K951987 · Smith & Nephew Richards, Inc. · JWH · Aug 22, 1995 · Orthopedic
Device Facts
Record ID
K951987
Device Name
GENESIS II TOTAL KNEE SYSTEM
Applicant
Smith & Nephew Richards, Inc.
Product Code
JWH · Orthopedic
Decision Date
Aug 22, 1995
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Genesis II Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3. Failed osteotomies, unicompartmental replacement, or total knee replacement. 4. The posterior stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. The Genesis II Total Knee System is indicated for use only with cement and is a single use device.
Device Story
Genesis II Knee System is a total knee replacement prosthesis; components include femoral components (porous/non-porous), tibial inserts, tibial trays, and patellar components. Used in primary and revision knee arthroplasty; implanted by orthopedic surgeons in clinical/hospital settings. Device replaces damaged joint surfaces to restore function and reduce pain. Femoral components are Co-Cr-Mo; tibial inserts/patellar components are UHMW PE; wedges are Ti-6Al-4V. System is strictly for cemented fixation; single-use. Clinical benefit derived from mechanical restoration of knee joint articulation.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing including device constraint, contact area, tibial tray fatigue, patello-femoral resistance to lateral subluxation, static/fatigue tibial component lock strength, static wedge testing, and porous coating characterization.
Technological Characteristics
Modular knee prosthesis. Materials: Co-Cr-Mo (femoral), UHMW PE (inserts/patellar), Ti-6Al-4V (wedges). Fixation: Cemented only. Components: Porous/non-porous femoral, various tibial inserts/trays, patellar components. Mechanical testing per knee draft guidance.
Indications for Use
Indicated for patients with rheumatoid, post-traumatic, or osteoarthritis, or degenerative arthritis requiring total knee replacement; also for failed prior knee surgeries. Posterior stabilized version indicated for primary/revision surgery with incompetent cruciate ligaments and intact collateral ligaments. Restricted to cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Genesis I Total Knee System (Smith & Nephew Richards Inc.)
Profix Total Knee System (Smith & Nephew Richards Inc.)
NexGen Knee System (Zimmer)
P.F.C. Modular Knee System (Johnson & Johnson)
Omnifit Total Knee System (Osteonics)
Whiteside Ortholoc II Modular Knee System (Dow Corning Wright)
AMK/Coordinate Knee System (Depuy)
Submission Summary (Full Text)
{0}
K9-1(45)
AUG 22 1995
## Summary of Safety and Effectiveness
Smith & Nephew Richards Inc.
Genesis II Knee System
### Substantial Equivalent Information
The Genesis II Knee System is similar to the following knee systems:
1. Genesis I Total Knee System - Smith & Nephew Richards Inc.
2. Profix Total Knee System - Smith & Nephew Richards Inc.
3. NexGen Knee System - Zimmer
4. P.F.C. Modular Knee System - Johnson & Johnson
5. Omnifit Total Knee System - Osteonics
6. Whiteside Ortholoc II Modular Knee System - Dow Corning Wright
7. AMK/Coordinate Knee System - Depuy
All of the devices listed above are indicated for total knee replacement and are similar in design to the Genesis II Knee System. The safety and effectiveness of the Genesis II Knee System is based on the long history of use of these devices in the market place.
### Device Description
The Genesis II Knee System consists of the following components: Porous Cruciate Retaining Femoral Component, Non-Porous Cruciate Retaining Femoral Component, Low Demand Femoral Component, Porous Posterior Stabilized Femoral Component, Non-Porous Posterior Stabilized Femoral Component, Cruciate Retaining Tibial Insert, Dished Tibial Insert, Posterior Stabilized Tibial Insert, Titanium Tibial Tray, Cobalt Chromium Tibial Tray, All Polyethylene Tibial Component, Posterior Femoral Wedge, Anglec Hemi Tibial Wedge, Stepped Hemi Tibial Wedge, Medial-Lateral Tibial Wedge, Biconvex Patellar Component, Resurfacing Patellar Component, All Polyethylene Patellar Component with Flex-Lok Peg, and Revision Biconvex Patellar Component. The femoral components are manufactured from Co-Cr-Mo and available with and without a Co-Cr-Mo porous coating. The porous coating on the femoral component is the same as that on the Genesis I Femoral Component. The tibial inserts, all poly tibial components, patellar components, and Flex Lok Pegs are manufactured from UHMW PE. Wedges are manufactured from Ti-6Al-4V.
G2032251/22
EXHIBIT 6
EXHIBIT 7
EXHIBIT 8
Exhibit 4
Exhibit 5
{1}
# Indications for Use
The Genesis II Knee System is indicated for:
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. The posterior stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
The Genesis II Total Knee System is indicated for use only with cement and is a single use device.
# Mechanical Testing
Mechanical testing was performed according to the requirements in the knee draft guidance document. Following is a list of the testing that supports the safety and effectiveness for the Genesis II Knee System: device constraint, device contact area, tibial tray fatigue, patello-femoral resistance to lateral subluxation, static and fatigue tibial component lock strength testing, static wedge testing, and porous coating characterization. All of the test results indicate that the Genesis II Knee System is capable of withstanding in vivo loading without failure.
G2C3225 /23
Exhibit 5
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
AUG 22 1995
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Thomas L. Craig
Director, Clinical and Regulatory Affairs
Orthopaedic Implant Division
Smith & Nephew Richards Inc.
1450 Brooks Road
Memphis, Tennessee 38116
Re: K951987
Genesis II Total Knee System
Regulatory Class: II
Product Code: JWH
Dated: April 26, 1995
Received: April 27, 1995
Dear Mr. Craig:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone cement." You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations:
1. The thinnest tibial insert available is the nominal "9mm" sized insert, which has a minimum polyethylene thickness under the condyles of 6.7mm, and the thinnest all polyethylene tibial component available is the nominal "9mm" sized component, which has a polyethylene thickness under the condyles of 9.6mm.
2. This device may not be labeled or promoted for non-cemented use.
3. All labeling for this device, including package label and labeling included within the package, must prominently state that the device is intended for cemented use only.
4. Any non-cemented fixation of this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for non-cemented fixation must receive
{3}
Page 2 - Mr. Thomas L. Craig
approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice regarding labeling for your device in accordance with 21 CFR Part 801, promotion, or advertising please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-302) at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Kimber C. Richter, M.D.
Acting Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.