← Product Code [FOZ](/productcode/FOZ) · K946170

# COMFORT-Q SUBCUTANEOUS INJECTION SETS (K946170)

_B. Braun of America, Inc. · FOZ · Nov 3, 1995 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K946170

## Device Facts

- **Applicant:** B. Braun of America, Inc.
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Nov 3, 1995
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The intended use for administration of fluids or other therapeutic agents subcutaneously.

## Device Story

Comfort-Q Subcutaneous Injection Set; 23 gauge catheter over 26 gauge introducer needle; 42 inch microbore tubing; includes adhesive dressing. Used for subcutaneous delivery of fluids/therapeutic agents. Operated by healthcare professionals. Device provides fluid path for infusion; ensures precise administration. Benefits patient via reliable subcutaneous access.

## Clinical Evidence

Bench testing only. Includes biocompatibility testing per Tripartite Guidance, USP Class 6, cytotoxicity, hemolysis, and USP Physio-chemical testing. Finished products undergo sterility, pyrogenicity (LAL method), and physical testing per established Quality Control Test Procedures.

## Technological Characteristics

23 gauge catheter; 26 gauge introducer needle; 42 inch microbore tubing; adhesive dressing included. Materials meet USP Class 6, cytotoxicity, and hemolysis requirements. Sterilization and physical testing performed per GMP and Device Master Records.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- Subcutaneous Infusion Sets (Pharmacia Deltec) ([K801941](/device/K801941.md))
- Microcath Intravenous Catheter and Placement Unit and Catheter (Burron Medical, Inc.) (K9738885)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

K 94 6170

NOV

10(k) Summary of Safety and Effectiveness
in Accordance with SMDA '90

Summary of Safety and Effectiveness
in Accordance with SMDA '90

Braun of

er ca, Inc

December 16, 1994

Twelfth

enje

hlehem, P

18018

) 691-540

TACT: Ma

S. Alsberge, Regulatory Affairs Associate

DUCT NAME

Subcutaneous Injection Sets

E NAME:

mfort-Q Subcutaneous Injection Sets

SS I FICATI

I ME :

General

forital

Class 1

FPA, Intravascular Administration Set

21 CFR

3, 5440

STANTIAL

JULANCE TO

510 (k) num

K801941

K9738885

Device DESC

B. Bra
inters
Sets.
gauge
needle
dress i
per for
admini
subcut

|   | Name  |
| --- | --- |
|   | MICROCATH INTRAVENOUS CATHETER PLACEMENT UNIT AND CATHETER  |
|   | SUBCUTANEOUS INFUSION SET  |

Applicant

BURRON MEDICAL INC.

PHARMACIA DELTEC

PT ON:

0' America Inc. intends to introduce into
te commerce the Comfort-Q Subcutaneous Injection
the Comfort-Q Subcutaneous Injection et 1s a 23
3/1 in. (0.9cm.) catheter over 26 gauge intorducer
th 42 in. (106.7cm.) microbore tubing. Adhesive
closed. These all have similar designs and
characteristics. The intended use for
ation of fluids or other therapeutic agents
ecisly.

ATTACHMENT 2

ATT

{2}

MATERIAL :

B. Braun of America certifies that the biocompatibility tests recommended in the Tripartite Guidance for this category of contact duration will be completed for all the materials listed above for use in the manufacture of the device. All the materials listed above have also past requirements for USP Class 6, cytotoxicity, Hemolysis, and USP Physio-chemical testing.

SUBSTANTIAL EQUIVALENCE:

From a regulatory standpoint, the Comfort-Q Subcutaneous Injection Sets are equivalent in materials, form, and intended use to the Pharmacia Deltec (Subcutaneous Infusion Sets) and in materials to the Burron Medical, Inc. (Microcath Intravenous Catheter and Placement Unit and Catheter). There are no new issues of safety or effectiveness raised by the Comfort-Q Subcutaneous Injection Sets.

SAFETY AND EFFECTIVENESS:

All finished products are tested and must meet all required release specifications before distribution. The array of testing required for release include, but are not limited to; sterility, pyrogenicity (endotoxin/ LAL Method), physical testing, visual examination (in process and finished product).

The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures and parameters which conform to the product design specifications. The physical testing for the The Comfort-Q Subcutaneous Injection Sets is defined in detail in the "Device Master Records".

The testing instruction records for each of the individually required procedures are approved, released, distributed and revised in accordance with document control GMP's.

ATTACHMENT 2

ATT

---

**Source:** [https://fda.innolitics.com/device/K946170](https://fda.innolitics.com/device/K946170)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).