NIPRO ARTERIAL & VENOUS BLOOD TUBING SET FOR HEMODIALYSIS

K954676 · Nissho Corp. · KOC · Mar 4, 1997 · Gastroenterology, Urology

Device Facts

Record IDK954676
Device NameNIPRO ARTERIAL & VENOUS BLOOD TUBING SET FOR HEMODIALYSIS
ApplicantNissho Corp.
Product CodeKOC · Gastroenterology, Urology
Decision DateMar 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2

Intended Use

The blood tubing sets are used to transport blood or fluids to a patient's vascular access device from a hemodialyzer. The reverse is also true in that the tubing set is used as to transport blood or fluids from a patient's vascular access device to a hemodialyzer.

Device Story

Disposable blood tubing set for hemodialysis; includes arterial and venous lines, drip chambers, infusion tubing, monitor lines, and ports. Used to transport blood/fluids between patient vascular access and hemodialyzer; facilitates blood pumping, air/debris capture, medication infusion, blood sampling, and pressure monitoring. Operated by clinical staff in a dialysis setting. Connects to dialysis machines via luer locks. Benefits patient by providing a sterile, closed-circuit pathway for extracorporeal blood circulation during dialysis treatment.

Clinical Evidence

Bench testing only. No clinical data. Performance testing included priming volume, pull force at joints, torque force, pump segment endurance, and air leakage (positive and negative pressure). Biocompatibility testing included pyrogenicity, cytotoxicity, acute toxicity, intracutaneous reactivity, sensitization, hemolysis, Ames mutagenicity, and implantation. All results met specifications and were comparable to predicate devices.

Technological Characteristics

Materials: PE, PVC, ABS, POM, PP, PC, PEHD. Components: drip chambers, infusion tubing, monitor lines, ports, pump segments. Connectivity: luer lock connections. Sterilization: sterile, single-use. Configuration: arterial (red) and venous (blue) lines.

