SERIES 9100 PLUS AUTOMATED DIFFERENTIAL HEMATOLOGY ANALYZERS
K953598 · Serono Diagnostics, Inc. · GKZ · Mar 11, 1996 · Hematology
Device Facts
Record ID
K953598
Device Name
SERIES 9100 PLUS AUTOMATED DIFFERENTIAL HEMATOLOGY ANALYZERS
Applicant
Serono Diagnostics, Inc.
Product Code
GKZ · Hematology
Decision Date
Mar 11, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
Each of the four models within the Series 9100+ PLUS (i.e., Model 9120, Model 9118, Model 9110, and Model 9100 Rx) is intended for use as a fully automated hematology cell counter for the in vitro diagnostic testing of whole blood specimens. The Series 9100+ PLUS also is intended to flag or identify specimens containing abnormal blood cells.
Device Story
Series 9100+ PLUS comprises four automated hematology analyzer models (9120, 9118, 9110, 9100 Rx) for whole blood analysis. Input methods include whole blood mode, prediluted (fingerstick), platelet-rich plasma, cap piercer, or optional AutoSampler with bar code scanning. Device performs automated dilution and analysis; transforms sample data into white cell, red cell, and platelet histograms. Software compares histogram patterns against pre-set limits to provide interpretive or quantitative reports; alerts users to abnormal cell concentrations or sizes; categorizes suspected abnormalities. Used in clinical laboratory settings; operated by laboratory personnel. Output assists clinicians in identifying hematological abnormalities, supporting diagnostic decision-making. System 9120 offers 20 parameters; other models utilize identical hardware with software-suppressed features. Benefits include standardized, rapid (1.33 min/sample) automated blood cell counting and flagging of potential pathologies.
Clinical Evidence
Bench testing only. Performance data included linearity, within-run and day-to-day precision, accuracy (whole blood mode vs. cap piercer/predilute/AutoSampler modes), and carryover testing. Results demonstrated the device functioned as intended and performed in a substantially equivalent manner to predicate devices.
Technological Characteristics
Automated hematology analyzer; 4 models with shared hardware/electronics. Features internal dilution system, cap piercer, and optional AutoSampler. Connectivity includes bar code scanner. Software-based parameter suppression/upgradability. Uses specific hematology reagents. Measures 20 parameters (Model 9120).
Indications for Use
Indicated for in vitro diagnostic testing of whole blood specimens to perform automated hematology cell counts and to flag or identify specimens containing abnormal blood cells.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
Serono Baker Diagnostics, Inc.’s System 9000+ Interpretive Diff Model Cell Counters (K911626/B)
Serono Baker Diagnostics, Inc.’s System 9000 Automated Cell Counter (K870603)
Submission Summary (Full Text)
{1}
MAR 11 1978
K 11 MAR
# ATTACHMENT B
## 510(K) SUMMARY
{2}
# SERONO DIAGNOSTICS, INC.'S SYSTEM 9100+ PLUS AUTOMATED HEMATOLOGY ANALYZERS
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Serono Diagnostics, Inc.
100 Cascade Drive
Allentown, Pennsylvania 18103
Phone: 610-264-2800
Facsimile: 610-266-4361
Contact Person: Laurence A. Potter
Serono Diagnostics, Inc.
Phone: 610-266-4315
Facsimile: 610-266-4361
Howard M. Holstein, Esq.
Regulatory Counsel for Serono Diagnostics, Inc.
Hogan & Hartson, L.L.P.
555 Thirteenth Street, N.W.
Washington, DC 20004-1109
Phone: 202-637-5813
Facsimile: (202) 637-5910
Date Prepared: July 17, 1995
## Name of Device and Name/Address of Sponsor
Series 9100+ PLUS Automated Differential and Non-Differential Hematology Analyzers
The Series 9100+ PLUS is comprised of the following three differential and one non-differential models:
- Model 9120 Interpretive 5-Part Differential Hematology Analyzer;
- Model 9118 Quantitative 3-Part Differential Hematology Analyzer;
- Model 9110 Quantitative Non-Differential Hematology Analyzer; and
- Model 9100 Rx Quantitative 1-Part Differential Hematology Analyzer
\DC-04050/6-0178650.01
{3}
# Sponsor
Serono Diagnostics, Inc.
