COMFORT-Q SUBCUTANEOUS INJECTION SETS

K946170 · B. Braun of America, Inc. · FOZ · Nov 3, 1995 · General Hospital

Device Facts

Record IDK946170
Device NameCOMFORT-Q SUBCUTANEOUS INJECTION SETS
ApplicantB. Braun of America, Inc.
Product CodeFOZ · General Hospital
Decision DateNov 3, 1995
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use for administration of fluids or other therapeutic agents subcutaneously.

Device Story

Comfort-Q Subcutaneous Injection Set; 23 gauge catheter over 26 gauge introducer needle; 42 inch microbore tubing; includes adhesive dressing. Used for subcutaneous delivery of fluids/therapeutic agents. Operated by healthcare professionals. Device provides fluid path for infusion; ensures precise administration. Benefits patient via reliable subcutaneous access.

Clinical Evidence

Bench testing only. Includes biocompatibility testing per Tripartite Guidance, USP Class 6, cytotoxicity, hemolysis, and USP Physio-chemical testing. Finished products undergo sterility, pyrogenicity (LAL method), and physical testing per established Quality Control Test Procedures.

Technological Characteristics

23 gauge catheter; 26 gauge introducer needle; 42 inch microbore tubing; adhesive dressing included. Materials meet USP Class 6, cytotoxicity, and hemolysis requirements. Sterilization and physical testing performed per GMP and Device Master Records.

Indications for Use

Indicated for the subcutaneous administration of fluids or other therapeutic agents.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Submission Summary (Full Text)

{1} K 94 6170 NOV 10(k) Summary of Safety and Effectiveness in Accordance with SMDA '90 Summary of Safety and Effectiveness in Accordance with SMDA '90 Braun of er ca, Inc December 16, 1994 Twelfth enje hlehem, P 18018 ) 691-540 TACT: Ma S. Alsberge, Regulatory Affairs Associate DUCT NAME Subcutaneous Injection Sets E NAME: mfort-Q Subcutaneous Injection Sets SS I FICATI I ME : General forital Class 1 FPA, Intravascular Administration Set 21 CFR 3, 5440 STANTIAL JULANCE TO 510 (k) num K801941 K9738885 Device DESC B. Bra inters Sets. gauge needle dress i per for admini subcut | | Name | | --- | --- | | | MICROCATH INTRAVENOUS CATHETER PLACEMENT UNIT AND CATHETER | | | SUBCUTANEOUS INFUSION SET | Applicant BURRON MEDICAL INC. PHARMACIA DELTEC PT ON: 0' America Inc. intends to introduce into te commerce the Comfort-Q Subcutaneous Injection the Comfort-Q Subcutaneous Injection et 1s a 23 3/1 in. (0.9cm.) catheter over 26 gauge intorducer th 42 in. (106.7cm.) microbore tubing. Adhesive closed. These all have similar designs and characteristics. The intended use for ation of fluids or other therapeutic agents ecisly. ATTACHMENT 2 ATT {2} MATERIAL : B. Braun of America certifies that the biocompatibility tests recommended in the Tripartite Guidance for this category of contact duration will be completed for all the materials listed above for use in the manufacture of the device. All the materials listed above have also past requirements for USP Class 6, cytotoxicity, Hemolysis, and USP Physio-chemical testing. SUBSTANTIAL EQUIVALENCE: From a regulatory standpoint, the Comfort-Q Subcutaneous Injection Sets are equivalent in materials, form, and intended use to the Pharmacia Deltec (Subcutaneous Infusion Sets) and in materials to the Burron Medical, Inc. (Microcath Intravenous Catheter and Placement Unit and Catheter). There are no new issues of safety or effectiveness raised by the Comfort-Q Subcutaneous Injection Sets. SAFETY AND EFFECTIVENESS: All finished products are tested and must meet all required release specifications before distribution. The array of testing required for release include, but are not limited to; sterility, pyrogenicity (endotoxin/ LAL Method), physical testing, visual examination (in process and finished product). The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures and parameters which conform to the product design specifications. The physical testing for the The Comfort-Q Subcutaneous Injection Sets is defined in detail in the "Device Master Records". The testing instruction records for each of the individually required procedures are approved, released, distributed and revised in accordance with document control GMP's. ATTACHMENT 2 ATT
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...