THE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS

K950419 · Imed Corp. · FRN · Jun 21, 1995 · General Hospital

Device Facts

Record IDK950419
Device NameTHE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS
ApplicantImed Corp.
Product CodeFRN · General Hospital
Decision DateJun 21, 1995
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Orion™ Infusion System is intended for use in a health care facility to pump standard IV fluids, medications, whole blood and packed red blood cells into a patient in a controlled manner. The system is electrically powered, with back up battery power available, and uses linear peristaltic pumping action to infuse fluids. The Orion System is capable of detecting air-in-line and occlusions. The Orion System is also indicated for Epidural infusions. Appropriate warnings and precautions to safely and accurately infuse Epidural are provided in the product labeling.

Device Story

Modular infusion system comprising interface units and linear peristaltic pump modules; used in healthcare facilities to deliver IV fluids, medications, blood products, and epidural infusions. System detects air-in-line and occlusions. Features include drug name display, multichannel coordinated infusion, battery run-time indicator, and PC card capability. Operated by healthcare personnel to manage fluid delivery; provides controlled infusion rates to support patient therapy.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by comparative functional testing and biocompatibility data.

Technological Characteristics

Modular electromechanical system; linear peristaltic pumping mechanism; electrically powered with backup battery. Materials meet Tripartite Biocompatibility Guidance for externally communicating devices (indirect blood contact). Features: drug name display, multichannel coordination, battery indicator, PC card interface. Complies with applicable safety/performance standards.

Indications for Use

Indicated for patients in health care facilities requiring controlled infusion of IV fluids, medications, whole blood, packed red blood cells, or epidural infusions.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0} K950419 # SAFETY AND EFFECTIVENESS SUMMARY # 1.0 SUBMITTER INFORMATION IMED Corporation 9775 Businesspark Ave. San Diego, CA 92131-1699 JUN 21 2005 Contact Person: Ahmad Sajadi, Manager, Regulatory Affairs # 2.0 DEVICE NAME Trade Name: Common Names: Orion™ Infusion Pump and Administration Sets Infusion Pump Intravascular Administration Sets Classification Names: Infusion Pumps (21 CFR 880.5725), Class II IV Administration Sets (21 CFR 880.5440), Class II # 3.0 PREDICATE DEVICE The IMED Orion Infusion Pump and Administration Sets are substantially equivalent to the IMED Gemini PC-2TX Infusion Pump and Administration Sets. The Gemini devices have been reviewed by FDA and are currently being marketed by IMED Corporation. # 4.0 DESCRIPTION OF THE SUBJECT DEVICE The Orion Infusion System is a modular system that will allow the user to assemble infusion systems which meet a wide range of patient needs. The electromechanical portion of the system is assembled from three building blocks. Two of these blocks are interchangeable Interface Units. The third is a pump module, which physically pumps fluid to the patient when used in conjunction with the dedicated administration sets. One through four pump modules can be assembled/connected to each Interface Unit. Various configurations of the IV administration sets are available. The Orion Infusion System consists of the following components: - Advanced Interface Unit - Basic Interface Unit - Large Volume Parenteral Pump Unit - IV Administration Sets 679 nc bc ec us on is fu ibl be np an mi {1} SAFETY AND EFFECTIVENESS SUMMARY (CONT'D) Page two # 0 INTENDED USE OF THE SUBJECT DEVICE The Orion™ Infusion System is intended for use in a health care facility to pump standard IV fluids, medications, whole blood and packed red blood cells into a patient in a controlled manner. The system is electrically powered, with back up battery power available, and uses linear peristaltic pumping action to infuse fluids. The Orion System is capable of detecting air-in-line and occlusions. The Orion System is also indicated for Epidural infusions. Appropriate warnings and precautions to safely and accurately infuse Epidural are provided in the product labeling. # 0 TECHNOLOGICAL ASPECTS OF THE SUBJECT DEVICE The technological aspects of the subject devices are substantially equivalent to the technological aspects of the IMED Gemini PC-2TX Infusion Pump and Administration Sets. This is supported by the comparison of the design and component materials of both systems. The performance data resulting from the comparative functional testing also support the substantial equivalence claim to the predicate device. The Orion System complies with the applicable safety and/or performance standards. The Orion System contains the following additional features: - Drug Name Display - Multichannel Coordinated Infusion - Battery Run Time Indicator - PC Card Capability The Orion Administration Sets are equivalent to the Gemini Administration Sets. The materials are equivalent in that per the device category definition in the Tripartite Biocompatibility Guidance For Medical Devices both are "Externally Communicating Devices for Blood Path Indirect and Short-term Contact Duration." Consequently all materials must meet the same testing criteria as outlined in the Tripartite Guidance document. Biocompatibility data to support this claim is provided. The conclusion drawn from the Performance Data demonstrates that the IMED Orion Infusion Pump and Administration Sets are equivalent to a legally marketed device and that they perform as well as or better than the predicate device. 680 {2} # SAFETY AND EFFECTIVENESS SUMMARY (CONT'D) Page three # 7.0 CERTIFICATION I hereby certify that to the best of my knowledge all information contained in this Premarket Notification is truthful and accurate and that no material fact has been omitted. Ahmad Sajadi Manager, Regulatory Affairs 681 {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service JUN 21 1995 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Amhad Sajadi Manager, Regulatory Affairs IMED Corporation 9775 Businesspark Avenue San Diego, California 92131-1699 Re: K950419 Trade Name: The IMED Orion Infusion Pump and Administration Sets Regulatory Class: II Product Code: FRN, FPA Dated: May 9, 1995 Received: May 10, 1995 Dear Mr. Sajadi: We have reviewed your Section 510(k) notification of intent to market the device reference above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (act). The general controls provisions of the act include requirements for registration, listing of devices, good manufacturing practices, and labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulation. {4} Page 2 - Mr. Sajadi In addition, on August 16, 1993 the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirement of the regulation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447. This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diagnostic devices), promotion, or advertising, please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-302) at (301) 594-4639. Other general information on your responsibilities under the act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597. Timothy A. Ulatowski Acting Director Office of Device Evaluation Center for Devices and Radiological Health Enclosures ![img-0.jpeg](img-0.jpeg)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...