The Denver® Paracentesis Needle, Paracentesis Needle with Drainage Line, and Paracentesis Tray are indicated for use in removing fluid from the peritoneal cavity (abdominal paracentesis). They are intended to be used for both diagnostic and therapeutic abdominal paracentesis procedures.
Device Story
Single-use, sterile paracentesis device; includes needle, drainage line, and tray configurations. Input: manual insertion into abdominal cavity. Mechanism: two-part trocar with blunt, rounded-end inner tube and sharp-beveled outer needle; blunt tip extends automatically upon entering low-resistance space (peritoneal cavity) to prevent organ puncture; red indicator confirms sharp needle exposure. Output: drainage of peritoneal fluid via cannula. Used by clinicians in clinical settings. Benefits: facilitates fluid removal while reducing risk of inadvertent organ puncture via automatic blunt-tip extension.
Clinical Evidence
Bench testing only. Drainage rates compared to Tri-Med Caldwell 15 GA Paracentesis Needle/Cannulas; results demonstrated similar drainage performance.
Technological Characteristics
Sterile, single-use device. Components: inner two-part trocar, outer drainage cannula, trocar/cannula housing. Features: automatic blunt-tip extension mechanism, red visual indicator for sharp needle exposure, sealed housing to prevent air/fluid leakage, luer port with sealing diaphragm. Non-pyrogenic fluid path.
Indications for Use
Indicated for patients requiring diagnostic or therapeutic abdominal paracentesis to remove fluid from the peritoneal cavity.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
Tri-Med Specialties, Inc. Paracentesis Tray (K932652)
Reference Devices
Tri-Med Caldwell 15 GA Paracentesis Needle/Cannulas (K932652)
Submission Summary (Full Text)
{1}
K960456
# Summary of Safety and Effectiveness
## Denver® Paracentesis Needle (#60-1200)
## Denver® Paracentesis Needle with Drainage Line (60-1100)
## Denver® Paracentesis Tray (#60-1000)
### IDENTIFICATION OF SUBSTANTIALLY EQUIVALENT DEVICE(S)
The Denver® Paracentesis Needle, Paracentesis Needle with Drainage Line, and Paracentesis Tray are substantially equivalent in intended use to the *Endovations, Inc. Paracentesis Tray (K874063)*. The Denver® and Endovations devices are similar in that they are both single-use, sterile devices used for draining fluid from the peritoneal cavity (abdominal paracentesis), both for diagnostic and therapeutic purposes. Both trays provide the primary components required to perform paracentesis. The two devices differ in that the Denver® needle has a metal cannula and a two-part, non-coring trocar (the inner rounded end tube extends beyond the outer sharp needle whenever the trocar is in an area of low resistance, such as when the abdominal cavity is entered). The Endovations needle has a plastic cannula and a single non-coring sharp needle trocar. In addition, the Endovations needle is only offered as part of a paracentesis tray, whereas the Denver® needle is also offered alone (Catalog No. 60-1200) and with a drainage line and connectors (Catalog No. 60-1100). Both the Denver® and Endovations needles have sealed housings that prevent leakage of air or fluid into or out of the lumen through which the trocar was placed. This seal also prevents reinsertion of the trocar in both needles. The drainage ports of both devices (a luer port with a sealing diaphragm in the Denver® needle; a three-way stopcock in the Endovations needle) also prevent passage of air or fluid in or out of the port until accessed.
The Denver® Paracentesis Needle, Paracentesis Needle with Drainage Line, and Paracentesis Tray are also substantially equivalent in intended use to the *Tri-Med Specialties, Inc. Paracentesis Tray (K932652)*. The Denver® and Tri-Med devices are similar in that they are both single-use, sterile devices used for draining fluid from the peritoneal cavity, and both incorporate a metal drainage cannula. Both trays provide the primary components required to perform paracentesis. The two devices differ in that the Denver® needle has a two-part, non-coring trocar and the Tri-Med needle has a single non-coring sharp needle trocar. In addition, the Denver® needle has a sealed housing that prevents leakage into or out of the lumen through which the trocar was placed, and prevents reinsertion of the trocar. The Tri-Med needle has no seal in the housing; thus
9t
{2}
fluid leakage out of the needle can occur with the trocar either in place or removed. Similar to the Denver® needle, the Tri-Med needle is offered alone, with a drainage line, and as part of a complete paracentesis tray.
## DEVICE DESCRIPTION:
The Denver® Paracentesis Needle is a sterile device, for single use only. It is not to be resterilized. Components in the fluid path are non-pyrogenic. The device consists of an inner two-part trocar, an outer drainage cannula, and a trocar/cannula housing. The inner two-part trocar is a rounded, closed end tube with side holes, coaxially housed within a sharp-beveled hollow needle and the outer drainage cannula. The outer drainage cannula has a beveled tip and distal side holes to facilitate drainage.
As the two-part trocar penetrates the abdominal wall, the blunt portion of the trocar is forced back within the shaft of the needle portion of the trocar, exposing the sharp-beveled needle. The trocar/cannula housing contains a red indicator which is visible when the sharp-beveled needle is exposed. When the needle encounters low resistance, such as when the abdominal cavity is entered, the blunt portion of the trocar automatically re-extends out beyond the sharp needle, thereby helping protect against inadvertent organ puncture. Once the blunt portion of the trocar re-extends beyond the needle, the red indicator is no longer visible. Proper position in an area of fluid may also be confirmed by attaching a syringe to the end of the device and attempting aspiration.
Once in the abdominal cavity, the two-part trocar is removed, leaving the outer drainage cannula and housing in place for the paracentesis procedure.
The Denver® Paracentesis Needle with Drainage Line is a sterile device intended for single use only. It is not to be resterilized. The product contains one Denver® Paracentesis Needle, one drainage tubing line, and connectors to allow attachment to wall suction or a vacuum bottle.
The Denver® Paracentesis Tray is a sterile device intended for single use only. It is not to be resterilized. The tray contains one Denver® Paracentesis Needle, one drainage tubing line, connectors to allow attachment to wall suction or a vacuum bottle, and the primary components needed for performing the paracentesis procedure.
91
{3}
Page 3 of 3
# INTENDED USE
The Denver® Paracentesis Needle, Paracentesis Needle with Drainage Line, and Paracentesis Tray are indicated for use in removing fluid from the peritoneal cavity (abdominal paracentesis). They are intended to be used for both diagnostic and therapeutic abdominal paracentesis procedures.
# SCIENTIFIC LITERATURE
Published studies indicate that abdominal paracentesis is a safe and effective procedure.
# PERFORMANCE TESTING:
Drainage rates for Denver® Paracentesis Needles were compared to Tri-Med Caldwell 15 GA Paracentesis Needle/Cannulas (K932652). The results indicate that the two devices have a similar drainage performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.