CORDIS BILIARY STENT
Device Facts
| Record ID | K955728 |
|---|---|
| Device Name | CORDIS BILIARY STENT |
| Applicant | Cordis Corp. |
| Product Code | FGE · Gastroenterology, Urology |
| Decision Date | May 8, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5010 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Cordis Biliary Stent intended use is to maintain patency of a bile duct which is obstructed from malignant neoplasms.
Device Story
Helical coil stent; flexible radiopaque 5400 Grade Tantalum; 7.0 mm diameter; 1.5 cm and 3.0 cm lengths. Supplied mounted/crimped on balloon catheter delivery system. Used by physicians in clinical settings to maintain bile duct patency in malignant obstruction cases. Stent expands upon balloon inflation to support ductal walls; prevents occlusion; improves biliary drainage.
Clinical Evidence
In-vitro and in-vivo testing conducted. Biocompatibility testing performed per ISO 10993 Part 1.
Technological Characteristics
Material: 5400 Grade Tantalum (Ta). Form factor: Helical coil stent, 7.0 mm diameter, 1.5/3.0 cm lengths. Delivery: Balloon catheter system. Radiopaque.
Indications for Use
Indicated for patients with bile duct obstruction caused by malignant neoplasms.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
- Johnson & Johnson Interventional Systems Palmaz™ Balloon Expandable Stent (K905720, K911581)
- Schneider Wallstent® Transhepatic Biliary Endoprosthesis (K885180, K896163, K911292, K914277)
- Cook Incorporated Gianturco-Rosch Biliary Z Stents™ (K921191, K903858, K882610)
Related Devices
- K052539 — COOK FORMULA 418 BILIARY STENT SYSTEM · Cook, Inc. · Dec 9, 2005
- K020682 — BARD LUMINEXX 6 FR BILIARY STENT AND DELIVERY SYSTEM · C.R. Bard, Inc. · Apr 2, 2002
- K012056 — CORDIS PALMAZ GENESIS TRANSHEPATIC BILIARY STENT ON SLALOM .018 DELIVERY SYSTEM · Cordis Europa, N.V. · Aug 1, 2001
- K963782 — BARD MEMOTHERM TRANSHEPATIC BILIARY ENDOPROSTHESIS · C.R. Bard, Inc. · May 2, 1997
- K032025 — SENTINOL NITINOL BILLARY STENT SYSTEM · Boston Scientific Corp · Sep 25, 2003