Indications for Use

Indicated for patients requiring hemodialysis. The device is a non-implanted blood access accessory used to transport blood between the patient's vascular access and a hemodialyzer. Restricted to sale by or on the order of a physician.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K954676 Page 1 of 5 Nipro NISSHO CORPORATION MAR - 4 1997 7205 Corporate Center Drive Suite 305 Miami, Florida 33126 Tel: (305) 599-7174 Fax: (305) 599-8454 SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO BRANDED DISPOSABLE BLOOD TUBING SET FOR HEMODIALYSIS §807.92 (a)(1) Submitter: Nissho Corporation Contact Person: Louis Candelario General Manager Date of Summary Preparation: September 16, 1996 §807.92 (a)(2) Trade Name: Nipro Arterial and Venous Blood Tubing Set for Hemodialysis Common Name: Blood tubing set Classification Name: Blood access device and accessories (21 CFR 876.5540) §807.92 (a)(3) Legally Marketed Substantially Equivalent Devices: Kawasumi Blood Tubing Line gambro® Blood line set for hemodialysis §807.92 (a)(4) Description of Device: The tubing sets that we intend to market include arterial and venous dialysis blood tubing (nonimplanted blood access device) as described in 21 CFR 876.5540. Various models of blood tubing sets are being manufactured for application with different dialysis machines. Twenty-seven arterial line models (A001 - A021, A026, A029-A032, and A035) and 14 venous line models are described (V600 - V607, V609 - V613, and V616). All components 146 {1} K 95 4676 Page 2 of 5 # page 2 - SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO BRANDED DISPOSABLE BLOOD TUBING SET FOR HEMODIALYSIS of blood tubing, including drip chambers, infusion tubing, monitor lines, ports and segments which are used to pump blood, retain and capture air and blood debris, infuse medications or fluids, sample blood, monitor pressure, and make connections to other devices are included. The materials used for the components include polyethylene (PE), polyvinylchloride (PVC), acrylonitrate butadiene styrene (ABS), polyoxymethylene (POM), polypropylene (PP), polycarbonate (PC) and polyethylene high density (PEHD). §807.92 (a)(5) Intended Use: The blood tubing sets are used to transport blood or fluids to a patient's vascular access device from a hemodialyzer. The reverse is also true in that the tubing set is used as to transport blood or fluids from a patient's vascular access device to a hemodialyzer. §807.92 (a)(6) Comparison of Technical Characteristics: The configuration of the subject device is similar to legally marketed devices from Kawasumi and gambro®. Colored parts (red - arterial and blue- venous) as well as tubing colors are similar. Accessory lines are designed similarly also. Caps are placed at both ends of the blood circuit to maintain sterility. Labeling for the competitor's devices is similar to the subject device. According to the device names on the labeling, the intended use for the competitors' products is similar to that of the subject device. They are arterial and venous blood tubing sets for hemodialysis. All of the devices are labeled as sterile and for single use only. The devices are restricted to sale by or on the order of a physician The materials used to fabricate the tubing sets are similar as well. IR spectra of component materials of the predicate devices were used to identify them as similar to those of the subject device. The conclusion is that the materials of which the subject device is composed are substantially equivalent to those of the legally marketed predicate devices. Results from chemical and biological {2} 954676 Page 3 of 5 # page 3 - SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO BRANDED DISPOSABLE BLOOD TUBING SET FOR HEMODIALYSIS tests demonstrate that Nissho's blood tubing sets are similar to those of gambro® and Kawasumi. §807.92 (b)(1) The following performance and biocompatibility tests were performed on the subject device as well as the legally marketed devices for comparison. ## 1. Priming Volume Test Auxiliary tubing is closed with clamp connectors and pathways are filled with distilled water. The tubing set is weighed and then the water is drained. Empty tubing sets are weighed again. Volume is expressed in milliliters (ml). Observations found to be within acceptable range and tests results were similar to values obtained for Kawasumi and gambro® blood tubing sets. ## 2. Pull Force at Connecting Joints (A tensile strength tester is used.) a. Between the main tubing (or saline line) and connected parts or other tubing. b. Between heparin lines and connected parts. A 50 mm sample of tubing with the connected part or other tubing is cut and set on the apparatus. After disengagement of the test tubing from the part or other tubing, the result is recorded from the printer. The specification is defined as ±10% variance from nominal values mean of outer diameter of the pump segment. The subject and the legally marketed devices conformed to this specification. ## 3. Torque Force The torque gauge is connected to the tubing which is connected with a female luer lock. The male luer lock cap is turned by hand and torque force read and recorded from dial gauge. ## 4. Endurance Test of the Pump Segment The diameter of the rolling tube is measured before and after pumping at 200 ml/min for 4 hours at 37°C. The percentage of diameter increase is calculated and recorded. The specification is defined as ±10% variance from nominal values mean of outer diameter of the pump segment. The subject and the legally marketed devices conformed to this specification. 148 {3} K9 > 4676 Page 4 of 5 # page 4 - SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO BRANDED DISPOSABLE BLOOD TUBING SET FOR HEMODIALYSIS ## 5. Air Leakage above Atmospheric Pressure Air compressor is set at 2.5 kgf/cm² G and connections of tubing attached appropriately. Sample is then immersed into the water bath and pressure applied at 0.1 kgf/cm² G within seconds. If there is no leakage at the connection, then pressure is gradually increased at 0.1 kgf/cm² G intervals. The pressure at which leakage occurs is recorded. Results were similar to those obtained with legally marketed devices. Leaks occurred at pressures more than 2.0 kgf/cm²G. No clogging was noted. ## 6. Air Leakage below Atmospheric Pressure The access port is punched 6 times in the same location using an 18G needle. The blood line is then attached to the appropriate fitting and the pathway filled with distilled water. The pressure is set 200 mmHg below atmospheric pressure at sea level. The presence or absence of air bubbles in the pathway is noted. No air bubbles occurred in samples tested and results were similar in legally marketed devices for comparison. The following tests were conducted on the finished sterilized device: pyrogenicity, cytotoxicity, acute toxicity (systemic injection) test, intracutaneous reactivity (skin reaction), sensitization, hemolysis, mutagenicity, and implantation. Specifications for these tests are outlined below. ### Biocompatibility Test Specifications | Test | Specification | | --- | --- | | Pyrogenicity | Nonpyrogenic | | Cytotoxicity (Elution Test) | No biological reactivity | | Acute Toxicity (systemic injection) | No biological reactivity/death | | Intracutaneous reactivity | No erythema, edema, or necrosis | | Sensitization | Not a sensitizer | | Hemolysis | Not to exceed 10% | | Ames Mutagenicity | Not mutagenic | | Implantation | No hemorrhage, film or encapsulation | Results indicate that the blood tubing sets conform to the specifications set forth. Predicate devices for comparison show similar results when tested using these methods. {4} KCS 46/6 Page 5 of 5 # page 5 - SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO BRANDED DISPOSABLE BLOOD TUBING SET FOR HEMODIALYSIS §807.92 (b)(3) Results from nonclinical tests performed on both the subject and legally marketed devices demonstrate that Nipro Disposable Blood Tubing Set for Hemodialysis is substantially equivalent. Similar results were obtained from tests on subject and predicate devices. 150
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%