100 Cascade Drive
Allentown, Pennsylvania 18103
Phone: 610-264-2800
Facsimile: 610-266-4361
Contact Person: Laurence A. Potter
Serono Diagnostics, Inc.
Phone: 610-266-4315
Facsimile: 610-266-4361
# Common or Usual Name
Automated interpretive differential cell counter (applies to Model 9120)
Automated quantitative differential cell counter (applies to Models 9118 and 9000 Rx)
Automated cell counter (applies to Model 9110)
# Classification Name
Automated differential cell counter (intended to flag or identify specimens containing abnormal blood cells)
Automated cell counter
# Predicate Devices
The Company’s Series 9100+ PLUS covered by this submission is substantially equivalent to other legally marketed automated differential and non-differential blood cell counters. Specifically, the differential hematology analyzers in the Series 9100+ PLUS are substantially equivalent to Serono Baker Diagnostics, Inc.’s System 9000+ Interpretive Diff Model Cell Counters (K911626/B), and the non-differential hematology analyzer in the Series 9100+ PLUS is substantially equivalent to Serono Baker Diagnostics, Inc.’s System 9000 Automated Cell Counter (K870603). (For convenience, these two predicate devices will hereafter be referred to together as the “predicate System 9000+/System 9000” whenever it is appropriate.)
\DC-64050/6-0178650.01
{4}
# Intended Use
Each of the four models within the Series 9100+ PLUS (i.e., Model 9120, Model 9118, Model 9110, and Model 9100 Rx) is intended for use as a fully automated hematology cell counter for the in vitro diagnostic testing of whole blood specimens. The Series 9100+ PLUS also is intended to flag or identify specimens containing abnormal blood cells.
# Technological Characteristics and Substantial Equivalence
## A. Description of the Series 9100+
The Series 9100+ PLUS is comprised of four models of hematology analyzers: Models 9120, 9118, 9110, and 9100 Rx. The most advanced hematology analyzer of the four models (i.e., the model with the greatest number of testing parameters) is the Model 9120, testing a total of 20 parameters. The other models in the Series 9100+ PLUS (i.e., Models 9118, 9110, and 9100 Rx) have exactly the same technological design and electronics as the Model 9120, except that each of these three models lacks one or more of the testing parameters and/or features offered by the Model 9120. This downgrading is achieved in models other than the Model 9120 simply by suppressing or eliminating the software functions that would otherwise allow each of those analyzers to test all of the possible parameters.
Three of the four models in the Series 9100+ PLUS (i.e., Models 9120, 9118, and 9100 Rx) contain a built-in 5-, 3-, or 1-Part Differential software package (respectively) that has the ability to perform interpretive 5-part and quantitative 3- or 1-part white cell histogram differentials. (It should be noted, however, that the non-differential Model 9110 can be upgraded to perform the same 5- or 3- part differential analyses performed by the Models 9120 and 9118, respectively, simply by replacing the relevant software modules in the lower tier model and by utilizing the Series 9100+ PLUS differential reagents instead of the non-differential reagents usually associated with the Model 9110.)
The interpretive report for the Model 9120 and the quantitative reports for the Model 9118 and the Model 9100Rx removes subjectivity and provides possible suggestions of interpretation of the white cell, red cell, and platelet histograms based on a computer comparison of the histogram pattern with a multiplicity of pre-set limits. The interpretive report for the Model 9120 does not provide positive identification of cell types, but it alerts the user to the possible presence of abnormally high or low concentrations of normal cell types, abnormal cell types, and abnormally small or large sizes of normal cell types. Both the interpretive and the quantitative reports also provide one or more messages that categorize suspected abnormalities.
\DOC-64050/6-0178650.01
{5}
The rate of determination for each model in the Series 9100+ PLUS is approximately one sample per 1.33 minutes. This rate includes the time from introduction of the whole blood sample to sample analysis, along with the display and printing of the test results. The system is totally automated with an internal dilution system.
