The Sonoclot Coagulation & Platelet Function Analyzer is an instrument for the in vitro measuring of coagulation and platelet function. The Sonoclot Analyzer performs in vitro whole blood and plasma analyses.
Device Story
Sonoclot Coagulation & Platelet Function Analyzer measures hemostasis in vitro using whole blood or plasma samples. Device utilizes a tubular probe oscillating vertically within a sample; electronic drive and detection circuitry measure mechanical resistance to probe motion as blood progresses through hemostasis stages. Microcomputer processes analog signals into a 'Clot Signal,' generating a qualitative 'Sonoclot Signature' graph and quantitative results for clot formation time (SonACT) and fibrin polymerization rate (Clot Rate). Used in clinical settings to identify coagulopathies, including platelet dysfunction, factor deficiencies, and hyperfibrinolysis. Healthcare providers use printed thermal graphs and quantitative outputs to inform clinical decision-making regarding patient coagulation status.
Clinical Evidence
Bench testing only. Comparative study performed using three current and three original Sonoclot analyzers. Samples included reference viscosity fluids, plasma, and blood. Results demonstrated no performance differences between automatically calculated results from the current analyzer and manually calculated results from the original analyzer.
Technological Characteristics
Mechanical oscillation sensing principle. Construction materials: machined aluminum (side panels, platen, transducer head), stamped stainless steel (top/bottom/front/rear panels), etched zinc (control panel), Dupont Delrin (hinge pin). Power: 100-130Vac or 210-260Vac. Output: thermal printer graph and quantitative digital results. Microcomputer-based signal processing.
Indications for Use
Indicated for in vitro measurement of coagulation and platelet function in whole blood and plasma samples to assess hemostasis, including coagulation, fibrin gel formation, clot retraction, and fibrinolysis, and to identify coagulopathies such as platelet dysfunction, factor deficiencies, anticoagulant effects, hypercoagulable tendencies, and hyperfibrinolysis.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Original Sonoclot Analyzer (marketed prior to May 28, 1976)
Submission Summary (Full Text)
{1}
SIENCO, Inc.
57C BURBANK STREET
BROOMFIELD, COLORADO 80020
(303)469-6450, FAX (303)469-7025
K952560
## 510(k) SUMMARY
SEP - 5 1995
1) Contact Person: Jon Henderson, Vice President, Sienco Inc.
Summary Preparation Date: May 22, 1995
2) Device Trade Name: Sonoclot® Coagulation & Platelet Function Analyzer
Common Name: Coagulation Analyzer
Classification Name: Multipurpose system for in vitro coagulation studies
Classification: Class II, Classification Number 81JPA, 21 CFR 864.5425.
3) Identification of predicate device to which substantial equivalence is being claimed: Original Sonoclot Analyzer marketed prior to May 28, 1976.
(4) and (5) Device Description and Intended Use: The Sonoclot Coagulation & Platelet Function Analyzer is an instrument for the in vitro measuring of coagulation and platelet function. The Sonoclot Analyzer performs in vitro whole blood and plasma analyses.
The Sonoclot Analyzer rapidly provides accurate information on the entire hemostasis process including coagulation, fibrin gel formation, clot retraction (platelet function) and fibrinolysis. The Sonoclot Analyzer generates both a qualitative graph, known as the Sonoclot Signature, and the quantitative results on the clot formation time and the rate of fibrin polymerization for identifying numerous coagulopathies including platelet dysfunction, factor deficiencies, anticoagulant effect, hypercoagulable tendencies and hyperfibrinolysis.
The detection mechanism within the Sonoclot Analyzer responds to mechanical changes that occur within the blood sample. This mechanism consists of a tubular probe that oscillates up and down within a blood sample. The electronic drive and detection circuitry senses the resistance to motion that the probe encounters as the blood sample progresses through the various stages of hemostasis. The resulting analog electronic signal is processed by a microcomputer within the Sonoclot Analyzer and is reported as the Clot Signal. The Clot Signal is printed as a graph on the thermal printer.
The materials used to construct the Sonoclot Analyzer and it's physical dimensions are covered in this summary under item (6).
(6) Comparison of the current to the original Sonoclot: See Table 1 attached. Testing additionally has been completed to demonstrate that the current Sonoclot Analyzer is equivalent to the original Sonoclot Analyzer. Several different reference viscosity fluids, plasma and blood were run side by side on three current and three original analyzers. No difference in performance was observed between the automatically calculated results obtained on the current analyzer and the manually calculated results obtained on the original analyzer. Moreover, the current analyzer eliminates the possibility of the user performing incorrect calculations.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Jon Henderson
Jon Henderson
Vice President
22 MAY 95
Date
441
{2}
Slenco, Inc.
Sonoclot Coagulation & Platelet Function Analyzer
510(K) NOTIFICATION
TABLE I - Table showing the equivalence of the current to the original SONOCLOT Analyzer
| ITEM | Current Sonoclot | Original Sonoclot |
| --- | --- | --- |
| USE | The current Sonoclot Coagulation & Platelet Function Analyzer is an instrument used for measuring coagulation and platelet function. It provides rapid, accurate information on the entire hemostasis process including coagulation, fibrin gel formation, clot retraction (platelet function) and fibrinolysis. The current Sonoclot requires less user intervention and many features have been automated and standardized. | The original Sonoclot Coagulation & Platelet Function Analyzer was an instrument used for measuring coagulation and platelet function. It provided accurate information on the entire hemostasis process including coagulation, fibrin gel formation, clot retraction (platelet function) and fibrinolysis. |
| RESULTS | Provides quantitative results for the clot formation time SonACT (Sonoclot Activated Clotting Time) and the rate of fibrin polymerization (Clot Rate). Provides qualitative graph known of the Sonoclot Signature. | Provided quantitative results for the clot formation time SonACT (Sonoclot Activated Clotting Time) and the rate of fibrin polymerization (Clot Rate). Provided qualitative graph known of the Sonoclot Signature. |
| DESIGN / MATERIALS | Side panels are made from machined aluminum and painted. Bottom, top, front and rear panels are stamped stainless sheet metal. Front angled control panel is made from etched zinc. The platen (heater block) is made from anodized aluminum. The transducer head and hinge assembly is machined anodized aluminum. The transducer head assembly hinge pin is machined Dupont delrin®. | Side panels were made from machined aluminum and painted. Bottom, top, front and rear panels were stamped stainless sheet metal. Front angled control panel was made from etched zinc. The platen (heater block) was made from anodized aluminum. The transducer head and hinge assembly was machined anodized aluminum. The transducer head assembly hinge pin was machined Dupont delrin®. |
| VISUAL IDENTITY | Brushed stainless steel, black side panels and clear anodized aluminum. Front panel graphics and text done in orange and black paint. | Brushed stainless steel, black side panels and clear anodized aluminum. Front panel graphics and text done in orange and black paint. |
| POWER | 100 to 130Vac or 210 to 260Vac user selectable 55 watts The current power requirements have been narrowed to fall within the original ranges. | 90 to 135Vac or 180 to 270Vac factory option only 55 watts |
44
{3}
[LOGO]
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
SEP - 5 1995
Mr. Jon Henderson
Vice President
Sienco, Inc.
557 Unit C Burbank Street
Broomfield, Colorado 80020
Re: K952560
Sonoclot® Coagulation & Platelet Function Analyzer
Regulatory Class: II
Product Code: JPA
Dated: May 22, 1995
Received: June 2, 1995
Dear Mr. Henderson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
]
{4}
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-320) at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A.
Acting Director
Division of Clinical
Laboratory Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.