COORDINATE REVISION KNEE SYSTEM

K955884 · Depuy, Inc. · JWH · Mar 13, 1996 · Orthopedic

Device Facts

Record IDK955884
Device NameCOORDINATE REVISION KNEE SYSTEM
ApplicantDepuy, Inc.
Product CodeJWH · Orthopedic
Decision DateMar 13, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DePuy Coordinate II Revision Knee System is indicated for use with bone cement as either a primary or revision tricompartmental knee replacement system.

Device Story

Tricompartmental knee prosthesis; modification of Coordinate Revision Knee System. Design changes: threaded stem extensions for femoral/tibial components; modified central posts to accept threaded stems; cast Co-Cr-Mo tibial trays; increased conformity of tibial insert articular surfaces. Used in orthopedic surgery for total knee arthroplasty; implanted by surgeons. Provides structural replacement for knee joint; restores joint function; benefits patients via improved stability and fixation compared to previous tapered designs.

Clinical Evidence

Bench testing only.

Technological Characteristics

Materials: Cast Co-Cr-Mo (tibial trays). Design: Semi-constrained, cemented, tricompartmental knee prosthesis. Features: Threaded stem extensions for femoral and tibial components; modified central posts; conforming tibial inserts. Classification: 888.3560.

Indications for Use

Indicated for patients requiring primary or revision tricompartmental knee replacement, to be used with bone cement.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0} K955884 # SUMMARY OF SAFETY AND EFFECTIVENESS | **NAME OF FIRM:** | DePuy Inc. 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 | | --- | --- | | **510(k) CONTACT:** | Cheryl Hastings Manager, Pre-clinical Device Evaluation | | **TRADE NAME:** | DePuy Coordinate II Revision Knee System | | **COMMON NAME:** | Knee Prosthesis | | **CLASSIFICATION:** | 888.3560 - Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer | | **DEVICE PRODUCT CODE:** | 87 JWH | | **SUBSTANTIALLY EQUIVALENT DEVICES:** | - DePuy AMK Landmark Revision Knee System and Full Wedge Tibial Tray (renamed the Coordinate Revision Knee System) - Johnson & Johnson P.F.C. Modular Knee System - DePuy Synatomic Variable Fit Tibial Plateau - DePuy AMK Congruency PS Tibial Inserts | ## INTENDED USE AND DEVICE DESCRIPTION: The Coordinate II Revision Knee System is a design modification of the Coordinate Revision Knee System which was previously cleared by FDA for cemented use. The design modifications consist of the following: stem extensions for femoral and tibial components will be threaded rather than tapered; central posts on the femoral components and tibial trays will be modified to accept the threaded stem extensions; all tibial trays will be manufactured from cast Co-Cr-Mo; and the articular surface of the tibial inserts will be modified to be slightly more conforming. The DePuy Coordinate II Revision Knee System is indicated for use with bone cement as either a primary or revision tricompartmental knee replacement system. ## BASIS OF SUBSTANTIAL EQUIVALENCE: The Coordinate II Revision Knee System is similar to the Coordinate Revision Knee System previously cleared by FDA with the exception of the design change listed under DEVICE DESCRIPTION. The material, manufacturing process, intended use (cemented total knee arthroplasty) and the basic design of the Coordinate II Revision Knee System, including sizes of femoral components, tibial trays, and tibial inserts will not be changed from those of the Coordinate Revision Knee System previously cleared by FDA. 000016 105
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...