K954564 · Gammex, Inc. · IYE · Apr 22, 1996 · Radiology
Device Facts
Record ID
K954564
Device Name
PID-1 PRECISION INTELLIGENT DOSIMETER
Applicant
Gammex, Inc.
Product Code
IYE · Radiology
Decision Date
Apr 22, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Intended Use
This device is intended to be used for precision measurement of ionizing radiation beams.
Device Story
Battery-powered, stand-alone electrometer/dosimeter; measures charge, charge rate (current), dose, and dose rate of ionizing radiation beams. Inputs: ionization chambers and solid-state detectors. Features: automatic time/date logging; manual temperature/pressure entry for measurement correction; data transfer to host computer. Used by medical physicists/technicians for radiation beam calibration. Output: numerical radiation measurements; data logs. Benefit: ensures accurate radiation therapy equipment calibration.
Clinical Evidence
Bench testing only. Device performance verified against calibrated equipment to confirm adherence to product literature specifications.
Technological Characteristics
Battery-powered electrometer; compatible with ionization chambers and solid-state detectors. Includes data logging and host computer interface. Stand-alone form factor.
Indications for Use
Indicated for precision measurement of ionizing radiation beams in clinical or laboratory settings for calibration of medical radiation equipment. Not for direct use on patients.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Victoreen Model 530 Electrometer/Dosemeter (K931927)
Related Devices
K150265 — Model 206 Electrometer/Dosimeter · Cnmc Company Incorporated · Sep 3, 2015
K092019 — MODEL 1014 PC ELECTROMETER · Sun Nuclear Corp. · Sep 15, 2009
K982937 — MODEL 580-006 RADIATION THERAPY IONIZATION CHAMBER · Victoreen, Inc. · Nov 19, 1998
K013548 — STANDARD IMAGING IVB 1000 WELL CHAMBER · Standard Imaging, Inc. · Nov 9, 2001
K964952 — VERIDOSE V (FIVE), MODEL NUMBER 37-705 · Victoreen, Inc. · Jul 21, 1997
Submission Summary (Full Text)
{0}
K954564
Gammex
GAMMEX RMI®
# SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.87(h).
## 1) SUBMITTER
Contact: Thomas A. Demke, D.D.S.
Company: Gammex, Inc.
2500 W. Beltline Highway at University Ave.
Middleton, WI 53562-0327
Telephone: (608) 828-7260
Date Summary Prepared: August 15, 1995
## 2) PRODUCT IDENTIFICATION
Brand Name: PID-1 Precision Intelligent Dosimeter
Common Name: Electrometer/Dosimeter
Classification Name: Radiation Therapy Calibration Device
## 3) MARKETED DEVICES
Predicate Device: Victoreen Model 530 Electrometer/Dosemeter (K931927)
## 4) DEVICE DESCRIPTION
The PID-1 Precision Intelligent Dosimeter is a battery-powered, stand-alone, precision electrometer for the calibration of medical radiation beams. It is designed to accommodate different types of ionization chambers and solid state detectors. The PID-1 also incorporates features for logging and transferring measured data to a host computer for analysis.
The PID-1 measures charge, charge rate (current), dose and dose rate. Time and date of measurement are automatically stored with the data. Temperature and pressure data can be entered for automatic correction of measurements.
## 5) INDICATIONS FOR USE
This device is intended to be used for precision measurement of ionizing radiation beams.
Gammex, Inc.
2500 West Beltline Hwy. at University Avenue
P.O. Box 620327
Middleton, WI 53562-0327 U.S.A.
1-608-828-7000
1-800-GAMMEX 1 (426-6391)
FAX 1-608-828-7500
Milwaukee, WI • Middleton, WI • Benicia, CA • Bad Münstereifel, Germany • Nottingham, England
{1}
Summary Of Safety And Effectiveness
PID-1
Page 2
6) COMPARISON WITH PREDICATE
The PID-1 has the same intended use as the Victoreen Model 530 Electrometer/Dosemeter. Both devices allow the operator to measure the charge, charge rate (current), dose (exposure) and dose rate from ionizing radiation beams. These devices are compatible with similar types of radiation probes. The devices are of similar size and weight. Neither device is intended for direct use on patients.
7) SUMMARY OF STUDIES
Tests conducted on calibrated equipment indicate that the PID-1 will measure to the specifications stated in the product literature.
8) CONCLUSIONS
Gammex considers the PID-1 to be substantially equivalent to the Victoreen Model 530 Electrometer/Dosemeter. Comparison of the features indicates that the intended use and effectiveness of the PID-1 do not significantly differ from the predicate device. The PID-1 does not raise any new potential safety risks.
16
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.