K960771 · Rymed Technologies, Inc. · FPA · Dec 10, 1996 · General Hospital
Device Facts
Record ID
K960771
Device Name
SLAV AND SLAV-ONE INJECTION PORT CONNECTOR SYSTEM
Applicant
Rymed Technologies, Inc.
Product Code
FPA · General Hospital
Decision Date
Dec 10, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Intended Use
The RyMed Technologies SLAV™ Injection Port Systems are intended for single patient use in intravenous and blood administration.
Device Story
SLAV™ Injection Port Systems consist of Heparin Lock Injection Ports or Y-Injection Site Ports featuring a two-piece valve mechanism. Designed for needleless intravenous and blood administration to comply with OSHA safety standards regarding needle stick injuries. Device operates by allowing fluid access through a silicone valve; supports minimum 100 activations without leakage. Used in clinical settings for IV therapy; provides high flow rates and low priming volumes to minimize dead space. Healthcare providers use the port to facilitate fluid delivery; benefits include reduced risk of accidental needle sticks and improved fluid management efficiency.
Clinical Evidence
Bench testing only. Comparative performance testing against Braun SafSite® and Medex Nu-Site™ evaluated fluid flow rate (≥150 cc/min), priming volume (<0.1 mls), and durability (100 activations). SLAV™ passed all tests; predicates failed priming volume requirements.
Technological Characteristics
Materials: ABS, Silicone, PVC, Polyethylene, Polypropylene. Sensing/Actuation: Mechanical two-piece silicone valve. Connectivity: None. Sterilization: ETO. Form factor: Heparin lock or Y-injection site ports, including J-loop and T-connector extensions.
Indications for Use
Indicated for single patient use in intravenous and blood administration.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Braun SafSite®
Medex Nu-Site™
Related Devices
K991653 — INVISION-PLUS INJECTION PORT CONNECTOR SYSTEM · Rymed Technologies, Inc. · Jun 24, 1999
K112893 — CLEARLINK LUER ACTIVATED VALVE, CLEARLINK SYSTEM NON-DEHP CATHETER EXTENSION SETS · Baxter Healthcare Corporation · Oct 18, 2011
K991599 — CADD ADMINISTRATION SET · Sims Deltec, Inc. · Jun 15, 1999
K120548 — HLIC NEEDLE-LESS VALVE · Health Line International Corporation · May 4, 2012
K031923 — ULTRASITE VALVE · B.Braun Medical, Inc. · Aug 11, 2003
Submission Summary (Full Text)
{0}
SLAV™ Injection Port Connector Systems Pre-Market Notification (510 k)
K 960111
Page 23 of 27
DEC 10 1996
# Summary of Safety and Effectiveness for the RyMed Technologies SLAV™ Injection Ports
submitted by
**RyMed Technologies, Inc.**
3110 Blue Sage Drive
Woodward, OK 73801
## Identification of a Legally Marketed Predicate Device
The RyMed Technologies SLAV™ Injection Port Systems are substantially equivalent Injection Ports Systems manufactured by:
- BRAUN MEDICAL, Bethlehem, Pennsylvania
- MEDEX, Hilliard, Ohio
## Device Description
The RyMed Technologies SLAV™ Injection Port Systems are composed of either a Heparin Lock Injection Port, or a Y-Injection Site Port that contain a specialized two piece valve made from a medical grade clear plastic and a medical grade Silicone material. This specialized valve completely eliminates the use of Needles forcing compliance with the new OSHA standards of safety avoiding accidental needle stick injuries, the valve also provides excellent fluid Flow Rates, requires very low Priming Volumes, and it eliminates “dead spaces” that could entrap blood. This valve has the ability to activate injection ports a minimum of 100 times without leaking or change in its performance.
## Intended for Use
The RyMed Technologies SLAV™ Injection Port Systems are intended for single patient use in intravenous and blood administration.
## Summary of Technological Characteristics
| Feature | SLAV™ Injection Port | Predicate Device
Braun SafSite® | Predicate Device
Medex Nu-Site™ |
| --- | --- | --- | --- |
| Intended Use | Intravenous and Blood Administration. | Intravenous and Blood Administration. | Intravenous and Blood Administration. |
| Disposable Single Patient Use | Yes | Yes | Yes |
RyMed Technologies, Inc.
{1}
SLAV™ Injection Port Connector Systems Pre-Market Notification (510 k) Page 24 of 27
Summary of Technological Characteristics continuation...
| Feature | SLAV™ Injection Port | Predicate Device
Braun SafSite® | Predicate Device
Medex Nu-Site™ |
| --- | --- | --- | --- |
| Flow Rate | 310 -350 mls / min. | > 250 mls / min. | > 250 mls / min. |
| Priming Vol. Required | 0.028 mls. | 0.30 mls | 0.12 mls. |
| Heparin Lock Injec. Ports | Yes | Yes | Yes |
| J - Loop Tube Extension | Yes | No | Yes |
| T - Connector/ Tube Ext. | Yes | No | Yes |
| IV Administration Sets | Yes | No | No |
| Multiple Access IV Manifold | Yes | Yes | Yes |
| Packaging | Blister Pack | Blister Pack | Blister Pack |
| Sterilization | ETO | Unknown, believed to be ETO | Unknown, believed to be ETO |
| Non-Pyrogenic | Yes | Yes | Yes |
| Materials | ABS, Silicone, PVC
Polyethylene/Polypropylene | Polycarbonate, Silicone
PVC, Polyethylene | ABS, Silicone, PVC
Polyethylene |
| Performance Standards | Unknown | Unknown | Unknown |
## Summary of Performance Data
| Test | SLAV™ Injection Port | BRAUN SafSite® | Medex Nu-Site™ |
| --- | --- | --- | --- |
| Fluid Flow Rate
≥ 150 cc / min. | Passed | Passed | Passed |
| Ideal Priming Volume
less than 0.1 mls | Passed | Failed | Failed |
| 100 Activation's with no
leaking or change in
performance capabilities | Passed | Passed | Passed |
RyMed Technologies, Inc.
{2}
SLAV™ Injection Port Connector Systems Pre-Market Notification (510 k) Page 25 of 27
# Conclusion
The SLAV™ Injection Port Systems meet the requirements to be considered similar in Technological Characteristics and in Performance to the predicate devices, we believe that the SLAV™ Injection Port Systems are safe and effective and perform as well as or better than the predicate devices. The SLAV™ will be manufactured per specifications and under Good Manufacturing Practices by an ISO 9000 Certified manufacturer to ensure the devices are safe and effective for its intended use.
RyMed Technologies, Inc.

RyMed Technologies, Inc.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.