K955324 · Respironics, Inc. · MNT · Dec 20, 1996 · Anesthesiology
Device Facts
Record ID
K955324
Device Name
BIPAP S/T-D 30 SYSTEM
Applicant
Respironics, Inc.
Product Code
MNT · Anesthesiology
Decision Date
Dec 20, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BiPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.
Device Story
Ventilation support system; augments spontaneous breathing; not for life support. Inputs: patient flow demand; internal blower generates airway pressure. Modes: CPAP, Spontaneous (S), Spontaneous/Timed (S/T), Timed (T). Operation: senses flow changes to trigger IPAP/EPAP transitions; adjusts flow to maintain set pressures; compensates for circuit leaks. Output: positive pressure ventilation via nasal/full-face mask or mouthpiece. Used in hospital/institutional settings; operated by clinicians. Benefits: noninvasive alternative to intubation for acute respiratory conditions. Supplemental oxygen can be bled into circuit; requires external monitoring.
Clinical Evidence
Prospective, randomized, controlled study (n=31) and prospective, non-randomized study (n=81) in adult patients with acute respiratory failure. Primary endpoint: success of noninvasive ventilation as alternative to intubation. Results showed statistical significance (p<0.05) supporting use as alternative to intubation. Adverse effects included nasal bridge abrasions, managed with skin barriers/foam spacers.
Technological Characteristics
Electromechanically controlled ventilation system; internal blower assembly; pressure range 4-30 cm H2O. Connectivity: standalone. Standards: ASTM F 1100-90 (referenced for comparison), CSA C22.2 125, UL544, IEC 68-2-6/27/34, IEC 601-1-2, MIL-STD-461D/462D, FCC Part 15.J.
Indications for Use
Indicated for assisted ventilation of spontaneously breathing adult patients with obstructive sleep apnea, respiratory failure, or respiratory insufficiency in hospital or institutional settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Healthdyne Quantum Pressure Support Ventilator (K950331)
Reference Devices
Puritan Bennett 7200
Submission Summary (Full Text)
{0}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
# Section 12 - 510(k) Summary

Date: September 3, 1996
Manufacturer: Respironics Inc.
1001 Murry Ridge Drive
Murrysville, PA 15668
Phone: (412) 733-0200
Official Contact: Francis X. Dobscha
Manager, Regulatory Affairs
Device Name: BiPAP® S/T-D 30
Common Name: Continuous Ventilator, Passive Exhalation Port, Critical Care
Classification: 868.5895, 73 MNT
Predicate Device: Respironics BiPAP® (K883825)
Respironics BiPAP® S/T-D System (K951264)
Detachable Control Panel (K905540)
Siemens Servo 900 (K811102)
Healthdyne Quantum™ Pressure Support Ventilator (K950331)
Page 1
{1}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
# Description of the BiPAP S/T-D 30 System
The BiPAP S/T-D 30 System is a ventilation support system designed to augment a patient's ability to breathe on a spontaneous basis. It is not intended for life support situations. The BiPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.
The BiPAP S/T-D offers four modes of operation:
- Continuous Positive Airway Pressure (CPAP)
- Spontaneous (S) mode. The unit cycles between Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) in response to the patient rate. The difference between IPAP and EPAP is the delivered pressure support level.
- Spontaneous/Timed (S/T) mode. The unit cycles between Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) in response to the patient rate. The difference between IPAP and EPAP is the delivered pressure support level.
- Timed (T) mode. The unit cycles between Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) levels based solely on set breaths per minute (BPM) and %IPAP Time Controls.
The BiPAP S/T-D 30 System provides ventilation support by applying a positive pressure (IPAP) during the inspiratory segment of a breathing cycle, and subsequently cycling to a lower pressure level (EPAP) during the expiratory segment of the cycle for spontaneous breaths. The transition point between inspiration and expiration is sensed by the BiPAP System as a function of a decrease in patient flow demand. The BiPAP S/T-D 30 System also senses the transition point between expiration and inspiration. As the patient begins to inhale, the BiPAP System senses the increase in flow and transitions from EPAP to IPAP. These points are called the trigger thresholds.
Levels of IPAP and EPAP can be adjusted by the clinician for individual patient requirements. The IPAP control setting is prohibited from being greater than EPAP by design of the control system. IPAP settings are always set equal to or greater than EPAP. By setting IPAP and EPAP levels the same, CPAP therapy is applied. If the EPAP level is set greater than the ambient atmospheric pressure (greater than 0), then a level of positive end expiratory pressure (PEEP) is applied to the patient's airway. The BiPAP S/T-D 30 System is a prescription device. IPAP and EPAP levels are determined by a physician or by a clinician under the direction of a physician.
The BiPAP S/T-D 30 System monitors delivered pressure and adjusts flow as applied to the patient circuit to maintain set IPAP and EPAP pressures. The System is also capable of compensating for most leaks around the patient
Page 2
{2}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
interface. Patient interfaces include specified nasal masks, full face masks covering both the mouth and nose, and a mouthpiece that is used to apply pressures orally.
