DSL ACTIVE I-HCG IRMA
Device Facts
| Record ID | K960357 |
|---|---|
| Device Name | DSL ACTIVE I-HCG IRMA |
| Applicant | Diagnostic Systems Laboratories, Inc. |
| Product Code | JHI · Clinical Chemistry |
| Decision Date | May 7, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1155 |
| Device Class | Class 2 |
Intended Use
The DSL I-hCG IRMA assay is intended for the quantitative determination of I-hCG in human serum. It is intended for in vitro diagnostic use by professional laboratory personnel as an aid in the detection of pregnancy.
Device Story
DSL 8300 I-hCG IRMA kit; non-competitive immunoradiometric assay (IRMA) for quantitative hCG measurement in human serum. Principle: 'sandwich' immunoassay using two antibodies; one immobilized on test tube wall, one radiolabeled for detection. Procedure: sample incubation; decanting/washing to remove unbound material; gamma counter analysis of bound counts per minute. Output: hCG concentration directly proportional to bound counts. Used in clinical laboratories by professional personnel. Results aid clinicians in pregnancy detection.
Clinical Evidence
Bench testing only. Comparative study of 53 patient serum samples analyzed by both DSL 8300 I-hCG IRMA and predicate DPC hCG IRMA. Results showed strong correlation (r=0.96) with linear regression equation Y = 0.86(X) + 9.5.
Technological Characteristics
Non-competitive immunoradiometric assay (IRMA). Components: antibody-coated test tubes, radiolabeled antibody reagent. Detection via gamma counter. In vitro diagnostic use.
Indications for Use
Indicated for the quantitative measurement of human chorionic gonadotropin (hCG) in human serum to aid in the detection of pregnancy in clinical laboratory settings.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
- DPC hCG IRMA
Related Devices
- K961911 — DSL ACTIVE I-HCG ELISA · Diagnostic Systems Laboratories, Inc. · Jun 28, 1996
- K223690 — iFlash-HCG; Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) · Shenzhen Yhlo Biotech Co., Ltd. · Dec 11, 2023