MODEL ME 300 MUSCLE TESTER
K943535 · Regulatory Strategies, Inc. · HCC · Jun 25, 1997 · Neurology
Device Facts
| Record ID | K943535 |
| Device Name | MODEL ME 300 MUSCLE TESTER |
| Applicant | Regulatory Strategies, Inc. |
| Product Code | HCC · Neurology |
| Decision Date | Jun 25, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.5050 |
| Device Class | Class 2 |
Intended Use
The intended use of the Model ME300 Muscle Tester is to record EMG sensory feedback signals to be used in various programs of muscle re-education and relaxation training.
Device Story
ME300 is a portable, battery-powered EMG recorder. It captures electrophysiological signals from muscle activity via conductive adhesive skin electrodes. The device features two-channel signal acquisition. It does not deliver electrical stimulation to the body. Signals are displayed or stored for later review. Data is transferred to a PC via a standard fiber optic cable. Used in clinical or training settings for muscle re-education and relaxation programs. Provides visual feedback to the user/clinician to assist in monitoring muscle activity, facilitating patient training and therapeutic progress.
Clinical Evidence
No clinical data. Bench testing only, including electrical and EMC safety testing per IEC 601-1-2, IEC 801-2, IEC 801-3, and EN55011.
Technological Characteristics
Portable, battery-powered EMG recorder. Two-channel signal acquisition via conductive adhesive skin electrodes. Fiber optic data transfer to PC. Complies with IEC 601-1-2, IEC 801-2, IEC 801-3, and EN55011 standards. Passive signal recording; no electrical stimulation output.
Indications for Use
Indicated for patients requiring muscle re-education and relaxation training via EMG sensory feedback. Device is for prescription use.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
Predicate Devices
- Myotrac from Thought Technology (K881416)
- Myoexcerciser from Verimed (K832714)
Related Devices
- K963202 — MEGA ME 3000 PROFESSIONAL MUSCLE TESTER · Mega Electronics , Ltd. · May 19, 1997
- K972487 — EMG RETRAINER · Chattanooga Group, Inc. · Sep 26, 1997
- K970051 — MUSCLESENSE II CLINICAL ELECTROMYOGRAPHY DEVICE · Smith and Nephew Donjoy, Inc. · Jul 10, 1997
- K013399 — MODEL MES-9000/EMG SYSTEM · Myotronics-Noromed, Inc. · Nov 13, 2001
- K111687 — EMG SYSTEM · Myotronics-Noromed, Inc. · Sep 29, 2011
Submission Summary (Full Text)
{0}
K943535
# 510(k) Summary
JUN 25 1997
## General Information
| Classification | Class II |
| --- | --- |
| Common Name | Biofeedback device, EMG Device |
| Trade Name | ME300 Muscle Tester |
| Manufacturer | Mega Electronics, Ltd.
Savilahdentie 6
PO Box 1750
70211 Kuopio
Finland |
| | tel: 358 71 282 8959 |
| Submitted by | Regulatory Strategies, Inc.
Consultant to Mega Electronics, Ltd. |
## Predicate Devices
| Myotrac from Thought Technology | K881416 |
| --- | --- |
| Myoexcerciser from Verimed | K832714 |
## Device Description
The ME300 is a portable, battery powered EMG recorder.
## Technological Characteristics
The ME300 is an electronic device used to gather and display electrophysiological signals obtained from muscle activity. These signals are used in programs related to muscle re-education and relaxation training.
The signals are received via conductive adhesive skin electrodes and may be displayed or stored for display at a later time. Signals may be gathered from two channels. Data downloading or transfer is via a standard fiber optic cable to a PC. The product does not relay any electrical or stimulatory signals to the body. It is only a signal recorder with display capabilities.
{1}
# Intended Use
The intended use of the Model ME300 Muscle Tester is to record EMG sensory feedback signals to be used in various programs of muscle re-education and relaxation training.
# Testing
The ME300 was tested for electrical and EMC safety by a recognized third party testing facility. The product meets the requirements of IEC 601-1-2, IEC 801-2, IEC 801-3, EN55011 and the EMC Directive.
# Summary of Equivalence
The ME300 offers the user a means to gather electrophysiological signals (EMG) and display these signals for the purpose of muscle re-education and relaxation training. The indications for use and the basic overall design are either identical or substantially equivalent to the predicate products.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Gregory J. Mathison
Regulatory Strategies, Inc.
1660 Riverton Point
Eagan, Minnesota 55122
JUN 25 1997
Re: K943535
Trade Name: Model ME300 Muscle Tester
Regulatory Class: II
Product Code: 84HCC
Dated: March 28, 1997
Received: March 28, 1997
Dear Mr. Mathison:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Mr. Gregory J. Mathison
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
{4}
Page 1 of 1
510(k) Number (if known): K943535
Device Name: ME300 Muscle Tester
Indications For Use: ME300 is an electronic device used to gather and display electrophysiological signals obtained from muscle activity. These signals are used in programs related to muscle re-education and relaxation training.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K943535
Prescription Use ☑
(Per 21 CFR 801.109)