Archived bone marrow specimens from clinical patients
Archived clinical specimens were used in a multi-center, blinded, controlled study to determine the sensitivity and specificity of the CEP 8 assay compared to standard cytogenetic analysis.
Archived specimens; Bone marrow; Hematological disorders; Methods comparison
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-center comparative study; Multi-center, blinded, controlled, comparative study
Patients with hematological disorders (AML, MPD, MDS, CML, and other hematological disorders); Sample Size: 368; Number of Sites: 4
Standard cytogenetic analysis
Sensitivity and specificity of CEP 8 assay relative to standard cytogenetic analysis
Indications for Use
The CEP 8 SpectrumOrange DNA Probe is a SpectrumOrange fluorescent labeled DNA probe specific for the centromeric region of chromosome 8. This assay is designed to provide a method for the detection and quantification of chromosome 8 in both interphase nuclei and metaphase spreads by fluorescence in situ hybridization (FISH).
Device Story
CEP 8 SpectrumOrange DNA Probe is a fluorescently labeled DNA probe targeting the centromeric region of chromosome 8. Used in clinical laboratories by cytogeneticists/technicians to detect trisomy 8 via fluorescence in situ hybridization (FISH). Input consists of bone marrow specimens; processed via hybridization to interphase nuclei or metaphase spreads. Output is visual enumeration of fluorescent signals (tri-signaled nuclei/spreads) under a microscope. Results are compared against a 2.2% cutoff to classify specimens as positive or negative for trisomy 8. Provides diagnostic information for hematological disorders; assists clinicians in identifying cytogenetic abnormalities; serves as an alternative or adjunct to standard karyotyping.
Clinical Evidence
Multi-center, blinded, controlled study of 368 archived bone marrow specimens (AML, MPD, MDS, CML, HDNOS). Compared FISH interphase/metaphase analysis to standard cytogenetics. Interphase analysis: 96.03% relative sensitivity, 98.01% relative specificity (cutoff 2.2%). Metaphase analysis: 89.19% relative sensitivity, 91.30% relative specificity (cutoff ≥2 spreads). Correlation coefficient between FISH and cytogenetics was 0.91. Reproducibility studies showed inter-site/observer variation due to visual enumeration subjectivity.
Technological Characteristics
SpectrumOrange fluorescent labeled DNA probe specific to chromosome 8 centromere. Utilizes fluorescence in situ hybridization (FISH) principle. Form factor: liquid reagent kit. Analysis performed via manual microscopic visualization of fluorescent signals. No automated software or electronic connectivity described.
Indications for Use
Indicated for the detection and quantification of chromosome 8 in interphase nuclei and metaphase spreads in patients with hematological disorders, including acute myeloid leukemia (AML), myeloproliferative disorders (MPD), myelodysplastic syndrome (MDS), and chronic myelogenous leukemia (CML).
Regulatory Classification
Identification
An automated FISH enumeration system is a device that consists of an automated scanning microscope, image analysis system, and customized software applications for FISH assays. This device is intended for in vitro diagnostic use with FISH assays as an aid in the detection, counting and classification of cells based on recognition of cellular color, size, and shape, and in the detection and enumeration of FISH signals in interphase nuclei of formalin-fixed, paraffin-embedded human tissue specimens.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.4700 with special controls. The special control guidance document " Class II Special Controls Guidance Document: Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems" is available at www.fda.gov/cdrh/oivd/guidance/1550.pdf.
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Automated Fluorescence*in situ* Hybridization (FISH) Enumeration Systems.” See § 866.1(e) for the availability of this guidance document.
Submission Summary (Full Text)
{1}
NOV 29 1996
K953591
# SUMMARY: SAFETY AND EFFECTIVENESS INFORMATION FOR CEP 8 SpectrumOrange DNA Probe Kit
The CEP 8 SpectrumOrange DNA Probe is a SpectrumOrange fluorescent labeled DNA probe specific for the centromeric region of chromosome 8. This assay is designed to provide a method for the detection and quantification of chromosome 8 in both interphase nuclei and metaphase spreads by fluorescence *in situ* hybridization (FISH).
Standard cytogenetic analysis detects cytogenetic abnormalities such as trisomy 8 by karyotyping metaphase spreads after staining the chromosomes with a dye in cultured tissue cells.
Safety and effectiveness issues relevant to FISH CEP 8 assay may include cross-reactivity, poor sensitivity, poor specificity, or poor reproducibility. The following summarize the specific performance characteristics of CEP 8 assay:
## Analytical Sensitivity and Specificity
### Hybridization Efficiency
In a pilot study, the average percentage of cells with no hybridization signal was 0.22% (s.d.=0.20%) on 35 bone marrow (BM) specimens. In a pivotal study, the average percentage of cells with no hybridization signal was 0.11% (s.d.=0.21%) on 60 BM specimens. Thus, <2% cells with no signal is a realistic standard of acceptance, especially for FISH metaphase analysis.