Each model in the Series 9100+ PLUS has a cap piercer module that permits sampling of the blood specimens without removing the stopper from the blood vacuum collection tube. An "AutoSampler" module (denoted by the term "Ax") is optional and allows for the automatic mixing and sampling of multiple EDTA-anticoagulated blood specimens from a batch of closed 5 mL vacuum tubes. The AutoSampler also has a Bar Code Scanner option which enables the automatic identification entry of tubes labeled with a standard bar code. Additionally, a manual Bar Code Scanner is available which functions independently of the the AutoSampler, and the user may add the Bar Code Scanner option to each model of the Series 9100+ PLUS without adding the AutoSampler module.
A blood sample can be introduced to the hematology analyzer by one of five methods: (1) the whole blood mode (through the sample probe); (2) the prediluted (fingerstick) sample mode; (3) the platelet-rich plasma mode; (4) the cap piercer mode; or (5) the "Ax" option. Each model of hematology analyzer within the Series 9100+ PLUS uses the same methods to measure the given parameters. The Series 9100+ PLUS makes use of several reagents and supplies.
## B. Substantial Equivalence of the Series 9100+ PLUS to the Serono Baker Diagnostics, Inc. System 9000+ Interpretive Diff Model Cell Counter and the Serono Baker Diagnostics, Inc. System 9000 Automated Cell Counter
The predicate devices also provide the same or similar functions, characteristics, and accessories as described above for the Series 9100+. Like the Series 9100+, the predicate Serono Baker Diagnostics, Inc. System 9000+ and System 9000 ("System 9000+/System 9000") are intended for use as fully automated hematology cell counters for the in vitro diagnostic testing of whole blood specimens. Like the Series 9100+, the predicate System 9000+/System 9000 also are intended to flag or identify specimens containing abnormal blood cells. Similarly, both the Series 9100+ PLUS and its predicate devices have similar performance specifications and testing results for such analyses as correlation coefficient, linearity, precision (both within-run and day-to-day), and accuracy.
In fact, the Series 9100+ PLUS and its predicates are almost identical in their design, functions, characteristics, and accessories. Along with having the same intended use, both the Series 9100+ PLUS and the predicate System 9000+ offer one model with a 5-part interpretive differential software package and two other models with a 1- or 3-part quantitative differential software package
\DC-64050/6-0178650.01
{6}
(respectively) to perform interpretive 1-, 3-, or 5-part white cell histogram differentials (i.e., Model 9020, Model 9018, and Model 9000 Rx, respectively, for the predicate System 9000+). Likewise, both the Series 9100+ PLUS and the predicate System 9000 offer a non-differential hematology analyzer which measures the same 10 parameters. Both the Series 9100+ PLUS and its predicates have a rate of determination of approximately one sample per minute, a cap piercer function, and an optional AutoSampler module.
Both the Series 9100+ PLUS and the predicates System 9000+/System 9000 have identical cycles of operation, and all offer the same methods for blood sample introduction to the hematology analyzer. The Series 9100+ PLUS and the predicate System 9000+/System 9000 also employ all of the same testing methods to measure each blood sample parameter. Furthermore, the predicate System 9000+/System 9000 utilize the exact same reagents and supplies/accessories as do the hematology analyzers in the Serono Series 9100+.
Although there are some differences between the Series 9100+ PLUS and its predicate System 9000+/System 9000, these differences are minor and raise no new questions of safety and effectiveness. The differences between the Series 9100+ PLUS and the predicate devices include: (1) removal of the on-board printer; (2) replacement of reagent level sensors with a software reagent tracking system; (3) reduction in initial sample volume; (4) expansion of claimed platelet linearity; and (5) addition of a Bar Code Scanner capability. However, these few differences in technological characteristics between the Series 9100+ PLUS and its predicate device are minor and do not present any new issues of safety or effectiveness.
## Performance Data
The Series 9100+ PLUS was subjected to the following performance testing:
- Linearity Testing (in the whole blood mode)
- Precision, Within-Run
- Precision, Day-to-Day
- Accuracy, Whole Blood Mode
- Accuracy, Cap Piercer/Predilute/AutoSampler Modes (through mode-to-mode comparison with sample run in the whole blood mode)
- Carryover Testing
In all instances, the Series 9100+ PLUS functioned as intended. These test results demonstrated that the Series 9100+ PLUS is both effective for its intended use and functions in a substantially equivalent manner to the predicate devices.
\DC-64050/6-0178650.01
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.