Each patient interface system uses an exhalation leak to direct the patient's exhaled gasses to the atmosphere. This leak is incorporated either as an accessory valve or bleed port.
## Intended Uses of the BiPAP S/T-D 30 System
The BiPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.
The BiPAP S/T-D 30 System can be used to administer oxygen by using an external oxygen tank and bleeding oxygen into the patient circuit. This therapy is not metered or otherwise regulated. Patient monitoring must be performed to ensure effective administration of oxygen therapy.
The BiPAP S/T-D 30 System is similar to the BiPAP described in K883825, and other continuous ventilators, such as the Siemens Servo 900 and the Puritan Bennett 7200. However, there are specific differences between the BiPAP S/T-D 30 and these traditional life-support ventilators as noted below.
## Comparison of the BiPAP to Predicate Devices
The Respironics BiPAP S/T-D 30 System is equivalent to the current BiPAP S/T-D System as described in K951264. The major difference between these two devices is that the BiPAP S/T-D System is limited to a maximum working pressure of 20cm H₂O while the BiPAP S/T-D 30 System can be set for a maximum pressure of 30cm H₂O.
The major technical difference between the BiPAP S/T-D 30 System and most continuous ventilators is in the technical approach used to achieve patient support ventilation. Like the Healthdyne Quantum PSV, the BiPAP S/T-D 30 System has a self-contained, internal blower assembly to generate airway pressure. Most comparable predicate devices utilize an external pressure source.
The BiPAP S/T-D 30 System is an electromechanically controlled system that controls the pressure applied to the patient based on patient demand, while the flow is adjusted to maintain a set inspiratory and expiratory pressure. Changes between inspiratory phases and expiratory phases are determined by changes in patient flow patterns.
The BiPAP S/T-D 30 System has an electronic mechanism for compensating for minor leaks around the patient's mask or in the patient circuit.
Page 3
{3}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
The pressure supplied by the BiPAP S/T-D 30 System is constant within +0.8 to -1.5 cm H₂O in the flow ranges of -60 to 100 Liters per Minute (LPM). (The -60 LPM designation represents patient exhalation into the BiPAP S/T-D 30 patient circuit.) Most comparable continuous ventilators require a manual operator adjustment to the flow rate control to compensate for leaks in the masks and other areas of the patient circuit.
There are differences in operational characteristics between the BiPAP S/T-D 30 system and predicate devices that are fully compliant with ASTM F 1100-90, the voluntary standard for critical care ventilators:
- The BiPAP S/T-D 30 System provides continuous positive airway pressure (CPAP) and positive pressure ventilation and is indicated for assisted ventilation of adults. This system does not provide mandatory ventilation with guaranteed tidal volume delivery. Patients requiring mandatory ventilation at predetermined tidal volumes are not candidates for pressure support or pressure-limited ventilation.
- The BiPAP S/T-D 30 System requires an intentional leak port, instead of an actively controlled exhalation valve to remove exhaled air from the circuit. Therefore, specific masks and circuits using an intentional leak port are required for normal operation. The pressurized air from the BiPAP S/T-D 30 System causes a continuous flow of air to exhaust from the leak port to flush the exhaled air from the circuit. The machine should be turned on and the intentional leak port should be checked before the mask is applied.
- Pressurized air from the BiPAP S/T-D 30 System causes a continuous flow of air to exhaust from the leak port to flush exhaled air from the circuit. The ability to completely exhaust exhaled air from the circuit is dependent upon the EPAP setting and I:E ratio. At low EPAP settings or with short expiratory times (e.g., high breathing rates) the leak rate through the intentional leak port may be inadequate to clear all exhaled gas from the circuit. Some rebreathing may occur.
- A controlled method for setting and measuring inspired oxygen concentrations cannot be accomplished while using the BiPAP S/T-D 30 System. Instead, oxygen administration is accomplished by using an external low flow oxygen source that is titrated into the mask. At a fixed flow rate of supplemental oxygen, the inspired oxygen concentration will vary, depending on the IPAP and EPAP settings, patient breathing pattern, mask fit and the leak rate. Continuous patient monitoring is recommended while administering oxygen.
- The Airway Pressure Monitor is recommended for use while operating the BiPAP S/T-D 30 System. The Airway Pressure Monitor will monitor proximal airway pressures and provide audible and visual alerts for high and low pressure conditions, and alert the user to a low internal battery,
Page 4
{4}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
and incidents when the monitor's power switch is inadvertently turned OFF while the BiPAP S/T-D 30 System is in use. Each alarm function should be checked before use by performing the Performance Verification procedure described in Chapter 8 of the Clinical Manual.
- The BiPAP alarm system differs from some ventilators in the following respects:
> Although the low pressure alarm on the Airway Pressure Monitor will sound in the event of a power failure, the BiPAP System does not have an independent loss-of-power alarm.
> The Airway Pressure Monitor cannot detect complete or partial obstruction of the exhalation port.
