METAL-LINED TRANSCEND CERAMIC BRACKET

K944286 · Unitek Corp. · DYW · Nov 29, 1994 · Dental

Device Facts

Record IDK944286
Device NameMETAL-LINED TRANSCEND CERAMIC BRACKET
ApplicantUnitek Corp.
Product CodeDYW · Dental
Decision DateNov 29, 1994
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2

Device Story

Metal-Lined Transcend™ Ceramic Bracket; orthodontic device used for tooth alignment; ceramic bracket body with metal liner; bonded to teeth by orthodontist; provides mechanical attachment for orthodontic archwires; facilitates tooth movement through application of force.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Ceramic bracket body with metal liner; orthodontic application; class II device; product code DYW.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Submission Summary (Full Text)

{1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service MAR 24 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Marlyn Scheff Regulatory Affairs Manager 3M Unitek 3M Dental Products Division 2724 South Peck Road Monrovia, California 91016-5097 Re: K944286 Trade/Device Name: Metal-Lined Transcend™ Ceramic Bracket Regulation Number: 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: DYW Dated: September 1, 1994 Received: September 2, 1994 Dear Ms. Scheff: This letter corrects our substantially equivalent letter of September 1, 1994. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2} Page-2 Ms. Scheff Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Chiu Lin Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...