CERTI COOL INSTANT COLD COMPRESS

K960624 · Certified Safety Mfg., Inc. · IMD · May 6, 1996 · Physical Medicine

Device Facts

Record IDK960624
Device NameCERTI COOL INSTANT COLD COMPRESS
ApplicantCertified Safety Mfg., Inc.
Product CodeIMD · Physical Medicine
Decision DateMay 6, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 890.5710
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Certi*Cool Instant Cold Compress is a self contained disposable unit that provides temporary cold therapy for body surfaces.

Device Story

Disposable cold pack; self-contained unit; inner plastic pouch with water, outer pouch with ammonium nitrate. Activation via manual rupture of inner pouch; endothermic chemical reaction produces temperatures below 40°F for minimum 15 minutes. Used for temporary cold therapy on body surfaces. Single-use device.

Clinical Evidence

Bench testing only; device performance verified by chemical reaction temperature output (below 40°F) and duration (minimum 15 minutes).

Technological Characteristics

Plastic pouch construction; ammonium nitrate and water chemical reactants; endothermic reaction principle; single-use disposable form factor.

Indications for Use

Indicated for temporary cold therapy for body surfaces in patients requiring localized cooling.

Regulatory Classification

Identification

A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.

Predicate Devices

Submission Summary (Full Text)

{0} Attachment # 5 # 510(k) SUMMARY THIS SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION IS BEING SUBMITTED IN ACCORDANCE WITH THE REQUIREMENTS OF THE SAFE MEDICAL DEVICE ACT OF 1990. K960624 510(k) Summary of Safety and Effectiveness Certi*Cool Instant Cold Compress MAY - 6 1996 Submitted by: Certified Safety Manufacturing, Inc. 1400 Chestnut Kansas City, MO. 64127 (816) 483-9090 Fax (816)-483-9091 Contact: Howard Gerson, Vice President Date Prepared: February 8, 1996 Trade Name: Certi*Cool Instant Cold Compress Common Name: Disposable Cold Pack Classification Name: Disposable Cold Packs are classified as Class I medical devices, and are listed in 21 CFR 890.5710 as cold disposable pack. The Certi*Cool Instant Cold Compress is substantially equivalent in design, composition and function to other disposable cold packs on the marketplace that employ a chemical reaction to produce cold. It is also substantially equivalent to our Edcool and Edcool J. Instant Cold Packs which were found SE by FDA on March 25, 1987 and assigned #K871006. The Certi*Cool Instant Cold Compress is a self contained disposable unit that provides temporary cold therapy for body surfaces. Each unit is labeled with instructions for use as well as specific warning statements. Our product consists of an inner plastic pouch containing a green-colored water and an outer pouch containing ammonium nitrate. To activate the product, simply rupture the inner pouch by using an external force, such as a fist. Then, as the two ingredients mix, a chemical reaction results by producing temperatures below 40°F for a minimum of 15 minutes. These are single use devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...