The Biocircuits IOS™ T4 cartridge is to be used for the quantitative determination of total thyroxine levels, as part of the Biocircuits IOS™ System. The IOS™ Thyroid Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ thyroid assays.
Device Story
Biocircuits IOS™ T4 Test Cartridge is a competitive immunoassay for quantitative total thyroxine (T4) measurement in serum. Device operates on the IOS™ instrument; uses dried reagents (monoclonal anti-T4 antibody, alkaline phosphatase-labeled T4, ANS, salicylate, substrate) within a plastic cartridge. Sample addition rehydrates reagents; T4 in sample competes with conjugate for antibody binding sites. Enzyme-substrate reaction produces fluorescent signal; fluorescence intensity is inversely proportional to T4 concentration. Instrument performs buffer addition, wash steps, signal detection, and result calculation. Used in clinical laboratories or physician office laboratories; operated by laboratory personnel. Output provided via LCD display and printed report. Assists clinicians in diagnosing thyroid disorders (e.g., hyperthyroidism, hypothyroidism) by monitoring T4 levels.
Clinical Evidence
Bench testing and clinical correlation studies performed. Analytical sensitivity is 1.85 ug/dL. Precision studies (within-day, between-day, total) showed CVs ranging from 4.4% to 21.2%. Accuracy evaluated via spiked recovery (91-100% recovery) and method comparison. Correlation study (n=126) against a fluorescent enzyme immunoassay yielded r=0.937. Physician office laboratory study (n=43) yielded r=0.919 compared to the predicate device. Specificity testing confirmed minimal cross-reactivity with T3, T4 analogs, and common interfering substances (hemoglobin, bilirubin, cholesterol, triglycerides).
Technological Characteristics
Competitive immunoassay; plastic cartridge-based; dried reagents (monoclonal anti-T4 antibody, alkaline phosphatase-labeled T4, ANS, salicylate, substrate). Fluorescence detection principle. Instrument-based automated buffer rehydration and wash steps. Factory-generated calibration. Room temperature storage (15-30°C). Microprocessor-controlled data analysis with LCD and printed output.
Indications for Use
Indicated for the quantitative determination of total thyroxine (T4) levels in human serum to assist in the assessment of thyroid function, including hyperthyroidism and hypothyroidism, in patients of all ages.
Regulatory Classification
Identification
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Stratus T4 test (Dade International)
Submission Summary (Full Text)
{1}
K543837
# 510(k) Summary of Information Respecting Safety and Effectiveness
A. Submitter:
Biocircuits Corporation
1324 Chesapeake Terrace
Sunnyvale, CA 94089
(408) 745-1961
Contact: Sheila Ramerman
Date Prepared: October 7, 1995
- 0 : 295
B. Device Names:
Proprietary Name: Biocircuits IOS™ T4 Test Cartridge
Biocircuits IOS™ Thyroid Controls
Common Names: Reagents for total thyroxine assay
Quality control material (assayed and unassayed)
Classification Name: Total thyroxine test system
c. Legally Marketed Device:
The IOS™ T4 Test Cartridges are substantially equivalent to the Stratus T4 test currently in commercial distribution by Dade International.
D. Device Description:
Thyroxine (T4) is a hormone that is synthesized and stored in the thyroid gland, plays a vital role in growth and maturation, and is essential in the control of metabolism in all bodily organs. More than 99% of T4 is reversibly bound to three plasma proteins in the blood: thyroxine-binding globulin (TBG), thyroxine-binding prealbumin (TBPA), and albumin. The balance of T4 is in the free, unbound state in blood at any one time.
Malfunction of the thyroid gland may result in abnormal concentrations of T4. Increased levels of T4 have been found in hyperthyroidism due to Grave's disease, toxic multinodular goiter, and in acute and subacute thyroiditis. Low levels of T4 have been associated with congenital hypothyroidism, myxedema, chronic thyroiditis (Hashimoto's disease), and with some genetic abnormalities.
Principle 01 the Test:
T4 Test: The IOS™ T4 test is a competitive immunoassay in which T4 in the sample is first released from the carrier proteins by 8-anilino-1-naphthalene sulphonic acid (ANS) and salicylate, and then competes with alkaline phosphatase-labeled T4 (conjugate) for binding sites on immobilized monoclonal anti-T4 antibody. After a short incubation, excess sample and conjugate are washed away. Substrate is added, which reacts with the alkaline phosphatase conjugate bound to antibody. The product of the enzyme-substrate reaction produces a fluorescent signal. The level of fluorescence is directly proportional to the amount of conjugate bound to the surface, and inversely proportional to the amount of T4 present in the patient sample. All reagents necessary to perform the test are dried in the IOS™ cartridge, and are dehydrated by the addition of patient sample by the operator, or by the addition of buffer by the instrument.
