← Product Code [LNH](/productcode/LNH) · K960181

# VERSION 3.7 OPERATING SYSTEM SOFTWARE (K960181)

_Hitachi Medical Systems America, Inc. · LNH · Mar 5, 1996 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K960181

## Device Facts

- **Applicant:** Hitachi Medical Systems America, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Mar 5, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

## Device Story

Magnetic Resonance (MR) diagnostic device; utilizes hydrogen proton excitation via RF pulses and magnetic field gradients to generate spatial images. Version 3.7 software update adds Dynamic Scan measurement/analysis, MTC and SSP for 3D TOF MRA, additional 2D/3D SARGE sequences, expanded SE/FSE flip angle ranges, Fast IR/High-Res Fast IR, and post-processing filters/coil sensitivity correction. Operated by trained clinicians in stationary or mobile configurations. Input: RF echoes from proton relaxation; Output: 2D/3D cross-sectional anatomical images. Physician interprets images to assess internal structures, aiding clinical diagnosis without ionizing radiation.

## Clinical Evidence

No clinical data provided. Bench testing only.

## Technological Characteristics

MR imaging system; proton excitation; RF pulse/gradient-based spatial encoding. Software-defined pulse sequences (SE, FSE, IR, FIR, GE, SARGE, TOF MRA). Post-processing includes adaptive filters and receiver coil sensitivity correction. Connectivity/platform details not specified beyond operating system software versioning.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Hitachi AIRIS with Version 3.4C Operating System Software
- Hitachi MRP-7000 with Version 3.3B Operating System Software

## Reference Devices

- [K905834](/device/K905834.md)
- K920441A
- [K925009](/device/K925009.md)
- [K926397](/device/K926397.md)
- [K934485](/device/K934485.md)
- [K935664](/device/K935664.md)
- [K935671](/device/K935671.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

510(k) SUMMARY

K960181

# 1.0 SUBMITTER INFORMATION:

1.1 Submitter: Hitachi Medical Systems America
1963 Case Parkway
Twinsburg, OH 44087
PH: 216 425-1313
FX: 216 425-1410

1.2 Contact: James Jochen Rogers
1.3 Date: January 11, 1996

# 2.0 DEVICE NAME:

2.1 Magnetic Resonance Diagnostic Device
2.2 Classification Name: System, Nuclear Magnetic Resonance Imaging
2.3 Classification Number: 90LNH
2.4 Trade/Proprietary Name: Version 3.7 Operating System Software
2.5 PREDICATE DEVICE(s):

Hitachi AIRIS with Version 3.4C Operating System Software
Hitachi MRP-7000 with Version 3.3B Operating System Software

# 3.0 DEVICE DESCRIPTION:

3.1 FUNCTION

The AIRIS Operating System Software is revised to Version 3.7 to increase the clinical utility of the AIRIS in the stationary configuration. The MRP-7000 Operating System Software is revised to Version 3.7 to increase the clinical utility of the MRP-7000 in both stationary and mobile configurations.

Version 3.7 Operating System revisions include the addition of Dynamic Scan measurement and analysis, MTC (magnetization transfer contrast) and SSP (sloped slab profile) for 3D TOF MRA, additional 2D TOF, 2D, and 3D SARGE sequences (steady-state acquisition with gradient rephasing), increased SE and FSE Flip Angle range to $60^{\circ} - 120^{\circ}$, addition of Fast IR and High Resolution / High Definition Fast IR image acquisition software, addition of High Resolution / High Definition FSE, three additional adaptive filter image post-processing techniques, and Receiver Coil sensitivity correction image post-processing.

For previous FDA 510(k) submissions, option software such as Fast Spin Echo, 2D and 3D TOF MRA, and Dual Slice, and option software enhancements [modifications requiring the submission of a 510(k) pre-market notification], have traditionally been documented through separate FDA 510(k) pre-market notification submissions; (Cf. K905834, K920441A, K925009, K926397, K934485, K935664, K935671). For the purposes of simplification, current enhancements to option software which would otherwise be documented through separate 510(k) submissions will also be included as part of this 510(k) submission.

281

{2}

## 3.2 SCIENTIFIC CONCEPTS

Magnetic Resonance (MR) is based on the fact that certain atomic nuclei have electromagnetic properties which cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nucleus used in current imaging experiments in magnetic resonance. When placed in a magnetic field, there is a slight net orientation or alignment of these atomic nuclei with the magnetic field. The introduction of a short burst of radiofrequency (RF) excitation of wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the proton's magnetization vector. When the RF excitation is removed, the proton relaxes and returns to its original orientation. The rate of relaxation is exponential, and varies with the character of the proton and its adjacent molecular environment. This reorientation process is characterized by two exponential relaxation times called T1 and T2 which can be measured.

These relaxation events are accompanied by an RF emission or echo which can be measured and used to develop a representation of these emissions on a three dimensional matrix. Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT.

For magnetic fields up to 1.5T, the RF frequencies commonly used range up to 65MHz. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The time-varying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.

## 3.3 PHYSICAL AND PERFORMANCE CHARACTERISTICS

MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.

## 4.0 DEVICE INTENDED USE:

The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

{3}

- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited: Proton
- Diagnostic uses:
- 2D T1- / T2-weighted imaging
- T1, T2, proton density measurements
- MR Angiography
- image processing
- Imaging capabilities:
- 2D, 3D Spin Echo (SE)
- 2D Fast Spin Echo (FSE)
- 2D Inversion Recovery (IR)
- 2D Fast Inversion Recovery (FIR)
- 2D, 3D Gradient Echo (GE)
- 2D, 3D Gradient Echo with Rephasing (GR)
- 2D, 3D Steady state acquisition with rewinded GE (SARGE)
- 2D Dual Slice acquisition (DS)
- MR Angiography (2D TOF, 3D TOF, MTC, SSP, half echo, high resolution/high definition)

## 5.0 DEVICE TECHNOLOGICAL CHARACTERISTICS:

Identical to the Predicate Device.

---

**Source:** [https://fda.innolitics.com/device/K960181](https://fda.innolitics.com/device/K960181)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
