← Product Code [QEY](/productcode/QEY) · K950615

# KAPLAN-SIMPSON INFUSASLEEVE CATHETER (K950615)

_Localmed, Inc. · QEY · Apr 19, 1995 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K950615

## Device Facts

- **Applicant:** Localmed, Inc.
- **Product Code:** [QEY](/productcode/QEY.md)
- **Decision Date:** Apr 19, 1995
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The InfusaSleeve is intended to infuse fluids such as heparanized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with commercially available balloon dilatation catheters to facilitate access (selective and subselective) and regional infusion.

## Device Story

InfusaSleeve catheter; percutaneous infusion device; tracks over standard balloon dilatation catheters. Enables solution delivery (heparinized saline, contrast media, thrombolytic agents) into coronary vasculature without requiring device exchange. Used in clinical settings by physicians; compatible with standard guiding catheters, guide wires, and dilatation catheters. Benefits patient by maintaining access during infusion, reducing procedure time and complexity associated with catheter exchanges.

## Clinical Evidence

Bench testing and animal testing performed. Bench testing included biocompatibility, hub-to-shaft and tip-to-shaft tensile strength, infusion pathway fatigue, burst pressure, infusate flow, and balloon fatigue. Results verified performance equivalent to or exceeding commercially marketed infusion catheters.

## Technological Characteristics

Sterile, nonpyrogenic infusion catheter designed for percutaneous use. Features include infusion pathway for fluid delivery and compatibility with standard guiding catheters, guide wires, and balloon dilatation catheters. Mechanical design facilitates tracking over existing dilatation catheters.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- USCI Dorros Infusion/Probing Catheter
- Target Therapeutics Tracker Infusion Catheters
- SCIMED DISPATCH Coronary Infusion Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

**FDA U.S. FOOD & DRUG**
ADMINISTRATION

February 3, 2022

LocalMed, Inc.
Eleanor Chiu
Director, Regulatory Affairs
1820 Embarcadero Rd.
Palo Alto, California 94303

Re: K950615

Trade/Device Name: Kaplan-Simpson InfusaSleeve Catheter
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy catheter
Regulatory Class: Class II
Product Code: QEY, KRA

Dear Eleanor Chiu:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 19, 1995. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell@FDA.HHS.gov.

Sincerely,

**Gregory W. O'connell -S**
Digitally signed by
Gregory W. O'connell -S
Date: 2022.02.03
14:33:30 -05'00'

Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary
and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{2}

[LOGO]

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

APR 19 1995

Food and Drug Administration

1390 Piccard Drive

Rockville MD 20850

Ms. Eleanor V. Chiu

Director, Regulatory Affairs

LocalMed

1820 Embarcadero Road

Palo Alto, California 94303

Dear Ms. Chiu:

Re: K950615

Kaplan-Simpson

Infusasleeve Catheter

Regulatory Class: II

Product Code: KRA 74

Dated: February 9, 1995

Received: February 10, 1995

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-300) at (301) 594-4639. Other general information on your responsibilities under the Act

{3}

may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D.

Acting Director

Division of Cardiovascular,

Respiratory, and Neurological Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

8

{4}

K

APR 19 1995

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
(Pursuant to Section 12 of the Safe Medical Devices Act of 1990)

1. The trade name of the device is LocalMed's Kaplan-Simpson InfusaSleeve™ Catheter. The common name is Infusion Catheter.
2. The InfusaSleeve is intended to infuse fluids such as heparanized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with commercially available balloon dilatation catheters to facilitate access (selective and subselective) and regional infusion. The contents of the device package are supplied sterile and nonpyrogenic.

The InfusaSleeve is designed for percutaneous use and for tracking over standard dilatation balloon catheters. The InfusaSleeve provides solution delivery without device exchange, i.e., replacing the dilatation catheter with an infusion catheter. The InfusaSleeve is compatible with standard guiding catheters, guide wires, and dilatation catheters.

3. Testing and evaluation included biocompatibility, hub to shaft tensile strength, tip to shaft tensile strength, infusion pathway fatigue, burst pressure, infusate flow, balloon fatigue, and animal testing. The test results verified that the performance of the InfusaSleeve is equivalent to or exceeds other commercially marketed infusion catheters.

4. The InfusaSleeve is similar in design and construction to the USCI Dorros™ Infusion/Probing Catheter manufactured by USCI, a division of C.R. Bard, the Target Therapeutics Tracker® Infusion Catheters and the SCIMED® DISPATCH™ Coronary Infusion Catheter.

---

**Source:** [https://fda.innolitics.com/device/K950615](https://fda.innolitics.com/device/K950615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com