52
{2}
**Thyroid Controls**: The use of materials derived from human blood to monitor quality control of clinical chemistry testing in the clinical laboratory has been widely established over the past several years. The Biocircuits IOS™ Thyroid Controls are two levels of blood-based material for use with Biocircuits IOS™ thyroid assay test cartridges.
To run a control, the operator inserts the Thyroid Control Cartridge (packaged with the controls) into the IOS™ instrument. The instrument reads the lot number and ranges of acceptable values for the control solutions from the Control Cartridge barcode, and then ejects the Control Cartridge. The operator then inserts a T4 or T4/TU test cartridge and follows the instrument prompts to identify the control level, apply control solutions, and begin the test sequence. The IOS™ instrument performs the required buffer additions to rehydrate assay reagents and perform wash steps as necessary, reads the fluorescence signal generate,, and calculates and prints the control result just as it would if the cartridge were used to test a patient sample.
# E. Intended Use
The Biocircuits IOS™ T4 cartridge is to be used for the quantitative determination of total thyroxine levels, as part of the Biocircuits IOS™ System.
The IOS™ Thyroid Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ thyroid assays.
# F. Comparison with the Predicate Device:
Table I summarizes the comparative features of both the IOS™ and Stratus T4 assays.
# G. Performance Data:
# 1. T4 Test Cartridges:
Non-clinical testing performed in the manufacturer's laboratories gave the following results:
**Sensitivity**: The analytical sensitivity of the IOS™ T4 assay is defined as the smallest concentration of T4 that can be distinguished from zero. Ten replicates of a zero calibrator were run using IOS™ T4 cartridges. The mean value and standard deviation of the signal were calculated. The sensitivity was obtained by interpolating two standard deviations above the mean signal of the zero calibrator. The sensitivity of the IOS™ T4 assay is 1.85 ug/dL.
# **Specificity**: Cross-Reactivity
The specificity of the monoclonal antibody used in the IOS™ T4 assay was determined by evaluating the cross-reactivity with substances similar in structure to T4 or those known to interfere with the immunological reaction. Results are listed below. Pooled human serum with a value of 8.5 ug/dL T4 was used.
| Compound | Cross-Reactivity |
| --- | --- |
| D-thyroxine (D-T4) | 100% |
| 1-triiodothyronine (L-T3) | <470 |
| 1-triiodothyronine (D-T3) | < 2% |
| 5-diiodothyronine | < 0.1% |
| 5-diiodotyrosine | <0.170 |
| Phenytoin (diphenylhydantoin), at 10,000 ug/dL | < 0.1% |
| Propylthiouracil, at 10,000 ug/dL | <0.170 |
| Phenylbutazone, at 10,000 ug/dL | < 0.1% |
| Salicylic acid, at 50 mg/dL | <0.170 |
| Acetylsalicylic acid (ASA), at 50 mg/dL | <0.170 |
52
{3}
# Specif@~ Interfering Substances
The following substances were tested at the levels specified and found not to significantly affect the T4 result obtained using the IOS™ T4 cartridge. A pool of human serum with a value of 5.8 ug/dL T4 was used.
| Substance | Interference |
| --- | --- |
| Hemoglobin @ 500 mg/dL | 6% |
| Bilirubin (unconjugated) @ 20 mg/dL | 6% |
| Cholesterol @ 500 mg/dL | 1% |
| Triglycerides @ 750 mg/dL | 4% |
Precision: The following results were obtained from a laboratory study performed at the manufacturer for within-day, between-day, and total imprecision:
| Control Level | 1 | 2 | |
| --- | --- | --- | --- |
| Mean (ug/dL T4) | 8.1 | 12.9 | 4.65 |
| SD, overall (ug/dL) | 0.72 | 1.14 | 0.56 |
| % CV, within-day (n=10) | 8.5% | 8.8% | 14.4% |
| % CV, between-day (n=40) | 5.0% | 4.4% | 6.9% |
| % CV, total | 8.9% | 8.8% | 12.1% |
# Accuracy: Spiked Recovery
A normal human serum pool with a known T4 value was spiked with different quantities of T4 and then assayed in duplicate using IOS™ T4 cartridges. The results are as shown:
| Endogenous [T4] (ug/dL) | Added [T4] (ug/dL) | Average Observed (ug/dL) | Average Recovery (%) |
| --- | --- | --- | --- |
| 4.25 | 3.5 | 7.55 | 93% |
| 4.25 | 6.0 | 10.2 | 99% |
| 4.25 | 9.2 | 13.5 | 100% |
| 4.25 | 12.4 | 16.5 | 99% |
| 4.25 | 18.0 | 20.7 | 91% |
# Accuracy: Correlation
A comparison of methods obtained by testing 126 patient samples in the manufacturer's laboratories using the IOS™ T4 cartridges and a commercially available fluorescent enzyme immunoassay gave a correlation coefficient (r) of 0.937, with the line of linear regression described by the equation $$y = 0.2 \cdot 9 + 0.943x$$. The samples tested ranged from 1.8 ug/dL to 24.6 ug/dL T4.
