← Product Code [GEI](/productcode/GEI) · K960918

# VIDAMED TUNA (TRANSURETHRAL NEEDLE ABLATION) SYSTEM (K960918)

_Vidamed, Inc. · GEI · Oct 8, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K960918

## Device Facts

- **Applicant:** Vidamed, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 8, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The TUNA System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cc.

## Device Story

TUNA System treats BPH via transurethral needle ablation; utilizes RF energy to ablate prostate tissue. Components: TUNA 3 Catheter, Model 7205 RF Generator, return electrode, rectal monitoring tape, optics. Generator delivers up to 30W (15W/channel); monitors tissue temperature via thermocouple sensors in electrodes; adjusts power output accordingly. Used in clinical setting by physicians. Provides controlled thermal ablation to reduce prostate volume; alleviates urinary obstruction symptoms.

## Clinical Evidence

Confirmatory clinical data provided to support safety and effectiveness for BPH treatment.

## Technological Characteristics

RF electrosurgical generator; 30W max power (15W/channel); thermocouple-based temperature sensing/feedback control. Compliance with ANSI, AAMI, and UL 544 standards. Class II device.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- VIDAMED TUNA System ([K951245](/device/K951245.md))
- VIDAMED TUNA System ([K955035](/device/K955035.md))
- Radionics (pre-amendments)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

OCT - 8 1996

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

$K960918$

# VIDAMED TUNA® (Trans Urethral Needle Ablation) System

## Indication

The TUNA System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cc.

## Device Description

The TUNA® System consists of the TUNA® 3 Catheter, the VIDAMED Model 7205 RF Generator, the VIDAMED Return Electrode, the VIDAMED Rectal Monitoring Tape, and the TUNA® Optics.

The electrosurgical generator is capable of delivering up to 30W (15W/channel) of power and is designed for use with VIDAMED accessory electrodes including the TUNA® 3 Catheter. When used with VIDAMED accessory electrodes, the electrosurgical generator is capable of reading temperature from thermocouple sensors within the electrode and controlling power output to the tissue.

## Substantial Equivalence

The VIDAMED TUNA® 3 Catheter and VIDAMED Model 7205 RF Generator is substantially equivalent to devices provided by VIDAMED (K951245 and K955035), and Radionics (pre-amendments). Information is provided in the premarket notification to demonstrate that the properties of the TUNA® 3 Catheter and VIDAMED Model 7205 RF Generator are equivalent to equipment from the manufacturers listed above. Confirmatory clinical data is also provided to support the safety and effectiveness of the device for use in the treatment of symptoms due to urinary outflow obstruction secondary to BPH.

## Standards/Classifications

The VIDAMED Model 7205 RF Generator and accessories are designed to be in compliance with ANSI, AAMI, and UL 544 standards. The VIDAMED Model 7205 RF Generator and accessories are Class II medical devices.

---

**Source:** [https://fda.innolitics.com/device/K960918](https://fda.innolitics.com/device/K960918)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
