QUICKELS QS 100 SYSTEM

K954773 · Ferguson Medical · DRX · Jul 11, 1996 · Cardiovascular

Device Facts

Record IDK954773
Device NameQUICKELS QS 100 SYSTEM
ApplicantFerguson Medical
Product CodeDRX · Cardiovascular
Decision DateJul 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2360
Device ClassClass 2

Intended Use

The Quickels QS 100 device is intended for use as a device to provide a firm and reliable attachment of an AgCl skin surface electrode to a patient or subject being measured during standard electrocardiography workups. The electrodes can be reused.

Device Story

Quickels QS 100 system; suction-based electrode attachment device for standard electrocardiography. Replaces adhesive methods for AgCl electrodes. System utilizes suction pump to maintain electrode contact with patient skin; includes pressure regulation to prevent excessive/insufficient negative pressure; filtration mechanism prevents system contamination. Used in clinical settings by healthcare providers to facilitate ECG signal acquisition. Benefits include reliable electrode placement and reusability.

Clinical Evidence

Bench testing only.

Technological Characteristics

Suction pump system; AgCl electrodes; pressure regulation mechanism; filtration system for contamination control. Non-electronic/mechanical suction principle.

Indications for Use

Indicated for patients undergoing standard electrocardiography workups requiring AgCl skin surface electrode attachment.

Regulatory Classification

Identification

An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FERGUSON MEDICAL JUL 11 1996 916-342-4133 FAX: 916-343-4541 # K954773 20 September 1995 # 510(k) SUMMARY The 510 (k) summary information required by 21 CFR 807.92 is as follows: A. Classification name: Unknown Common/usual name: Suction electrode system Proprietary name: Quickels QS 100 System B. Substantial equivalence: Welsh cup electrode (preamendment device) and Niess Applicard System (K864260). C. Device description: The Quickels QS 100 System is a device that allows reusable AgCl electrodes to be attached to the skin surface using Suction rather than an adhesive. D. Intended use: The Quickels QS 100 device is intended for use as a device to provide a firm and reliable attachment of an AgCl skin surface electrode to a patient or subject being measured during standard electrocardiography workups. The electrodes can be reused. E. Technological characteristics: The proposed Quickels QS 100 device utilizes a suction pump to obtain the required suction to hold the electrode in place. The suction is regulated against too much or too little negative pressure. Contaminants are prevented from entering the system by a filtration mechanism. 3407 Bay Avenue • Chico, California 95926 • USA {1} The proposed device is similar to predicates in that all utilize suction in order to hold the electrode in place on the patient's skin, and also in that the electrodes are reusable. The proposed device is different in predicates in that it utilizes a direct dynamic suction with a pumping mechanism whereas predicates utilize indirect or static suction methods. Submitted, FERGUSON MEDICAL Establishment Registration Number 2937794 Frank Ferguson Official Correspondent
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...