RELSIA SCL-70 ANTIBODY TEST SYSTEM
Device Facts
| Record ID | K955602 |
|---|---|
| Device Name | RELSIA SCL-70 ANTIBODY TEST SYSTEM |
| Applicant | Immuno Concepts, Inc. |
| Product Code | LLL · Immunology |
| Decision Date | Apr 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5100 |
| Device Class | Class 2 |
Intended Use
This test system is for in vitro diagnostic use for the detection of antibodies to nuclear antigen Scl-70 in human serum.
Device Story
RELISA® Scl-70 Antibody Test System; enzyme immunoassay (EIA) for detection of Scl-70 antibodies in human serum. Device utilizes microwells coated with Scl-70 autoantigen. Laboratory-based test; performed by trained technicians. Patient serum sample added to microwells; antibody-antigen binding detected via enzymatic reaction. Output is qualitative/semi-quantitative result indicating presence of Scl-70 antibodies. Results used by clinicians to support diagnosis of autoimmune disorders. Benefits include standardized, specific detection of Scl-70 antibodies compared to broader screening assays.
Clinical Evidence
Bench testing only. Comparison study against predicate RELISA® Screening Assay (K935129) using 177 samples. Results: 37 positive, 3 borderline, 7 borderline, 1 borderline, 129 negative. Assuming borderline as positive: relative sensitivity 97.9%, relative specificity 100%, overall agreement 99.4%.
Technological Characteristics
Enzyme immunoassay (EIA) system. Microwells coated with Scl-70 autoantigen. Includes calibrator serum. Manual or automated plate reader detection. In vitro diagnostic use.
Indications for Use
Indicated for the detection of Scl-70 antibodies in human serum to aid in the diagnosis of systemic sclerosis or related autoimmune conditions.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
- RELISA® ENA Antibody Screening Tests System (K935129)
Related Devices
- K990797 — AUTOSTAT II ANTI-SCL-70 ELISA · Cogent Diagnotics , Ltd. · Apr 28, 1999
- K983921 — AUTOSTAT II ENA SCREEN ELISA · Cogent Diagnotics , Ltd. · Dec 28, 1998
- K970239 — IS-ANTI-SCL-70 TEST SYSTEM (720-280) · Diamedix Corp. · Apr 8, 1997
- K983423 — THE APTUS (AUTOMATED) APPLICATION OF THE SCL-70 ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE · Zeus Scientific, Inc. · Nov 18, 1998
- K113143 — IMMULISA SCI-70 ANTIBODY ELISA · Immco Diagnostics, Inc. · Dec 2, 2011