### Analytical Sensitivity
The analytical sensitivity of the CEP 8 probe was tested in the reproducibility study described below. In that study, the 0% specimen was estimated with a mean of 0.95% (s.d.=0.51%) tri-signaled nuclei and the 5% specimen, 5.37% (s.d.=0.98%). There was little overlap between the 0% and 5% specimens; the upper 95% confidence limit for the 0% specimen was 1.95% and the lower 95% confidence limit for the 5% specimen was 3.45%. Thus, the limit of detection for CEP 8 in interphase cells is estimated to be 4.0%.
### Analytical Specificity
Locus specificity studies were performed with metaphase spreads according to standard Vysis QC protocols. A total of 62 metaphase spreads were examined sequentially by G-banding to identify chromosome 8, followed by FISH. No cross-hybridization to other chromosome loci was observed in any of the 62 cells examined; hybridization was limited to the centromere region of chromosome 8.
Safety and Effectiveness Statement
Page 1 of 4
{2}
Safety and Effectiveness Statement
Page 2 of 4
# Reproducibility
In a pilot study, the reproducibility of CEP 8 interphase analysis for the %tri-signaled cells was assessed for inter-site, inter-day, and inter-observer reproducibility, using one lot of the CEP 8 DNA probe. One normal bone marrow specimen was evaluated for the %tri-signaled cells according to the instructions for signal enumeration in the package insert. Statistically significant site-to-site and observer-to-observer variations were observed, reflecting the subjectivity of the visual enumeration process. The results of classification of slides as positive or negative for trisomy 8 (using a cutoff of 2.2%) was 96% correct (1 of 24 had 2.4% tri-signaled nuclei) for this normal specimen. For the overall variation, the N, mean, SD, and percent CV of the observed %tri-signaled nuclei was 24, 0.99%, 0.57%, and 58%, respectively.
In a pivotal study, the CEP 8 interphase analysis for the %tri-signaled cells was again assessed for inter-site, inter-lot, inter-day and inter-observer reproducibility. One low level (approximately 7%) trisomy 8 bone marrow specimen was evaluated for the %tri-signaled cells according to the instructions for signal enumeration in the package insert. Statistically significant site-to-site and observer-to-observer variations were observed, reflecting the subjectivity of the visual enumeration process. The results of classification of slides as positive or negative for trisomy 8 (using a cutoff of 2.2%) was 100% correct for this low level (7%) trisomy 8 bone marrow specimen. For the overall variation, the N, mean, SD, and percent CV of the observed percentage of tri-signaled nuclei was 24, 7.70%, 1.45%, and 19%, respectively.
# Methods Comparison: Clinical Specimens
A multi-center, blinded, controlled, comparative study was conducted to further define the performance characteristics of the CEP 8 probe. The objective of the study was to determine the sensitivity and specificity of the CEP 8 assay relative to standard cytogenetic analysis, the standard of care. Four laboratories provided 368 archived bone marrow specimens for assay at three investigation sites. Site 1 provided 101 specimens; Site 2 provided 57 specimens; Site 3 provided 130 specimens; Site 4 provided 80 specimens. Specimens from Site 4 were analyzed at Site 3. By standard cytogenetic analysis, 151 of these specimens were classified as positive for trisomy 8; 201 negative for trisomy 8; and 16 ambiguous for trisomy 8 (1 trisomy 8 per 30 metaphases analyzed). Fifteen of the 16 ambiguous cases were selected "purposefully" after study completion. These specimens were derived from patients with one of the following diagnoses.
1. Acute myeloid leukemia (AML): 102 specimens
2. Myeloproliferative disorder (MPD), including polycythemia vera: 44 specimens
3. Myelodysplastic syndrome (MDS): 80 specimens
4. Chronic myelogenous leukemia (CML): 72 specimens
5. Hematological disorder, not otherwise specified (HDNOS): 70 specimens (including hyperproliferative states such as leukemoid reaction, lymphoproliferative disorders or chronic lymphocytic leukemia, without trisomy 8).
At one trial site, approximately 50% of the archived bone marrow specimens failed to produce informative FISH results. Further examination of a subset of these specimens revealed a lack of specimen integrity and it was determined that specimens at this site were stored at 4°C rather than at the recommended temperature of -20°C. The conclusion was that some specimens and/or slide preparations were inadequate.