Clinical testing performed at a typical physicians office laboratory gave the following results:
Precision:
| Control Level | 1 | 2 | 3 |
| --- | --- | --- | --- |
| number of replicates | 22 | 24 | 22 |
| Mean (ug/dL T4) | 7.46 | 11.75 | 4.79 |
| SD, overall (ug/dL) | 0.75 | 1.40 | 1.01 |
| % CV, total | 10.1% | 11.9% | 21.2% |
# Accuracy: Correlation
A comparison of methods was performed by users in a typical physicians' office laboratory. A total of 43 patient samples were tested using the IOS™ T4 cartridges in the office laboratory; the samples were split and sent to the manufacturer's laboratory for retesting on both the IOS™ and on the predicate device. These studies gave a correlation coefficient (r) of 0.919 with the predicate d, with the line of regression described by the equation $$y = 1.915 + 0.771x$$. The samples tested ranged from 1.8 ug/dL to 20.6 ug/dL T4.
54
{4}
# **Expected Values:**
In a Preference range study of 125 normal, healthy adults (98 males and 27 females) in the manufacturer's laboratories, the range of T4 values for the IOS™ T4 assay was determined to be 4.6 - 1.3 ug/dL.
T4 levels are generally depressed in hypothyroid patients and generally elevated in hyperthyroid patients. T4 levels may also be elevated in neonates and infants.
Since T4 values may vary with many factors, such as geographic location, local population, diet, etc., these ranges are provided for general informational purposes and may not apply to your specific patient population. We recommend that each laboratory establish its own reference ranges for its patient population.
# 2. Thyroid Controls
The following ranges for the IOS™ Thyroid Controls were determined in studies in the manufacturer's laboratories. To establish the ranges, the controls were tested in a total of 40 cartridges each, over at least 10 days, using three IOS™ instruments. These values only apply to this lot of IOS™ Thyroid Controls. Different lots of Thyroid Controls will likely have slightly different ranges. Your laboratory should establish its own range for these controls over time.
| Control | T4 (ug/dL) | | T-Uptake (%) | |
| --- | --- | --- | --- | --- |
| | Mean | Range | Mean | Range |
| 1 | 8.1 | 6.48-9.72 | 32.4 | 29.25-35.55 |
| 2 | 12.9 | 10.32-15.48 | 39.7 | 36.85-42.55 |
It is self-evident from the data and information presented here that the Biocircuits IOS™ T4 Test Cartridges are as safe, as effective, and perform as well as the Stratus T4 assay in commercial distribution by Dade International.
Attachment: Table I: Assay Comparison
r k.
{5}
TABLE 1
Baxter STRATUS vs. Biocircuits IOS™
Assay Comparison
| ATTRIBUTE | STRATUS T4 | IOS T4 |
| --- | --- | --- |
| Technology | Radial partition immunoassay | Intrusion, 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% of the 100% |
| Assay format | Sequential | Competitive |
| Enzyme label | Alkaline phosphatase | Alkaline phosphatase |
| Substrate | Methylumbelliferyl phosphate | Methylumbelliferyl phosphate |
| Reagents | | |
| Immobilization Medium | Reaction tab | Plastic cartridge |
| Dry | Monoclonal antibody only | Monoclonal antibody, ANS, T4-AP, substrate |
| Wet | 3, loaded by operator at start of each run | 1, continuously on board |
| Delivery | Fully automated | Fully automated |
| Calibration | User-generated | Factory-generated |
| Calibration Stability | 14 days (minimum) | 60 days minimum |
| Storage | Refrigerated (2-8°C) | Room Temperature (15-30°C) |
| Sample | | |
| Type | Serum or plasma | Serum |
| Volume | 0.2 ml (minimum) | 0.033 ml |
| Dilution | Performed by instrument | Performed by instrument |
| Operating environment | 22°-320 C | 15°-300 c |
| Data analysis | Microprocessor-controlled Stored standard curves | Microprocessor-controlled Stored standard curves |
| Data output | LCD display Printed alphanumeric hard copy | LCD display Printed alphanumeric hard copy |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.