ARROW BIPOLAR PACING/BALLOON WEDGE PRESSURE CATHETER
Device Facts
| Record ID | K960479 |
|---|---|
| Device Name | ARROW BIPOLAR PACING/BALLOON WEDGE PRESSURE CATHETER |
| Applicant | Arrow Intl., Inc. |
| Product Code | LDF · Cardiovascular |
| Decision Date | Oct 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.3680 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The ARROW catheter is indicated for use in sampling blood for oxygen levels and measuring pressure in the right heart (including central venous pressure, right ventricular pressure, pulmonary artery pressure, and pulmonary artery wedge pressure). The second pressure lumen allows infusion of a solution with simultaneous measurement of pressure. The catheter is intended for temporary use in electrophysiology studies for intracardiac simulation and/or ECG recording only.
Device Story
Radiopaque polymeric catheter; features two pressure lumens and balloon inflation lumen; two 2mm platinum electrodes for electrical conduction; 10cm increment markings. Used for right heart pressure measurement (CVP, RVP, PAP, PAWP), blood oxygen sampling, and temporary intracardiac stimulation/ECG recording. Operated by clinicians in clinical settings. Provides pressure data and electrical signals to external monitoring/recording equipment; facilitates solution infusion. Benefits include simultaneous pressure monitoring and pacing/recording capabilities.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included catheter flow rate, column strength, balloon inflation, torque, flexural modulus, lead pull, electrical continuity, signal-to-noise ratio, impedance, high pot, capacitance/inductance, and radiopacity.
Technological Characteristics
Radiopaque polymeric tubing; two 2mm platinum electrodes; two pressure lumens; balloon inflation lumen; female Luer fittings; 10cm increment markings. Electrical connectivity via proximal extension leads. Sterilization method not specified.
Indications for Use
Indicated for patients requiring right heart pressure monitoring, blood oxygen sampling, and temporary intracardiac stimulation or ECG recording during electrophysiology studies.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Predicate Devices
- Elecath® Ventricular Pacewedge® Pacing and Pressure Catheter (K832850, K844003)
- Arrow Balloon Temporary Bipolar Pacing Catheter (pre-enactment)
- Arrow Balloon Wedge Pressure Catheter (pre-enactment)