POSTERIOR ANNULOPLASTY BAND (MODEL 607)

K960356 · The Heart Valve Div. Medtronic Cardiovascular Surg · KRH · Aug 27, 1996 · Cardiovascular

Device Facts

Record IDK960356
Device NamePOSTERIOR ANNULOPLASTY BAND (MODEL 607)
ApplicantThe Heart Valve Div. Medtronic Cardiovascular Surg
Product CodeKRH · Cardiovascular
Decision DateAug 27, 1996
DecisionST
Submission TypeTraditional
Regulation21 CFR 870.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND is indicated for the reconstruction and/or remodeling of the pathological mitral valve. Combined mitral valve stenosis and insufficiency may be corrected by appropriate commissurotomy and valvular remodeling.

Device Story

The Medtronic Posterior Annuloplasty Band is a surgical implant used during mitral valve repair. It consists of a polyester fabric-covered, pliable silicone elastomer strip impregnated with barium sulfate for radiopacity. The device is 3 mm in diameter and 63 mm in length, featuring colored suture markers for surgical orientation. It is implanted by a cardiac surgeon during open-heart surgery to remodel the mitral valve annulus. By providing structural support to the posterior annulus, the band helps correct mitral valve stenosis and insufficiency. It is a passive implant; it does not require external power or software. The device is provided sterile for single use.

Clinical Evidence

No new clinical trials were conducted. Evidence relies on the established clinical record of the predicate Duran Flexible Annuloplasty Ring (over 15 years of documented performance) and a published report comparing posterior annuloplasty to commissural and full ring annuloplasty. Biocompatibility testing (cytotoxicity and hemolysis) was performed per USP requirements.

Technological Characteristics

Materials: Polyester fabric, silicone elastomer, barium sulfate. Dimensions: 3 mm diameter, 63 mm length. Radiopaque. Sterilization: Sterile double aseptic transfer pouch. Passive implant; no energy source or software.

Indications for Use

Indicated for patients requiring reconstruction or remodeling of pathological mitral valve due to stenosis or insufficiency. Contraindicated in patients with heavy valvular calcification, valvular retraction with severely reduced mobility, or active bacterial endocarditis.

Regulatory Classification

Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{1} K960.356 3/27/96 # Summary of Safety and Effectiveness ## MEDTRONIC® POSTERIOR ANNULOPLASTY BAND ### I. General Information Device Generic Name: Annuloplasty Band Device Trade Name: MEDTRONIC® POSTERIOR ANNULOPLASTY BAND Applicant's Name: MEDTRONIC® Heart Valves, Inc. Applicant's Address: 18011 South Mitchell Irvine, CA 92714 ### II. Description of the Diseases and Conditions for Which the Device is Intended The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND is indicated for the reconstruction and/or remodeling of the pathological mitral valve. Combined mitral valve stenosis and insufficiency may be corrected by appropriate commissurotomy and valvular remodeling. The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND is contraindicated for use under the following conditions: the presence of heavy valvular calcification, valvular retraction with severely reduced mobility and the presence of active bacterial endocarditis. ### III. Device Description The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND is constructed of a polyester fabric enclosing an open strip of pliable radiopaque silicone elastomer impregnated with barium sulfate which enables radiographic visualization. The band is uniformly 3 mm in diameter and is provided in a single length of 63 mm. Colored suture markers, which facilitate implantation, are circumferentially located at the center of the band and at each end. The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND will be offered in one length, 63 mm and will not be mounted onto a holder. No sizers will be offered with the posterior band configuration as only one length is available. MEDTRONIC POSTERIOR ANNULOPLASTY BAND JANUARY 24, 1996 A2-1 9600018 {2} The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND will be supplied in a sterile double aseptic transfer pouch. The annuloplasty band will remain sterile if the pouch is undamaged and unopened. The MEDTRONIC® POSTERIOR ANNULOPLASTY BAND uses the same materials and is processed, packaged and sterilized similarly to the MEDTRONIC® DURAN FLEXIBLE ANNULOPLASTY RING. ## IV. Alternatives Alternatives for patients requiring mitral valve annuloplasty is valve replacement surgery. Alternatives to use of the MEDTRONIC® POSTERIOR ANNULOPLASTY BAND is commissural annuloplasty and other commercially available annuloplasty bands or rings. ## V. Potential Adverse Effects Potential adverse effects of annuloplasty include thromboembolic events, ring dehiscence, hemolysis, stenosis, uncorrected or residual incompetence, heart block, and endocarditis. ## VI. Summary of Studies **Clinical Experience:** Appendix 3 contains a copy of a published report comparing the clinical experience of posterior annuloplasty (This report details clinical experience utilizing 2/3's of a MEDTRONIC® DURAN FLEXIBLE ANNULOPLASTY RING.) with that of commissural annuloplasty and full ring annuloplasty. In addition, the MEDTRONIC® DURAN FLEXIBLE ANNULOPLASTY RING has an established clinical record of documented performance for over 15 years. The MEDTRONIC® Posterior Annuloplasty Ring is composed of identical materials and is manufactured utilizing the same assembly, packaging and sterilization operations as the Duran Ring. Thus, the established clinical history of the Duran Flexible ring is germane to the MEDTRONIC® POSTERIOR ANNULOPLASTY BAND. Copies of publications documenting this performance can be found in Appendix 3. **Biocompatibility of Materials/Finished Device:** Cytotoxicity and Hemolysis testing was performed. This testing was performed in accordance with USP requirements. MEDTRONIC POSTERIOR ANNULOPLASTY BAND JANUARY 24, 1996 A2 - II 00019 {3} Qualification: The integrity of the Posterior Annuloplasty Band is considered to be the strength of the band, as measured by the "Wire Pull Out Test". "Wire Pull Out Test" Data for the Duran Flexible Annuloplasty Ring which had been subjected to a 100% Sterilization process (Appendix 9) was presented in an FDA notification provided in October, 1995. The data available for the Duran Flexible Annuloplasty Ring is considered to be germane for the Posterior Annuloplasty Band as the materials and manufacturing processes (inclusive of packaging and sterilization) are similar. The open configuration of the Posterior Annuloplasty Band has no bearing on the results of the "wire pull out test". For your convenience, a summary of this test data can be located in Appendix 5. A process qualification was performed in order to determine if the band could be manufactured to specifications. The process qualification demonstrated that a pliable band could be produced in accordance with the established product specifications. The process qualification report can be found in Appendix 5. V. Conclusions The testing performed for the MEDTRONIC® POSTERIOR ANNULOPLASTY BAND and examination of the clinical experience of the predicate device provides reasonable assurance that the MEDTRONIC® POSTERIOR ANNULOPLASTY BAND will perform in a safe and effective manner when used as indicated. I believe, to the best of my knowledge, that all data and information submitted in this 510(k) are truthful and accurate, and that no material fact has been omitted. Kathleen T. Boehm Product Regulation Manager MEDTRONIC® Heart Valves, Inc. 1-24-96 MEDTRONIC POSTERIOR ANNULOPLASTY BAND JANUARY 24, 1996 A2 - III 00020
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...