# BRAUN HEMOSTASIS INTRODUCERS (K955820)

_B.Braun Medical, Inc. · DYB · Mar 25, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K955820

## Device Facts

- **Applicant:** B.Braun Medical, Inc.
- **Product Code:** [DYB](/productcode/DYB.md)
- **Decision Date:** Mar 25, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.

## Device Story

Device consists of 16 and 18 French catheter introducers and vessel dilators; includes silicone gasket to prevent blood backflow. Used by clinicians for percutaneous vascular access. Introducer sheath (FEP with BaSO4) and dilator (high-density polyethylene with bismuth subcarbonate) facilitate catheter placement. Device functions as a mechanical conduit for vascular procedures.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Sheath: FEP with BaSO4. Dilator: High-density polyethylene with bismuth subcarbonate. Gasket: Silicone. Sizes: 16 and 18 French. Lengths: 10-40 cm. Packaging: Tyvek/Polymylar pouch. Sterilization: Not specified.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- Angeion Hemostasis Valve Introducer Kit ([K894446](/device/K894446.md))
- Burron Cardiovascular Angestat ([K931743](/device/K931743.md))
- ANGETEAR Tear-away Introducer Kits ([K871153](/device/K871153.md))
- Fluoro-tip Hemostasis Introducer Set ([K930057](/device/K930057.md))
- Fluoro-tip Hemostasis Introducer Set ([K945396](/device/K945396.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

3
90

# II. SUMMARY OF SAFETY AND EFFECTIVENESS

Applicant Name &amp; Address: B. Braun Medical Inc., Cardiovascular Division, 13000 Hwy 55
Minneapolis, MN 55441-3813 (612) 559-3370

Contact: Barbara Atzenhoefer, Regulatory Affairs Manager

Date Prepared: December 19, 1995

Common or Usual Name: Catheter introducer and Vessel dilator for percutaneous catheterization

Device Classification Name:
Introducer: 21 CFR Reference- 870.1340, Class II Panel- 74 DYB
Dilator: 21 CFR Reference- 870.1310, Class II, Panel- 74 DRE

Substantial Equivalence:

|  510(k) # | Product | Manufacturer  |
| --- | --- | --- |
|  K894446 | Angeion Hemostasis Valve Introducer Kit | B. Braun Cardiovascular - MN  |
|  K931743 | Burron Cardiovascular Angestat | B. Braun Cardiovascular - MN  |
|  K871153 | ANGETEAR Tear-away Introducer Kits | B. Braun Cardiovascular - MN  |
|  K930057 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN  |
|  K945396 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN  |

Device Description:
16 &amp; 18 french versions of the Braun Hemostasis Introducer with silicone gasket, in lengths of 10 - 40 cm.

Statement of Intended Use:
For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.

Summary/Comparison of Technological Characteristics:

|  Characteristic | Braun Hemostasis Introducers | K894446 Hemostasis Introducer | K931743 Angestat | K871153 ANGETEAR  |
| --- | --- | --- | --- | --- |
|  A) Labeling/Pkg  |   |   |   |   |
|  Labels & IFUs | See Attachment B | All Braun Hemostasis Introducer Label formats are identical | All Braun Hemostasis Introducer Label formats are identical | same information provided on pouch and box labels as on hemostasis labels  |
|  Expiration Dating | 2 years | 2 years | 2 years | 4 years  |
|  Pyrogen Free | Yes | Yes | Yes | Yes  |
|  Packaging | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tray w/ tyvek lid w/in a tyvek/polymylar pouch  |
|  B) Intended Use | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature  |
|  C) Physical Char.  |   |   |   |   |
|  Available french sizes | 4-18 french | 5-14 french | 4-15 french | 4-16 french  |
|  Lengths | 10-40 cm | 12-24 cm | 15-32 cm | 16-30 cm  |
|  gasket material | silicone | Natural rubber | silicone | none  |
|  sheath material | FEP with BaSO4 | FEP teflon with BaSO4 | FEP with BaSO4 | polyethylene with bismuth subcarbonate  |
|  dilator material | high density polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | FEP with BaSO4, or polyethylene with bismuth subcarbonate  |
|  D) Anatomical Sites | vasculature | vasculature | vasculature | vasculature  |

---

**Source:** [https://fda.innolitics.com/device/K955820](https://fda.innolitics.com/device/K955820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).