3I STND THREADED/SELF TAPPING THREADED/3I MINIPLANT IMPLANTS/3 I EMERGENCE PROF. TEMP. HEALING ABUTMENTS
K960417 · Implant Innovations, Inc. · DZE · Aug 1, 1996 · Dental
Device Facts
Record ID
K960417
Device Name
3I STND THREADED/SELF TAPPING THREADED/3I MINIPLANT IMPLANTS/3 I EMERGENCE PROF. TEMP. HEALING ABUTMENTS
Applicant
Implant Innovations, Inc.
Product Code
DZE · Dental
Decision Date
Aug 1, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A single-stage surgical process by which a 3i implant is placed into the bone and a 1 or 2 piece (EP) Temporary Healing Abutment is immediately attached to the implant, may be considered in those cases where the implant and healing abutment can be adequately protected from mastication forces throughout the healing period, by opposing dentition, denture or bridge that if required, may be modified to provide such protection.
Device Story
3i Implant System consists of titanium/titanium alloy endosseous implants and Emergence Profile (EP) Temporary Healing Abutments. Originally designed for two-stage surgery; new indication allows single-stage procedure. Surgeon places implant into jawbone and immediately attaches EP healing abutment; mucoperiosteal flaps sutured around abutment. Soft tissues heal to abutment contours, creating natural emergence profile. Requires protection from mastication forces via modified denture or bridge during healing. Eliminates second surgery for abutment placement; reduces infection risk; shortens reconstruction timeline. Used by dental surgeons in clinical settings. Output is a healed implant site ready for final prosthetic restoration.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on the established clinical history of the 3i implant system and the clinical rationale for single-stage surgical protocols.
Technological Characteristics
Materials: Titanium and titanium alloy implants with Titanium Plasma-Sprayed or Hydroxylapatite coatings. Form factor: Standard/Self-Tapping Threaded Screw-Type and cylindrical press-fit implants. Components: 1 and 2 piece Temporary Healing Emergence Profile (EP) Abutments. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring dental implants where adequate bone/soft tissue width and height exist to secure the implant/abutment. Contraindicated in patients with poor bone quality, poor oral hygiene, heavy smoking/tobacco abuse, uncontrolled diabetes, blood disorders, or cases where the implant cannot be protected from premature mastication loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}
3i
IMPPLANT INNOVATIONS
3071 Continental Drive
West Palm Beach, FL 33407
1-800-443-8166 (407) 840-2600
FAX (407) 840-2660
AUG - 1 1996
k960417
# 510(k) SUMMARY
NEW INDICATION FOR USE: SINGLE STAGE SURGERY USING DEVICES ORIGINALLY INDICATED FOR A TWO-STAGE SURGICAL PROCEDURE.
To the Requestor:
This information is taken directly from the original Pre-Market Notification [510(k)], submission, provided to the United States Food and Drug Administration. No information regarding safety or efficacy has been deleted from that submission, for this summary.
1. CLASSIFICATION NAME: Endosseous Dental Implants
2. COMMON/USUAL NAMES: Standard Threaded Implants, Self-Tapping Threaded Implants, Titanium Plasma Sprayed Cylindrical Implants, Miniplants (TM), with 1 and 2 Piece Temporary Healing, Emergence Profile (EP) Abutments.
3. PROPRIETARY NAME: 3i Standard Threaded Implant, Self-Tapping Threaded Implants and 3i Miniplant (TM) Implants. 3i Emergence Profile (EP) Temporary Healing Abutments.
The 3i implant system was submitted in a Pre-Market Notification that has since been amended and modified by subsequent submissions. The 3i implant system was originally determined substantially equivalent on or about May 11, 1988. For the purpose of this submission, all 3i implants and One and Two Piece Temporary Healing, Emergence Profile (EP) Abutments are to be considered.
4. CLASSIFICATION: Endosseous dental implants (including abutments), per 872.3640 have been classified as class III devices. PMA's may be required for endosseous dental implants in the future, but no effective date has been established for such submissions.
5. PERFORMANCE STANDARDS: Not applicable at this time.
Page 1 of 6
{1}
6. PRIOR INDICATIONS FOR USE: The 3i Implant System is designed for use in dental implant surgery. It includes a variety of types and sizes of specially designed bone implantable titanium and titanium alloy implants, including Standard and Self-Tapping Threaded Screw-Type and cylindrical press-fit implants with Titanium Plasma-Sprayed or Hydroxylapatite coatings. These implants are surgically inserted into the upper and/or lower jawbones.