> The Airway Pressure Monitor will sound if pressures exceed the user-set threshold, but there is no mechanism to relieve excess pressures generated in the event of a ventilator control system failure.
> The BiPAP's alarms provide indication only while the alarm condition exists. There is no mechanism to retain the alarm message after a transient alarm condition returns to normal.
- Unlike most ventilators, the oxygen inlet (located on the mask) will accept general purpose tubing. Caution must be exercised to ensure that only oxygen is connected to the oxygen inlet.
- The BiPAP System can be set to extreme inverse I:E ratios. Use of such ratios may have adverse physiological consequences and are not recommended.
The BiPAP Systems have been functionally tested to meet their published specifications for applied pressure, patient airflow, under varying environmental conditions. Specifications that were met include:
| | BiPAP S/T-D 30 System | Environmental Conditions |
| --- | --- | --- |
| Pressure Regulation/Flow Levels | +.8, -1.5 cm H₂O above setpoint over flow rates of -60 to 100 LPM | Nominal 115 VAC
132 VAC. 100 VAC,
72-75° F Nominal
105° F High |
| Control Accuracy, IPAP and EPAP Settings | ±2 cm H₂O of setting in the range of 4 to 30 cm H₂O | Nominal temperature and input voltage |
Page 5
{5}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
| Control Setting Accuracy, Breaths Per Minute (BPM) | ±2 BPM over range of 4 to 30 BPM, or ±10 of setpoint, whichever is greater | Nominal temperature and input voltage |
| --- | --- | --- |
| Control Setting Accuracy %IPAP | ±10% of setpoint measured at 90% and 10% IPAP settings | Nominal temperature and input voltage |
Electromagnetic environmental testing and shock / vibration testing were conducted by independent laboratories to the following National and International Standards for electrical safety for medical devices:
- Canadian Standards Association (CSA) C22.2 125, as tested by CSA.
- UL544 Medical and Dental Equipment Standards, as tested by the City of LA Test Labs.
- German Postal Law Requirements for Conducted and Radiated Emissions per VFG 234/1991, as tested by EMACO for TüV.
- Shock and Vibration testing per IEC 68-2-6 (sinusoidal vibration), IEC 68-2-27 Shock, and IEC 68-34 Random Wide Band Vibration, as tested by East West Labs.
- Shock and vibration for the final packaged configuration per National Safe Transit Association Test Labs.
- Electromagnetic compatibility testing to IEC 601-1-2 for medical electrical equipment. This is a standard that includes IEC 801-1, -2, -3 and CISPIR 14 requirements identified in the FDA Reviewer's Guidance Document. The BiPAP S/T-D 30 System was also tested to MIL-STD-461D and MIL-STD-462D for magnetic field susceptibility. EMI and magnetic field testing was conducted by Canadian Standards Association.
- FCC Class B, Part 15.J Emissions.
Results of the performance testing indicate that the BiPAP S/T-D 30 System will operate within its specifications.
Results of environmental testing indicate that the BiPAP S/T-D 30 System will continue to function after being subjected to shock and vibration, and electromagnetic interference.
Radiation emissions meet the requirement of FCC Class B, Part 15.J.
Page 6
{6}
Premarket Notification, Part A
Section 16, 510(k) Summary
BiPAP S/T-D System
## Clinical Study Summary
Clinical data are presented that verify the suitability of BiPAP Systems for use in treating adult patients with obstructive sleep apnea, respiratory insufficiency, or respiratory failure. These data indicate that the BiPAP S/T-D 30 System can be used in critical care situations with these patients to manage their acute respiratory condition. Data and analysis indicate with a high confidence (p<0.05) that there is a statistical basis to show that the BiPAP S/T-D 30 system can be used as an alternative to intubation in selected patients.
The clinical data includes 31 adult patients with acute respiratory failure who were enrolled in prospective, randomized, controlled clinical studies to determine, as a primary outcome measure, if noninvasive ventilation using the BiPAP System can successfully treat these patients. Additional data concerning 81 patients enrolled in prospective, non-randomized studies were included to substantiate the conclusions drawn from the controlled studies.
## Adverse Effects
Adverse effects identified during the clinical studies include abrasions on the bridge of the nose due to tight fitting masks. These were resolved by readjusting the mask, using foam spacers, or applying a skin barrier.
There was one report of a patient who had respiratory failure as a result of being taken off of the BiPAP S/T-D 30 System to enable transport to another procedure. It is advised that an alternate method of ventilation be available when patients are transported.
## Conclusions
The technical data presented indicate that the BiPAP S/T-D System is compatible with typical clinical or institutional environments. The BiPAP S/T-D 30 has been tested to meet technical safety and performance specifications for medical equipment as established by Canadian Standards Association and IEC.
Clinical data demonstrate that the BiPAP S/T-D 30 System can be used in the hospital environment and other institutional settings to provide ventilation to spontaneously breathing adult patients who suffer from obstructive sleep apnea, are in acute respiratory failure, or have acute respiratory insufficiency.
Page 7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.