ENDOSCOPIC ACCESS DEVICE
Device Facts
| Record ID | K960232 |
|---|---|
| Device Name | ENDOSCOPIC ACCESS DEVICE |
| Applicant | Applied Medical Technologies |
| Product Code | GCJ · Gastroenterology, Urology |
| Decision Date | Mar 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The Endoscopic Access Device is an instrument access portal for use during endoscopic surgery.
Device Story
Instrument access portal for endoscopic surgery; internal flange inserted into incision site; expands to retain position; allows passage of surgical instruments; external flange reduces site contamination and trauma; defines access region; shaft angulated 20°-90° (30° standard); single-use; sterile; physician-operated in clinical/surgical setting; provides stable access path for endoscopic tools; facilitates surgical procedures.
Clinical Evidence
Bench testing only; evaluated minimum pull-out force versus device strength to ensure mechanical integrity and functional performance.
Technological Characteristics
Injection molded biocompatible Dow Corning silicone (Q7-4840); angulated shaft (20°-90°); internal/external flange design; single-use; sterile.
Indications for Use
Indicated for use as an instrument access portal during endoscopic surgery in patients requiring endoscopic access.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- AMT Esophageal Overtube
Reference Devices
- AMT Button (K943392)
Related Devices
- K150705 — FloShield 10mm Endoscopic Cannula and Conical Blunt Obturator, FloShield 10mm Endoscopic Cannula, FloShield 10mm Conical Blunt Obturator · Minimally Invasive Devices, Inc. · Apr 13, 2015
- K965121 — CANNULA · Origin Medsystems, Inc. · Mar 20, 1997
- K252714 — C-Lant Port · Vigor Medical Technologies , Ltd. · Feb 27, 2026
- K212500 — ArtiSential Trocar · Livsmed, Inc. · Dec 3, 2021
- K963115 — TROCAR CANNULA · United States Surgical, A Division of Tyco Healthc · Feb 12, 1997