Temporary Healing, Emergence Profile (EP) Abutments were designed to be attached to the implant at a second stage surgical procedure, after a sufficient healing time for the implant/bone interface to form. Healing abutments were placed to help form the soft tissues to the proper diameter of the final restoration, creating a natural Emergence Profile of the prosthesis.
A successfully integrated implant will achieve a firm and direct connection between the living bone and the surface of the titanium or titanium alloy implant when surgically implanted under controlled conditions, per well known clinical studies.
Historically, the process for placing an implant, through final prosthesis, required a minimum of two surgical procedures. Stage one surgery included all activities required to place an implant/cover screw into the bone and suturing the mucoperiosteal flaps over the implanted devices. A healing period of between three and six months (or longer) followed, after which the site was surgically reopened (second stage surgery), and temporary healing abutments, provisional prosthetic components or final abutments were placed. Again, the mucoperiosteal flaps were sutured closed around the abutments and a healing period allowed for the soft tissues to heal.
7. PROPOSED NEW INDICATION FOR USE: Most clinicians recognize the need to protect the implant from premature loading throughout the required healing period. They also recognize the benefits of a single surgical procedure in cases that provide adequate protection of the implant/healing abutment; such as with use of a modified, soft refined denture or a bridge, preventing mastication forces from affecting the implanted devices.
In such cases, clinicians have used "Single-Stage" implant systems such as "The ITI Dental Implant System" distributed by The Straumann Company or, the "Immediate-Load Implant" distributed by Sargon Enterprises, Inc., as well as others, or have utilized two-stage implant systems offering temporary healing abutments, such as the 3i Implant/EP System.
Page 2 of 6
{2}
Page 3 of 6
A single-stage surgical process by which a 3i implant is placed into the bone and a 1 or 2 piece (EP) Temporary Healing Abutment is immediately attached to the implant, may be considered in those cases where the implant and healing abutment can be adequately protected from mastication forces throughout the healing period, by opposing dentition, denture or bridge that if required, may be modified to provide such protection.
The EP Temporary Healing Abutment may be a one or two piece abutment and, will be removed after the implant is firmly anchored into the healed implant site, without surgically altering the soft tissues.
Upon placement of the implant, instead of a cover screw, a Temporary (EP) Healing Abutment, that is of suitable height for the soft tissues, is secured to the implant. The mucoperiosteal flaps are then sutured closed around the healing abutment. During the healing phase, the soft tissues form to the diameter and contours of the healing abutment, forming the tissues to the "more natural emergence profile". This benefits both patient and surgeon in that only one surgical procedure is required. Both bone and soft tissues heal during the same time period, shortening the entire implant/reconstruction process by several months and reducing the chance for infection or other complications attributed to surgery.
Upon proper healing, impressions are taken using the 3i EP Impression system (K934126), and temporary or final prosthesis fabricated. No further surgery is required.
8. CONTRAINDICATIONS: A single stage surgical procedure using 3i implants and healing abutments is not indicated in cases where remaining bone or soft tissues are too diminished to provide adequate width or height to secure or surround the implant/abutment. Lack of integration and subsequent implant failure may occur in cases where there is an insufficient quantity of bone, poor bone quality, poor oral hygiene, heavy smoking/tobacco abuse, or numerous medical conditions including blood disorders and uncontrolled diabetes (See Sections 9-12).
A single surgical stage process utilizing 3i implants and healing abutments is not indicated in cases where the implants cannot be adequately protected from premature loading by mastication forces throughout the required healing period. Protection may be obtained by the use of a modified, soft relined denture or a bridge.
9. WARNINGS: For safe and effective use of 3i implants, it is strongly suggested that specialized training be undertaken since the surgical techniques required to place dental implants are highly specialized and complex
{3}
31/CUSTOMER SERV
TEL:4078402626
procedures. Improper patient selection and technique can cause implant failure and/or loss of supporting bone.
10. PRECAUTIONS: Thorough screening of prospective implant candidates must be performed. Visual inspection as well as panoramic and periapical radiographs are essential to determine anatomical landmarks, occlusal conditions, periodontal status, and adequacy of bone. Lateral cephalometric radiographs, CT Scans, and tomograms may also be beneficial.
11. ADVERSE EFFECTS: Loss of implant anchorage (failure to osseointegrate) and loss of the prosthesis are possible occurrences after surgery. Lack of quantity or quality of remaining bone, infections, poor patient oral hygiene or cooperation, and generalized diseases (diabetes, etc.) are some potential causes for loss of anchorage.
