HEMOGLOBIN ALC (HBA1C) & TOTAL ASSAYS FOR THE TECHNICON RA/OPERA SYSTEMS IN-VITRO DIAGNOSTIC SYSTEMS

K955087 · Bayer Corp. · LCP · Jun 13, 1996 · Hematology

Device Facts

Record IDK955087
Device NameHEMOGLOBIN ALC (HBA1C) & TOTAL ASSAYS FOR THE TECHNICON RA/OPERA SYSTEMS IN-VITRO DIAGNOSTIC SYSTEMS
ApplicantBayer Corp.
Product CodeLCP · Hematology
Decision DateJun 13, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

This in vitro diagnostic method is intended to quantitatively measure Hemoglobin A1c (HbA1c) and Total Hemoglobin (Thb) in human blood on the Technicon RA family/opeRA instrument systems. Measurements obtained through this procedure can be used for monitoring the long-term care of persons with diabetes.

Device Story

In vitro diagnostic assay for Technicon RA/opeRA instrument systems; measures HbA1c and Total Hemoglobin in human blood samples. Used in clinical laboratory settings by trained personnel. Provides quantitative results to clinicians for monitoring long-term glycemic control in diabetic patients. System utilizes automated photometric analysis to determine hemoglobin concentrations; results assist in clinical decision-making regarding diabetes management.

Clinical Evidence

Bench testing only. Comparison study (n=154) against Ames DCA 2000 showed correlation r=0.99, y=1.09x-0.39. Precision studies demonstrated total CV% between 2.0% and 3.1% across HbA1c ranges of 5.7-11.6%. Total Hemoglobin performance compared to NCCLS reference method (n=50) showed r=0.988.

Technological Characteristics

In vitro diagnostic assay reagents for use on Technicon RA/opeRA automated photometric instrument systems. Analytical range: HbA1c 2.5-18%, Total Hemoglobin 7-23 g/dL.

Indications for Use

Indicated for quantitative measurement of HbA1c and Total Hemoglobin in human blood for monitoring long-term care of patients with diabetes.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K955087 Enclosure 8 JUN 13 1996 Attachment 3 # SUMMARY OF SAFETY AND EFFECTIVENESS ## Hemoglobin A1c (%) Listed below is a comparison of the performance between the Technicon RA®/opeRA™ Systems Hemoglobin A1c/Total Hemoglobin assays and the similar device assay system that was granted FDA clearance of substantial equivalence (Ames DCA 2000® Hemoglobin method). The information provided in this summary was extracted from the Technicon RA®/opeRA™ Systems Hemoglobin A1c/Total Hemoglobin method sheet and the Ames DCA 2000® Hemoglobin A1c method sheet (Attachments 1 & 2). ## INTENDED USE This *in vitro* diagnostic method is intended to quantitatively measure Hemoglobin A1c (HbA1c) and Total Hemoglobin (Thb) in human blood on the Technicon RA family/opeRA instrument systems. Measurements obtained through this procedure can be used for monitoring the long-term care of persons with diabetes. | METHOD | RA/opeRA HbA1c | | DCA 2000 HbA1c | | | --- | --- | --- | --- | --- | | Precision | Mean % | Total CV% | Mean % | Total CV % | | | 5.7 | 3.1 | 5.2 | 3.5-5.6 | | | 8.9 | 2.6 | 8.2 | 2.4-4.9 | | | 11.6 | 2.0 | 12.0 | 3.3-4.9 | | Analytical Range | 2.5 - 18% | | 2.5 - 14.0% | | | Correlation y = RA/opeRA x = DCA 2000 | y = 1.09 x -0.39 n = 154 r = 0.99 Syx = 0.36 | | | | | Expected Values | 4.4 - 6.0 % | | 4.5 - 5.7% | | updated 5/21/96,gm. {1} Attachment 3 # SUMMARY OF SAFETY AND EFFECTIVENESS ## Total Hemoglobin (g/dl) | METHOD | RA/opeRA, Thb | NCCLS Reference Method, Thb Document H15 - A2, Vol 14, No. 6, May 1994 | | --- | --- | --- | | Precision | Mean g/dL Within-Run CV% | | | | 10.6 | 0.6 | | | 15.3 | 0.7 | | | 18.6 | 0.9 | | Analytical Range | 7 - 23 g/dL | | | Correlation | y = RA/opeRA x = NCCLS method | y = 0.96 x 0.13 n = 50 r = 0.988 Syx = 0.65 | | Expected Values | 12.3 - 17.9 g/dL | | updated 5/21/96,gm.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...