CLINICARE STERILE DISPOSABLE SYRINGE/NEEDLES
Device Facts
| Record ID | K955772 |
|---|---|
| Device Name | CLINICARE STERILE DISPOSABLE SYRINGE/NEEDLES |
| Applicant | Ttk Biomed, Ltd. |
| Product Code | FMI · General Hospital |
| Decision Date | Jul 8, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5570 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Clinicare Sterile Disposable Syringes and Needles are intended to be used for general purpose injections.
Device Story
Sterile disposable syringes and needles for general purpose injections. Device components: polypropylene barrels, thin-walled SS304 stainless steel capillary needles, natural rubber gaskets. Barrels lubricated with Polydimethyl Silicone. Used in clinical settings by healthcare professionals for medication administration or fluid aspiration. Single-use, disposable design.
Clinical Evidence
Bench testing only; compliance with ISO 7886, ISO 7864, ISO 594, ISO 9626, and ISO 6009 standards.
Technological Characteristics
Materials: Polypropylene (barrel), SS304 stainless steel (needle tubing), natural rubber (gasket). Lubricant: Polydimethyl Silicone. Standards: ISO 7886 (syringes), ISO 7864 (needles), ISO 594 (Luer fittings), ISO 9626 (tubing), ISO 6009 (needles). Single-use, sterile, disposable.
Indications for Use
Indicated for general purpose injections in patients requiring hypodermic administration of medication or fluid withdrawal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
- Sterile Disposable Syringe (Becton, Dickinson & Company)
- Sterile Disposable Syringes and Needles (Becton, Dickinson & Company)
- Sterile Disposable Hypodermic Needle (Becton, Dickinson & Company)
Related Devices
- K212207 — Sterile Hypodermic Syringe for Single Use · Shandong Weigao Group Medical Polymer Co., Ltd. · Mar 23, 2023
- K211482 — Disposable Syringe with Needle · Shandong Qinkai Medical Industry Co., Ltd. · Jan 24, 2022
- K212857 — Sterile Syringe for Single Use (with Needle), Sterile Hypodermic Needle for Single Use · Jiangsu Zhiyu Medical Instrument Co, Ltd. · May 9, 2023
- K220204 — Disposabel Syringe with permanently attached needle, Safety Syringe with permanently attached needle · Anhui Tiankang Medical Technology Co., Ltd. · Jul 14, 2022
- K212920 — Sterile Safety Syringe with Needle for Single Use, Sterile Safety Hypodermic Needle for Single Use,Sterile Auto-Disable Syringe with Needle for Single Use · Shandong Weigao Group Medical Polymer Co., Ltd. · Mar 11, 2022