HEMOGLOBIN ALC (HBA1C) & TOTAL ASSAYS FOR THE TECHNICON RA/OPERA SYSTEMS IN-VITRO DIAGNOSTIC SYSTEMS
Device Facts
| Record ID | K955087 |
|---|---|
| Device Name | HEMOGLOBIN ALC (HBA1C) & TOTAL ASSAYS FOR THE TECHNICON RA/OPERA SYSTEMS IN-VITRO DIAGNOSTIC SYSTEMS |
| Applicant | Bayer Corp. |
| Product Code | LCP · Hematology |
| Decision Date | Jun 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.7470 |
| Device Class | Class 2 |
AI Performance
| Output | Algorithm | Acceptance | Observed | Dev DS | Dev Readers | Test DS | Test Readers |
|---|---|---|---|---|---|---|---|
| Hemoglobin A1c | — | — | Precision: 2.0-3.1% CV; Correlation: y = 1.09x - 0.39 (r=0.99); Analytical Range: 2.5-18% | — | — | — | — |
| Total Hemoglobin | — | — | Precision: 0.6-0.9% CV; Correlation: y = 0.96x + 0.13 (r=0.988); Analytical Range: 7-23 g/dL | — | — | — | — |
Indications for Use
This in vitro diagnostic method is intended to quantitatively measure Hemoglobin A1c (HbA1c) and Total Hemoglobin (Thb) in human blood on the Technicon RA family/opeRA instrument systems. Measurements obtained through this procedure can be used for monitoring the long-term care of persons with diabetes.
Device Story
In vitro diagnostic assay for Technicon RA/opeRA instrument systems; measures HbA1c and Total Hemoglobin in human blood samples. Used in clinical laboratory settings by trained personnel. Provides quantitative results to clinicians for monitoring long-term glycemic control in diabetic patients. System utilizes automated photometric analysis to determine hemoglobin concentrations; results assist in clinical decision-making regarding diabetes management.
Clinical Evidence
Bench testing only. Comparison study (n=154) against Ames DCA 2000 showed correlation r=0.99, y=1.09x-0.39. Precision studies demonstrated total CV% between 2.0% and 3.1% across HbA1c ranges of 5.7-11.6%. Total Hemoglobin performance compared to NCCLS reference method (n=50) showed r=0.988.
Technological Characteristics
In vitro diagnostic assay reagents for use on Technicon RA/opeRA automated photometric instrument systems. Analytical range: HbA1c 2.5-18%, Total Hemoglobin 7-23 g/dL.
Indications for Use
Indicated for quantitative measurement of HbA1c and Total Hemoglobin in human blood for monitoring long-term care of patients with diabetes.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
Predicate Devices
- Ames DCA 2000 Hemoglobin method
Reference Devices
- NCCLS Reference Method (Document H15-A2, Vol 14, No. 6, May 1994)