{3}
Safety and Effectiveness Statement
Page 3 of 4
# FISH Interphase Analysis versus Standard Cytogenetics
From the same multi-center comparative study described above, the analysis of interphase nuclei by FISH compared to standard cytogenetics was performed. Based on the cutoff point of 2.2% tri-signaled nuclei that was validated by the same pivotal clinical study, the relative sensitivity was 96.03% (145/151) [95% C.I. 92.55 - 99.51%] and the relative specificity was 98.01% (197/201) [95% C.I. 96.08 - 99.94%] for the CEP 8 interphase analysis.
Among the 10 discrepant cases, the standard cytogenetic analysis results ranged from 0/20 to 8/22 metaphase cells with trisomy 8; the CEP 8 interphase results of % tri-signaled nuclei ranged from 0.6% to 6.0%. Among the 16 ambiguous cases, the range of % tri-signaled nuclei by CEP 8 interphase analysis was from 0.2% to 2.2%.
From the 368 clinical specimens, the correlation coefficient of trisomy 8 between standard cytogenetic metaphase analysis and CEP 8 interphase analysis was 0.91. The regression coefficient of CEP 8 assay on cytogenetic analysis was 0.71. The following equation describes the plot of CEP 8 assay results on cytogenetic analysis results:
$$
y = 1.2944 + 0.7112x
$$
where: x = percent metaphase spreads with trisomy 8 by standard cytogenetics
y = percent tri-signaled interphase nuclei by CEP 8 analysis
# FISH Metaphase Analysis versus Standard Cytogenetics
A total of 348 cases were included in the comparison of CEP 8 metaphase analysis to standard cytogenetics. Twenty of the 368 cases included in the CEP 8 interphase analysis were excluded due to insufficient number of metaphase spreads for analysis with the CEP 8 assay (a specimen must have ≥ 20 metaphase spreads).
By the rules of standard cytogenetic analysis, a case is declared positive for trisomy 8 if two or more metaphase spreads are trisomic for chromosome 8. Using this cutoff of ≥ 2 tri-signaled metaphases, the relative sensitivity was 89.19% (132/148) [95% C.I. 84.19 - 94.19%] and the relative specificity was 91.30% (168/184) [95% C.I. 87.22 - 95.37%] for the CEP 8 metaphase analysis. Among the 16 ambiguous cases by standard cytogenetic analysis, 15 were negative and 1 was ambiguous for trisomy 8 by CEP 8 metaphase analysis.
From the 348 clinical specimens, the correlation coefficient of trisomy 8 between cytogenetic metaphase analysis and CEP 8 metaphase analysis was 0.91. The regression coefficient of CEP 8 assay on cytogenetic analysis was 0.83. The following equation describes the plot of CEP 8 assay results on cytogenetic analysis results:
$$
y = 1.6979 + 0.8325x
$$
where: x = percent metaphase spreads with trisomy 8 by standard cytogenetics
y = percent tri-signaled metaphase spreads by CEP 8 analysis
{4}
Safety and Effectiveness Statement
Page 4 of 4
## FISH Interphase Analysis versus FISH Metaphase Analysis
In addition to the methods comparison between FISH and standard cytogenetics described above, a comparison between FISH interphase and FISH metaphase was made. There was a 90.8% [(133+183)/348] concordance between FISH interphase and FISH metaphase.
From the 348 clinical specimens, the correlation coefficient of trisomy 8 between CEP 8 interphase analysis and CEP 8 metaphase analyses was 0.95. The regression coefficient of CEP 8 metaphase analysis on interphase analysis was 0.81. The following equation describes the plot of CEP 8 assay metaphase results on CEP 8 interphase analysis results:
$$
y = 1.0942 + 0.8119x
$$
where:
x = percent tri-signaled interphase cells by CEP 8 analysis
y = percent tri-signaled metaphase spreads by CEP 8 analysis
## Conclusions
Consistent performance of CEP 8 is guaranteed by the Vysis Quality Control Procedures. When the CEP 8 SpectrumOrange DNA Probe is used as instructed in the package insert, the above statements describe its performance.
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
JAN 25 2012
Vysis
c/o Dr. Russel K. Enns
Vice President, Regulatory Affairs
3100 Woodcreek Dr.
Downers Grove, IL 60515
Re: k953591
Trade/Device Name: CEP 8 SpectrumOrange DNA Probe Kit
Regulation Number: 21 CFR §866.4700
Regulation Name: Automated fluorescence in situ hybridization (FISH) enumeration systems.
Regulatory Class: Class II
Product Code: OYU, KIR
Dated: October 21, 1996
Received: October 22, 1996
Dear Dr. Enns:
This letter corrects our substantially equivalent letter of November 29, 1996.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related
{6}
Page 2 – Dr. Russel K. Enns
adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Maria M. Chan, Ph.D.
Director
Division of Immunology and Hematology Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.