12. SURGICAL COMPLICATIONS: The implant procedure has risks, including localized swelling, dehiscence, tenderness of short duration, edema, hematoma, or bleeding. Numbness of the lower lip and chin region following lower jaw surgery, and of the tissue beside the nose following upper jaw surgery, is a possible side effect of the surgery. Though it would most probably be of a temporary nature, in very rare cases numbness has been permanent. Gingival/Mucosal (gum tissue) ulceration, tissue reaction, or infection may occur, but generally responds to local care.
13. LABEL/LABELING MATERIALS: The proposed new indication for use will not necessitate a revision to the device labels. Instruction sheets and marketing materials will be developed to reflect the use of the EP Temporary Healing Abutments in conjunction with 3i Implants at Stage 1 surgery. This material will specify that in those cases where the implant/abutment can be adequately protected from premature loading and mastication forces throughout the normal healing periods, a single stage surgical protocol may be considered.
14. SUBSTANTIAL EQUIVALENCE: A single-stage surgical protocol utilizing the 3i Implant/Temporary Healing (EP) Abutment systems, is substantially equivalent to cases indicated for the use of implants distributed by the Straumann Company, "Non-Submerged Implant System", a single surgical stage implant system and the "Immediate-Load Implant" distributed by Sargon Enterprises, Inc. whose claims assert that the implant achieves immediate bone integration and therefore may be immediately loaded with temporary crowns, in that both systems are designed for single stage surgical procedure.
Page 4 of 6
{4}
The 3i proposed single-stage surgical protocol is dissimilar from the Straumann system, in that it is a two or three component system utilizing the 3i 1 and 2 piece Temporary Healing (EP) Abutments in conjunction with the implant; whereas the Straumann implant is one-piece (w/ cover screw). Use of EP Healing Abutments improves on the Straumann system in that they offer a greater range in height and diameters, to more closely match the patient's soft tissues. The wide range of healing abutments also further assures that the coronal aspect of the implant/abutment unit can be more accurately placed at or just sub-gingivaly, affording greater protection from mastication forces. With the Straumann system the clinician must guess the height of the available soft-tissues when placing the one-piece implant. This may produce a variation in heights of the coronal aspect of the implant.
The 3i proposed single-stage surgical protocol is dissimilar from both Straumann and the Sargon system in that 3i at this time, does not recommend immediate loading under any circumstances.
15. 510(k) CERTIFICATION AND SUMMARY FOR SUBMISSION:
I certify that I have conducted a reasonable search of all information known or otherwise available to me about the types and causes of safety and/or effectiveness problems that have been reported for Endosseous Dental Implant systems.
Failure to osseointegrate or loss of osseointegration can be caused by improper patient selection (patients with systemic diseases which affect bone physiology, patients with habits such as bruxing or clenching, patients who are physically or psychologically unable to carry out proper implant hygiene, heavy smoking or alcohol use), by improper surgical technique (overheating of bone) or improper case planning or restorative technique (overloading of implants through improper placement, use of an insufficient number of implants or excessive cantilever). Improper implant processing by the manufacturer or improper handling by the customer, resulting in contamination, can also effect osseointegration.
Fracture of implants can occur, particularly in implants with apical cross-holes. Fracture occurs either on insertion of screw-type implants due to excessive torque (improper surgical technique such as an error in drill selection) or in service due to loss of bone.
Fracture of abutments and abutment screws occurs in implant systems and is usually attributed to factors within the control of the implant team, such as lack of
Page 5 of 6
{5}
passive fit of the restoration or excessive cantilever, or within the control of the patient, such as bruxing.
Other types of safety and efficacy problems which have been observed for endosseous dental implant systems are local soft tissue degeneration and bone resorption, paresthesia, perforation of the maxillary sinus, perforation of labial and lingual plates, local and systemic infection, prosthetic framework fracture, nerve injury, bone fracture, injury to adjacent teeth and their supporting bone, oroantral or oronasal fistula, gingival hyperplasia, soft tissue overgrowth, perforation of the gingiva by the healing screw, mucosal abscess, displacement of the implant into the mandibular canal, hemorrhage of the floor of the mouth due to transection of the sublingual artery and breakage of drill tip, requiring surgical removal.

William G. Conety
Director, Regulatory
Affairs/Quality Assurance
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.