired gas sampling monitors | Yes | Yes | page 3 of 21 {1} Non-Confidential Summary of Safety and Effectiveness (continued) March 1, 1996 page 2 of 2 Comparison to predicate devices: (continued) | Attribute | Marquest | Dryden (Gibeck) | | --- | --- | --- | | Use (continued) | | | | Intended for use in anesthesia | Yes | Yes | | Intended for use in respiratory therapy | Yes | Yes | | Indicated for single use | Yes | Yes | | Design | | | | Luer fittings | Yes | Yes | | Clear | Yes | Yes | | Various diameters of tubing | Yes | Yes | | Various lengths of tubing | Yes | Yes | | Maybe supplied with filter | Yes | Yes | | Materials | | | | Luers made of PVC, PC, ABS, Acrylic | Yes | Yes | | Tubing - PVC, PE, PVC/PE | Yes | Yes | | Performance Testing | | | | Luer fittings | Yes | Yes | ## B. Differences Any differences that do exist would not have a significant effect on the safety or effectiveness of the device. page 4 of 21
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...